Protein quantification is a biochemical experiment used to determine the concentration of proteins contained in a sample by using absorbance or color reactions.
Representative methods include the Bradford assay, Lowry method, BCA assay, Biuret method, and UV absorption method.
In many cases, a calibration curve is prepared using standard proteins of known concentration, and the concentration of an unknown sample is determined from its absorbance.
In a discussion of protein quantification, it is not sufficient simply to write that “the concentration was determined from absorbance” or “the calibration curve was linear.”
It is necessary to consider whether the calibration curve has good linearity, whether the absorbance of the unknown sample falls within the calibration-curve range, whether blank correction was performed appropriately, whether the dilution factor was correctly reflected, and whether salts, surfactants, reducing agents, or other substances in the sample affected the color reaction.
This article clearly explains how to interpret the results of protein quantification experiments, how to discuss calibration curves, absorbance, and unknown-sample concentrations, sources of error, points for improvement, and discussion examples that can be used in reports.
Note:
This article is a reference intended to assist with discussions of protein quantification results obtained in biochemical experiments at universities and similar institutions.
For the actual quantification method, standard protein, reagents, measurement wavelength, reaction time, dilution conditions, waste-liquid disposal, and safety precautions, always follow the instructions in your university’s laboratory manual and those given by your instructor or TA.
- What Is Protein Quantification?
- Main Items to Include in the Results
- Reference Experimental Values for Protein Quantification and Examples of Standard-Curve Analysis
- Reference Experimental Conditions
- Example Measurement of Standard Proteins
- Concept of Blank Correction
- Example Standard-Curve Equation
- Example Calculation of an Unknown Sample Concentration
- Example Measurement Results for Unknown Samples
- Checking the Linearity of the Standard Curve
- Handling Values Outside the Calibration-Curve Range
- Differences in Final Concentration Depending on Dilution Factor
- Example of an Outlier
- Effect of Errors in Standard-Solution Preparation
- Error Caused by the Sample’s Own Color or Turbidity
- Example Comparison of Concentration Differences Among Measurement Methods
- Examples of the Effects of Interfering Substances
- Example of Calculating Relative Standard Deviation From Repeated Measurements
- Example of How to Write the Results
- Points for Connecting the Results to the Discussion
- Example Discussion
- Summary
- What Is a Calibration Curve?
- Discussion of Calibration-Curve Linearity
- Relationship Between Absorbance and Protein Concentration
- Discussion of Blank Correction
- Concept of Determining the Concentration of an Unknown Sample
- Discussion of the Dilution Factor
- When the Unknown Sample Is Outside the Calibration-Curve Range
- Differences Between Standard Proteins and Unknown Samples
- Discussion of the Bradford Assay
- Discussion of the Lowry Method and BCA Assay
- Discussion of the UV Absorption Method
- Effects of Interfering Substances in the Sample
- Errors Caused by Pipetting
- Errors Caused by Reaction Time
- Errors Caused by Contamination of Cuvettes and Plates
- Discussion When Absorbance Is Too High
- Discussion When Absorbance Is Too Low
- Causes of Overestimating Protein Concentration
- Causes of Underestimating Protein Concentration
- Discussion of Variation in Repeated Measurements
- Discussion When Calibration-Curve Points Deviate
- Discussion of the Calibration-Curve Intercept
- Effects of Protein Denaturation and Degradation
- Buffer Conditions of the Standard Curve and Unknown Sample
- When the Results Can Be Considered Good
- Example Discussion When the Experiment Did Not Go Well
- How to Write Points for Improvement
- Difference Between a Superficial Discussion and a Good Discussion
- Examples of Expressions That Can Be Used in Reports
- Points to Check When Discussing Protein Quantification
- Summary
What Is Protein Quantification?
Protein quantification is an experiment used to numerically determine how much protein is contained in a sample.
In many quantification methods, proteins react with reagents to produce color, and the absorbance is measured.
Within the range where absorbance changes according to protein concentration, protein concentration can be determined from absorbance.
However, absorbance does not directly indicate the “amount of protein itself.”
A calibration curve must be prepared using a standard protein, and the relationship obtained must be used to convert the absorbance of an unknown sample into concentration.
Therefore, the accuracy of the calibration curve greatly affects the reliability of protein quantification.
Example Discussion:
In this experiment, the color reaction between proteins and reagents was used to determine the protein concentration in the sample from absorbance.
Although absorbance changes according to protein concentration, a calibration curve prepared using a standard protein is required to determine the concentration of an unknown sample.
Therefore, the linearity and measurement range of the calibration curve greatly affect the reliability of the quantification results.
Main Items to Include in the Results
In the results of protein quantification, organize the concentrations and absorbance values of the standard protein, the calibration curve, the absorbance of the unknown sample, the dilution factor, and the calculated protein concentration.
If multiple measurements were performed, showing the mean and variation also makes the discussion easier.
Main Items to Include in the Results
- Quantification method used
- Type of standard protein
- Standard protein concentration
- Absorbance of each standard solution
- Blank absorbance
- Absorbance after blank correction
- Calibration-curve equation
- Coefficient of determination R2
- Absorbance of the unknown sample
- Dilution factor
- Protein concentration of the unknown sample
- Mean of repeated measurements
- Standard deviation and variation
- Sources of error and points for improvement
Example of How to Write the Results:
A calibration curve was prepared from the concentrations and absorbance values of standard protein solutions.
The calibration curve showed an approximately linear relationship, and the absorbance of the unknown sample fell within the calibration-curve range.
Therefore, the protein concentration of the unknown sample was determined using the calibration-curve equation, and the concentration in the original sample was calculated by taking the dilution factor into account.
Reference Experimental Values for Protein Quantification and Examples of Standard-Curve Analysis
Here, standard curves, absorbance values, unknown-sample concentrations, dilution factors, and measurement errors obtained in protein quantification experiments are organized as reference experimental values that are easy to discuss in reports.
In protein quantification, standard proteins of known concentration are measured to prepare a standard curve, and the absorbance of an unknown sample is applied to that standard curve to determine its concentration.
When considering the reliability of measured values, it is important to confirm the linearity of the standard curve, blank correction, dilution factor, measurement range, and variation among repeated measurements.
Reference Experimental Conditions
| Item | Details |
|---|---|
| Samples measured | Protein standard solutions, unknown protein samples |
| Standard protein | BSA standard solution |
| Quantification methods | Bradford assay, BCA assay, Lowry method, etc. |
| Measurement wavelength | 595 nm, 562 nm, 750 nm, etc., depending on the measurement method |
| Evaluation items | Standard curve, unknown-sample concentration, dilution factor, blank correction, linearity, outliers, sources of error |
| Concentration expression | μg/mL, mg/mL, BSA-equivalent concentration |
Example Measurement of Standard Proteins
The following is a reference example in which the relationship between protein concentration and absorbance was measured using BSA standard solutions.
Here, the measurement wavelength is 595 nm.
| BSA Concentration | Absorbance 1st | Absorbance 2nd | Absorbance 3rd | Mean Absorbance | Standard Deviation |
|---|---|---|---|---|---|
| 0 μg/mL | 0.050 | 0.052 | 0.049 | 0.050 | 0.002 |
| 100 μg/mL | 0.128 | 0.131 | 0.129 | 0.129 | 0.002 |
| 200 μg/mL | 0.208 | 0.205 | 0.207 | 0.207 | 0.002 |
| 400 μg/mL | 0.361 | 0.365 | 0.363 | 0.363 | 0.002 |
| 600 μg/mL | 0.518 | 0.522 | 0.519 | 0.520 | 0.002 |
| 800 μg/mL | 0.668 | 0.671 | 0.666 | 0.668 | 0.003 |
| 1000 μg/mL | 0.808 | 0.814 | 0.812 | 0.811 | 0.003 |
Absorbance increases as the concentration of the standard protein increases.
Because the standard deviations of the repeated measurements are small, the reproducibility of the standard-solution measurements is considered relatively good in this example.
Concept of Blank Correction
Even a blank containing no protein may show absorbance because of the color of the reagent itself or the effects of the cuvette or plate.
Therefore, the absorbance of the blank is subtracted from the absorbance of the standard solutions and unknown samples.
Blank-corrected absorbance = Measured absorbance − Blank absorbance
| BSA Concentration | Mean Absorbance | Blank Absorbance | Corrected Absorbance |
|---|---|---|---|
| 0 μg/mL | 0.050 | 0.050 | 0.000 |
| 100 μg/mL | 0.129 | 0.050 | 0.079 |
| 200 μg/mL | 0.207 | 0.050 | 0.157 |
| 400 μg/mL | 0.363 | 0.050 | 0.313 |
| 600 μg/mL | 0.520 | 0.050 | 0.470 |
| 800 μg/mL | 0.668 | 0.050 | 0.618 |
| 1000 μg/mL | 0.811 | 0.050 | 0.761 |
By performing blank correction, background absorbance originating from reagents and the measuring instrument can be subtracted, making it easier to handle absorbance changes derived from proteins.
Example Standard-Curve Equation
Suppose the following approximation equation is obtained when a standard curve is prepared using blank-corrected absorbance values.
Corrected absorbance = 0.00076 × Protein concentration (μg/mL) + 0.002
Solving this equation for concentration gives the following equation for determining the protein concentration of an unknown sample.
Protein concentration (μg/mL) = (Corrected absorbance − 0.002) ÷ 0.00076
If the unknown sample lies within the range of the standard curve, its concentration can be calculated using this equation.
Example Calculation of an Unknown Sample Concentration
Consider a case in which Unknown Sample A was diluted 10-fold and measured, giving a measured absorbance of 0.432 and a blank absorbance of 0.050.
Corrected absorbance = 0.432 − 0.050 = 0.382
Concentration after dilution = (0.382 − 0.002) ÷ 0.00076 = 500 μg/mL
Because this is the concentration after 10-fold dilution, the original sample concentration is as follows.
Original sample concentration = 500 × 10 = 5000 μg/mL = 5.0 mg/mL
Therefore, the protein concentration of Unknown Sample A is determined to be 5.0 mg/mL.
Example Measurement Results for Unknown Samples
| Sample | Dilution Factor | Measured Absorbance | Corrected Absorbance | Concentration After Dilution | Original Sample Concentration | Judgment |
|---|---|---|---|---|---|---|
| Unknown Sample A | 10-fold | 0.432 | 0.382 | 500 μg/mL | 5.0 mg/mL | Within range |
| Unknown Sample B | 5-fold | 0.278 | 0.228 | 297 μg/mL | 1.49 mg/mL | Within range |
| Unknown Sample C | 20-fold | 0.715 | 0.665 | 872 μg/mL | 17.4 mg/mL | High-concentration side |
| Unknown Sample D | 2-fold | 0.092 | 0.042 | 52.6 μg/mL | 105 μg/mL | Low-concentration side |
Unknown Sample C lies on the high-concentration side of the standard curve, while Unknown Sample D lies on the low-concentration side.
Both values can be calculated, but adjusting the dilution factor so that the absorbance falls near the center of the standard curve allows a more reliable measurement.
Checking the Linearity of the Standard Curve
Before using a standard curve, it is necessary to confirm that it is linear within the measurement range.
On the high-concentration side, absorbance may become saturated and deviate from the straight line.
| Concentration Range | Approximation Equation | Coefficient of Determination R2 | Evaluation |
|---|---|---|---|
| 0–800 μg/mL | A = 0.00077C + 0.001 | 0.999 | High linearity |
| 0–1000 μg/mL | A = 0.00076C + 0.002 | 0.998 | Good |
| 0–1500 μg/mL | A = 0.00062C + 0.035 | 0.970 | Deviation on the high-concentration side |
| 400–1500 μg/mL | A = 0.00050C + 0.120 | 0.940 | Low linearity |
Even if the coefficient of determination is high, it does not necessarily mean that the entire range is accurate.
It is necessary to check whether the standard-solution points lie on the straight line and whether there is bias on the low- or high-concentration side.
Handling Values Outside the Calibration-Curve Range
If the absorbance of an unknown sample lies outside the standard-curve range, determining its concentration by extrapolation may cause a large error.
| Sample | Corrected Absorbance | Standard-Curve Range | Judgment | Response |
|---|---|---|---|---|
| Sample A | 0.382 | 0.000–0.761 | Within range | Use for calculation |
| Sample B | 0.820 | 0.000–0.761 | Too high | Dilute and remeasure |
| Sample C | 0.015 | 0.000–0.761 | Too low | Concentrate or increase sample amount |
| Sample D | 1.100 | 0.000–0.761 | Clearly outside the range | Dilute substantially and measure |
Outside the standard-curve range, there is less assurance that the relationship between absorbance and concentration remains valid.
Samples outside the range should be diluted or concentrated so that they fall within the measurement range and then remeasured.
Differences in Final Concentration Depending on Dilution Factor
Even when the concentration determined from the same absorbance is the same, the original sample concentration changes depending on the dilution factor used during measurement.
| Corrected Absorbance | Concentration After Dilution | Dilution Factor | Original Sample Concentration | Converted Value |
|---|---|---|---|---|
| 0.382 | 500 μg/mL | 1-fold | 500 μg/mL | 0.50 mg/mL |
| 0.382 | 500 μg/mL | 5-fold | 2500 μg/mL | 2.5 mg/mL |
| 0.382 | 500 μg/mL | 10-fold | 5000 μg/mL | 5.0 mg/mL |
| 0.382 | 500 μg/mL | 20-fold | 10000 μg/mL | 10.0 mg/mL |
In a report, clearly state whether the concentration determined from the standard curve is the “concentration of the diluted sample measured” or the “concentration converted back to the original sample.”
Example of an Outlier
If only one measurement deviates greatly during repeated measurements, possible causes include bubbles, insufficient mixing, pipetting errors, or contamination of the cuvette.
| Sample | 1st | 2nd | 3rd | Difference From Mean | Judgment |
|---|---|---|---|---|---|
| Standard 400 μg/mL | 0.361 | 0.365 | 0.363 | Small | Good |
| Unknown Sample A | 0.430 | 0.432 | 0.434 | Small | Good |
| Unknown Sample B | 0.278 | 0.280 | 0.410 | 3rd is high | Possible outlier |
| Unknown Sample C | 0.715 | 0.710 | 0.705 | Small | Good |
If only the third measurement of Unknown Sample B is high, bubbles or contamination may have caused the absorbance to appear high.
If an outlier is excluded, the reason should be clearly stated and remeasurement should be performed if possible.
Effect of Errors in Standard-Solution Preparation
If the concentrations of standard solutions are prepared incorrectly, the entire standard curve shifts and the unknown-sample concentrations are also calculated incorrectly.
| Condition | Characteristic of Standard Curve | Effect on Unknown Sample | Direction of Discussion |
|---|---|---|---|
| Standard solutions are correct | High linearity | Concentration calculation is reliable | Reference |
| Only high-concentration standard is too concentrated | High-concentration side shifts upward | Slope becomes larger | Possibility of underestimating unknown sample |
| Only low-concentration standard is too dilute | Low-concentration side shifts downward | Linearity decreases | Suspect error in standard preparation |
| All standard solutions are too dilute | Apparent slope is small | Possibility of overestimating unknown sample | Check stock concentration and dilution calculation |
Error Caused by the Sample’s Own Color or Turbidity
If an unknown sample has color or turbidity, absorbance other than that produced by the protein reaction is included in the measured value.
In this case, correction can be facilitated by using a sample blank.
| Measurement Condition | Absorbance | Meaning | Correction |
|---|---|---|---|
| Sample + quantification reagent | 0.520 | Protein reaction + sample’s own color | Overestimated if used as is |
| Sample + water | 0.090 | Sample’s own color and turbidity | Sample blank |
| After correction | 0.430 | Value closer to protein reaction | Use for concentration calculation |
In colored or turbid samples, failure to subtract the sample blank may cause the protein concentration to be overestimated.
Example Comparison of Concentration Differences Among Measurement Methods
Even for the same sample, the protein concentration obtained may differ slightly depending on the measurement method.
| Measurement Method | Measurement Wavelength | Unknown Sample A Concentration | Characteristic | Direction of Discussion |
|---|---|---|---|---|
| Bradford assay | 595 nm | 5.0 mg/mL | Simple and rapid | Differences in color development depending on protein type |
| BCA assay | 562 nm | 5.4 mg/mL | Relatively broad linear range | Be careful of reducing agents |
| Lowry method | 750 nm | 5.2 mg/mL | High sensitivity | Operation is somewhat complicated |
| A280 method | 280 nm | 4.7 mg/mL | No reagent required | Effects of nucleic acids and aromatic residues |
Protein quantification methods are based on different reaction principles.
Therefore, even if the values are not exactly the same, it is important to discuss the results while considering the characteristics of each measurement method.
Examples of the Effects of Interfering Substances
| Coexisting Substance | Trend in Measured Value | Possible Cause | Example Response |
|---|---|---|---|
| Surfactant | May appear low or high | Interferes with dye binding or reaction | Select an appropriate quantification method |
| Reducing agent | May appear high in the BCA assay | Affects reduction of copper ions | Confirm with a blank or another method |
| High salt concentration | Absorbance shifts | Affects reaction or solution state | Dilution, desalting |
| Nucleic acids | May appear high in the A280 method | Nucleic acids also absorb ultraviolet light | Check A260/A280 or use another method |
| Sample turbidity | Appears high | Light scattering | Centrifugation, filtration, sample blank |
When interfering substances are present, measured values may not reflect protein concentration alone.
Possible measures include preparing standard solutions in the same buffer as the sample, using a sample blank, or confirming the result with another method.
Example of Calculating Relative Standard Deviation From Repeated Measurements
To evaluate measurement variation, the mean, standard deviation, and relative standard deviation may be calculated.
| Measurement | Unknown Sample A Concentration |
|---|---|
| 1st | 4.95 mg/mL |
| 2nd | 5.02 mg/mL |
| 3rd | 5.05 mg/mL |
Mean = 5.01 mg/mL
Standard deviation = 0.05 mg/mL
Relative standard deviation (RSD) = 0.05 ÷ 5.01 × 100 = 1.0%
When the RSD is small, the reproducibility of repeated measurements is considered relatively good.
Example of How to Write the Results
A standard curve for protein quantification was prepared using BSA standard solutions.
The blank absorbance was 0.050, and this value was subtracted from the mean absorbance of each standard solution for correction.
As a result, absorbance increased with increasing protein concentration over the range of 0–1000 μg/mL, and the standard curve was expressed as A = 0.00076C + 0.002.
When Unknown Sample A was diluted 10-fold and measured, the measured absorbance was 0.432.
The absorbance after blank correction was 0.382, and substitution into the standard curve gave a concentration after dilution of 500 μg/mL.
Because the original sample had been diluted 10-fold, the original protein concentration was 5.0 mg/mL.
In Unknown Sample C, the corrected absorbance was 0.665 and was located on the high-concentration side of the standard curve.
In this case, although the value lies within the measurement range, it may be affected by saturation of color development or deviation from linearity.
For more accurate measurement, it is desirable to further dilute the sample and remeasure at an absorbance near the center of the standard curve.
Points for Connecting the Results to the Discussion
In a discussion of protein quantification, it is important to explain not only the calculated concentration but also the reliability of the standard curve, the measurement range, and the sources of error.
- Has a standard curve been prepared from the standard solutions?
- Has blank correction been performed to subtract background absorbance?
- Can the concentration be calculated by substituting the absorbance of the unknown sample into the standard-curve equation?
- Has the original sample concentration been restored by multiplying by the dilution factor?
- Has it been confirmed that the absorbance of the unknown sample lies within the standard-curve range?
- Can the possibility of absorbance saturation and reduced linearity on the high-concentration side be discussed?
- If an outlier is present, can bubbles, insufficient mixing, pipetting errors, contamination of the cuvette, and similar causes be explained?
- Can it be discussed that errors in preparing the standard solutions affect the entire standard curve?
- Can it be explained that the sample’s own color, turbidity, and interfering substances affect absorbance?
- Can differences in results among measurement methods be explained in relation to differences in reaction principles?
Example Discussion
In this experiment, a standard curve was prepared using BSA standard solutions, and the protein concentration in an unknown sample was determined.
Absorbance increased as the concentration of the standard solutions increased, and a good linear relationship was obtained over the range of 0–1000 μg/mL.
Therefore, when the absorbance of an unknown sample falls within this range, the concentration can be determined using the standard curve.
For Unknown Sample A, the measured absorbance of the 10-fold diluted solution was 0.432.
Subtracting the blank absorbance of 0.050 gave a corrected absorbance of 0.382, and the concentration after dilution was determined to be 500 μg/mL from the standard curve.
Multiplying this value by the dilution factor of 10 gave an original sample protein concentration of 5.0 mg/mL.
Therefore, both blank correction and dilution-factor correction are important in concentration calculations.
Possible sources of measurement error include errors in preparation of the standard solutions, variation in pipetting, differences in reaction time, insufficient mixing, bubbles, and contamination of the cuvette.
In absorbance measurements in particular, bubbles and turbidity may scatter light and produce absorbance values higher than the actual values.
In addition, if the concentrations of the standard solutions are prepared incorrectly, the slope of the standard curve changes and affects all calculations of unknown-sample concentrations.
When absorbance lies on the high-concentration side, as in Unknown Sample C, the color reaction may become saturated and the linear relationship between absorbance and concentration may break down.
If values outside the standard-curve range are extrapolated, the concentration may be greatly misestimated.
Therefore, unknown samples should be diluted as necessary and adjusted so that they are measured near the center of the standard curve.
Furthermore, if the sample itself has color or turbidity, absorbance unrelated to the protein reaction may be included in the measured value.
In this case, a sample blank must be measured and the absorbance originating from the sample itself subtracted.
In addition, because interfering substances such as surfactants, reducing agents, high concentrations of salts, and nucleic acids may affect measured values, it is important to select an appropriate quantification method according to the sample composition.
Summary
In protein quantification, a standard curve is prepared using a standard protein, and the concentration of an unknown sample is determined from its absorbance.
To obtain an accurate concentration, blank correction, dilution-factor correction, measurement within the standard-curve range, and confirmation of repeated measurements are important.
This reference example covered standard curves, blank correction, calculation of unknown-sample concentrations, linearity, handling of values outside the calibration-curve range, dilution factors, outliers, errors in standard-solution preparation, sample blanks, interfering substances, and differences among measurement methods.
In a report, it is useful to clearly show the concentration-calculation procedure and discuss both the reliability of the measured values and the sources of error.
What Is a Calibration Curve?
A calibration curve is a graph showing the relationship between a standard sample of known concentration and the measured value.
In protein quantification, the standard protein concentration is often plotted on the horizontal axis and absorbance on the vertical axis.
Within the range where absorbance is proportional to concentration, the measurement points approach a straight line.
By applying the absorbance of an unknown sample to the calibration curve, the protein concentration of the unknown sample can be determined.
However, if the absorbance of the unknown sample lies outside the calibration-curve range, the result becomes an extrapolation and reliability decreases.
In that case, the sample must be diluted or remeasured.
Example Discussion:
A calibration curve was prepared from the concentrations and absorbance values of the standard protein.
Within the measurement range, absorbance was approximately proportional to concentration and a linear relationship was obtained.
By using this calibration curve, the protein concentration of the unknown sample could be calculated from its absorbance.
Discussion of Calibration-Curve Linearity
If the calibration curve has good linearity, the relationship between standard protein concentration and absorbance is considered stable.
When the coefficient of determination R2 of the approximation line is close to 1, the measurement points are considered to fit the straight line well.
However, a high R2 alone does not necessarily mean that all measurements are correct.
On the low-concentration side, absorbance is small and the effects of blank correction and measurement noise may become large.
On the high-concentration side, the color reaction or absorbance measurement may become saturated and deviate from the straight line.
Therefore, it is important to confirm which concentration range can be used as a linear range in the calibration curve.
Example Discussion:
The R2 of the calibration curve was close to 1, and a good linear relationship was observed between standard protein concentration and absorbance.
Therefore, protein concentration is considered to be determinable from absorbance within the measurement range.
However, if measurement points on the high-concentration side deviate from the straight line, saturation of the color reaction or limitations of the absorbance measurement range may have affected the result.
Relationship Between Absorbance and Protein Concentration
In protein quantification, as the protein concentration increases, the color reaction often becomes stronger and absorbance becomes larger.
This is because the intensity of the color produced by the reaction between proteins and the reagent increases according to the amount of protein.
However, absorbance is proportional to concentration only within a certain range.
If the concentration is too high, the reagent may become insufficient, the color reaction may become saturated, or the absorbance may exceed the linear range of the measuring instrument.
In such cases, directly converting absorbance to concentration causes a large error.
Example Discussion:
Absorbance increased as the protein concentration increased.
This is considered to have occurred because the color reaction became stronger as the amount of protein increased, resulting in greater absorption at the measurement wavelength.
However, because the proportional relationship with concentration may break down in regions where absorbance is too high, unknown samples must be measured within the linear range of the calibration curve.
Discussion of Blank Correction
Blank correction is the operation of subtracting absorbance caused by reagents and solvents other than proteins.
A blank is often prepared by adding the same reagents under conditions without protein.
Blank correction reduces the effects of absorbance originating from the reagents, cuvette, and solvent.
If blank correction is insufficient, the absorbance of the unknown sample may be overestimated or underestimated.
Particularly for low-concentration samples, even a slight deviation in the blank can have a large effect on concentration calculations.
Example Discussion:
Blank correction was used to subtract absorbance originating from the reagents and solvent.
If blank correction is insufficient, absorbance not originating from proteins is also included when converting absorbance into concentration, which may cause the protein concentration to be overestimated.
Particularly for low-concentration samples, even a slight deviation in blank absorbance greatly affects the result.
Concept of Determining the Concentration of an Unknown Sample
The protein concentration of an unknown sample is determined by substituting its absorbance into the calibration-curve equation.
For example, if the calibration-curve equation is y = ax + b, the absorbance of the unknown sample is substituted for y and the concentration is determined as x.
If the sample was diluted before measurement, the original sample concentration is then determined by multiplying by the dilution factor.
Original sample concentration = Measured sample concentration × Dilution factor
In a report, it is important to distinguish whether the concentration determined from the calibration curve is the “concentration of the diluted measurement sample” or the “concentration of the original sample.”
Example Discussion:
The absorbance of the unknown sample was substituted into the calibration-curve equation to determine the protein concentration in the measurement sample.
Furthermore, because the sample had been diluted before measurement, the original sample protein concentration was calculated by multiplying the determined concentration by the dilution factor.
If the dilution factor is not taken into account, the original sample concentration will be underestimated, so caution is required.
Discussion of the Dilution Factor
In protein quantification, when the concentration of an unknown sample is too high, the sample is diluted so that it falls within the calibration-curve range.
Dilution allows the absorbance to be brought into an appropriate range, but the dilution factor must be correctly reflected in the calculation.
If the dilution factor is incorrect, the final protein concentration will deviate greatly.
For example, if the concentration obtained from a 10-fold diluted sample is written directly as the original sample concentration, the value will be one-tenth of the actual concentration.
Example Discussion:
The unknown sample was diluted so that its absorbance would fall within the calibration-curve range.
Although the concentration after dilution can be determined from the calibration curve, the dilution factor must be multiplied to determine the concentration of the original sample.
If the dilution factor is handled incorrectly, the protein concentration may be greatly underestimated or overestimated, so it is important to clearly show the calculation process.
When the Unknown Sample Is Outside the Calibration-Curve Range
If the absorbance of an unknown sample is higher than that of the highest concentration on the calibration curve, determining the concentration directly results in extrapolation.
Extrapolation assumes that the linear relationship of the calibration curve continues outside the measured range, so reliability decreases.
On the high-concentration side, the color reaction may also be saturated.
Conversely, if the absorbance of the unknown sample is too low, the effects of the blank and measurement noise become large.
In this case, possible measures include increasing the sample amount, concentrating the sample, or using a more sensitive method.
Example Discussion:
If the absorbance of the unknown sample lies outside the calibration-curve range, conversion to concentration becomes an extrapolation and reliability decreases.
Particularly on the high-concentration side, the color reaction may become saturated and absorbance may no longer be proportional to concentration.
Therefore, unknown samples must be appropriately diluted and measured within the linear range of the calibration curve.
Differences Between Standard Proteins and Unknown Samples
In protein quantification, proteins such as BSA may be used as standard proteins.
However, depending on the quantification method, the intensity of color development may differ depending on the type of protein.
In other words, even at the same mass concentration, the absorbance of a standard protein and an unknown sample may not be exactly the same.
In the Bradford assay, the dye binds mainly depending on specific amino acid residues and the properties of proteins, so the response may differ among protein types.
In the UV absorption method, differences in the content of aromatic amino acids such as tryptophan and tyrosine affect absorbance.
Example Discussion:
The BSA used as the standard protein and the proteins in the unknown sample may respond differently to the color reaction.
Depending on the protein quantification method, differences in amino acid composition and three-dimensional structure may cause different absorbance values even at the same concentration.
Therefore, the concentration determined from the calibration curve may contain errors caused by differences in properties between the standard protein and the unknown sample.
Discussion of the Bradford Assay
The Bradford assay is a quantification method that uses the change in absorbance caused when Coomassie Brilliant Blue dye binds to proteins.
Because it can be measured in a relatively short time and has high sensitivity, it is commonly used in biochemical experiments.
In the Bradford assay, the ease with which the dye binds may differ depending on the type of protein.
In addition, some substances such as surfactants may affect the measurement.
Because the linear range of the calibration curve is limited, dilution of unknown samples is important.
Example Discussion:
In the Bradford assay, protein concentration was determined from changes in absorbance caused by binding of the dye to proteins.
Within the range where calibration-curve linearity was obtained, the relationship between absorbance and protein concentration could be used.
However, differences in dye-binding properties among protein types and the effects of surfactants in the sample may prevent the measured value from accurately reflecting the concentration.
Discussion of the Lowry Method and BCA Assay
The Lowry method and BCA assay are quantification methods that develop color by using reactions such as reduction of copper ions by proteins.
These methods are highly sensitive and widely used, but may be affected by reducing agents, chelating agents, surfactants, buffer components, and similar substances.
If DTT, β-mercaptoethanol, EDTA, strong surfactants, or similar substances are present in the sample, they may affect the color reaction and cause the protein concentration to be overestimated or underestimated.
In a report, it is useful to include the components of the sample buffer in the discussion.
Example Discussion:
In the BCA assay, the concentration was determined using reduction of copper ions by proteins and the resulting color reaction.
However, if reducing agents or chelating agents are present in the sample, they may affect the color reaction and cause absorbance to change because of factors other than protein amount.
Therefore, the effect of the buffer components in the unknown sample on the quantification result must be considered.
Discussion of the UV Absorption Method
In the UV absorption method, protein concentration is determined using the fact that proteins absorb near 280 nm.
This absorption mainly originates from aromatic amino acids such as tryptophan and tyrosine.
Therefore, absorbance at 280 nm differs greatly depending on the type of protein.
In addition, nucleic acids such as DNA and RNA absorb near 260 nm, so contamination with nucleic acids may cause the protein concentration to be overestimated.
In the UV absorption method, it is important to consider the purity of the sample and the effects of contaminants.
Example Discussion:
In the UV absorption method, the concentration was determined using absorption near 280 nm originating from aromatic amino acids in proteins.
However, because the content of aromatic amino acids differs among proteins, absorbance may differ even at the same mass concentration.
In addition, contamination with nucleic acids affects UV absorption and may cause the protein concentration to be overestimated.
Effects of Interfering Substances in the Sample
In protein quantification, components in the sample may affect the color reaction or absorbance measurement.
Representative interfering substances include surfactants, reducing agents, chelating agents, salts, high-concentration buffers, organic solvents, and nucleic acids.
When these substances are present, absorbance may change independently of the amount of protein.
| Interfering Factor | Possible Effect | How to Write the Discussion |
|---|---|---|
| Surfactants | Affect dye binding or color development | Absorbance may not reflect the actual amount of protein |
| Reducing agents | Affect copper-ion reactions | Possibility of overestimation in the BCA or Lowry method |
| Chelating agents | Capture metal ions | Color reaction may become weaker |
| Nucleic acids | Affect UV absorption | Possibility of overestimation in 280 nm measurements |
| Turbidity or precipitate | Causes light scattering | Absorbance may appear high |
Example Discussion:
Buffer components or surfactants contained in the unknown sample may have affected protein quantification.
These components may interfere with the color reaction or absorbance measurement and change absorbance because of factors other than protein concentration.
Therefore, if the composition of the unknown sample differs from that of the standard sample, the concentration determined from the calibration curve may contain errors.
Errors Caused by Pipetting
In protein quantification, small volumes of standard solutions, unknown samples, and color reagents must be dispensed accurately.
Errors in micropipette operation cause deviations in concentration and reagent volume and affect absorbance.
Particularly when preparing a calibration curve, errors in the concentration series of standard solutions affect the entire approximation line.
Possible causes include liquid remaining at the tip, aspirating air bubbles, using the wrong plunger position, and difficulty accurately aspirating viscous samples.
In small-volume operations, even a small volume error can have a relatively large effect.
Example Discussion:
Errors in micropipette operation may have caused variation in the measurement points of the calibration curve.
If the dispensed volumes of standard solutions or color reagent differ, the actual protein concentration and reaction conditions change and affect absorbance.
Particularly in small-volume operations, even a slight volume error greatly affects concentration calculations, so accurate pipetting is important.
Errors Caused by Reaction Time
In color reactions used for protein quantification, reaction time may affect absorbance.
If the reaction time is too short, color development does not proceed sufficiently and absorbance becomes small.
Conversely, if the reaction time is too long, color development may proceed too far or stability may decrease, making comparisons between standard solutions and unknown samples inaccurate.
If the reaction times differ between standard solutions and unknown samples, they cannot be compared under the same conditions.
Therefore, it is important to standardize the time from the start of the reaction to measurement.
Example Discussion:
One possible cause of variation in absorbance is that the color-development reaction time was not consistent among samples.
In samples with a short reaction time, color development is insufficient and absorbance is measured as low.
On the other hand, when the reaction time becomes longer, color development may proceed further and absorbance may increase, so standard solutions and unknown samples must be measured after the same reaction time.
Errors Caused by Contamination of Cuvettes and Plates
In absorbance measurements, contamination, scratches, fingerprints, and bubbles on cuvettes or microplates affect the measured values.
If contamination or bubbles are present in the optical path, light transmission is hindered and absorbance may appear higher than the actual value.
Turbidity and precipitates in the sample also cause light scattering and affect absorbance.
Example Discussion:
Contamination on the cuvette surface or bubbles may explain why the absorbance was measured as high.
If contamination or bubbles are present in the optical path, light transmission is hindered and absorbance not originating from protein is added.
As a result, the protein concentration may have been overestimated.
Discussion When Absorbance Is Too High
If the absorbance of an unknown sample is too high, the protein concentration may exceed the calibration-curve range.
At high absorbance, the linearity of the measuring instrument decreases and the proportional relationship with concentration is more likely to break down.
Light scattering caused by sample turbidity or precipitation and color development caused by interfering substances may also increase absorbance.
Example Discussion:
If the absorbance of the unknown sample exceeded the upper limit of the calibration curve, the obtained concentration would be an estimate based on extrapolation and therefore have low reliability.
In the high-concentration range, the color reaction may become saturated and absorbance may cease to be proportional to protein concentration.
Therefore, the sample must be appropriately diluted and remeasured within the linear range of the calibration curve.
Discussion When Absorbance Is Too Low
If the absorbance of an unknown sample is too low, possible causes include a low protein concentration, excessive dilution of the sample, insufficient progression of the color reaction, and protein degradation or precipitation.
At low absorbance, the relative effects of blank correction and measurement noise become larger.
Example Discussion:
One possible reason the absorbance of the unknown sample was low is that the protein concentration in the sample was low.
Absorbance may also become small if the sample was excessively diluted or the color reaction did not proceed sufficiently.
Because the effects of the blank and measurement noise become larger in the low-absorbance range, the reliability of the concentration calculation may decrease.
Causes of Overestimating Protein Concentration
Causes of overestimating protein concentration include insufficient blank correction, contamination of the cuvette, sample turbidity, color development caused by interfering substances, errors in preparing standard-solution concentrations, and errors in the dilution factor.
In particular, if absorbance becomes high because of factors other than proteins, the concentration determined from the calibration curve also becomes high.
Example Discussion:
The protein concentration may have been overestimated because sample turbidity or contamination of the cuvette caused the absorbance to be measured as high.
In addition, if blank correction was insufficient and reagent-derived absorbance was not subtracted, the absorbance of the unknown sample would also be overestimated.
As a result, the protein concentration determined from the calibration curve is considered to have become higher than the actual value.
Causes of Underestimating Protein Concentration
Causes of underestimating protein concentration include protein degradation, precipitation, insufficient extraction, inhibition of the color reaction, excessive dilution of the sample, insufficient reaction time, and deterioration of the standard solution or reagent.
If proteins have precipitated, measuring only the supernatant gives a value lower than the actual total amount of protein.
Example Discussion:
Partial precipitation of proteins in the sample may have caused the protein concentration to be underestimated.
Precipitated proteins are not uniformly present in the measurement solution and therefore are not sufficiently reflected in absorbance.
In addition, if components in the sample inhibited the color reaction or the color-development time was short, absorbance would become low and cause the concentration to be underestimated.
Discussion of Variation in Repeated Measurements
In protein quantification, the same sample may be measured multiple times and the mean value determined.
If repeated measurements agree well, the reproducibility of the operation and measurement is considered high.
On the other hand, if variation is large, possible causes include pipetting, insufficient mixing, differences in reaction time, bubbles, and differences in plate position.
Example Discussion:
Variation was observed among repeated measurements of the same sample.
Possible causes include differences in dispensing volume caused by micropipetting, insufficient mixing of the reaction solution, bubbles during measurement, and differences in color-development reaction time.
Therefore, to improve reproducibility, dispensing operations and reaction times must be standardized and the presence or absence of bubbles should be checked before measurement.
Discussion When Calibration-Curve Points Deviate
Some standard points on a calibration curve may deviate greatly from the approximation line.
Possible causes include dilution errors in the standard solutions, pipetting errors, insufficient mixing of the color reagent, bubbles, and errors in reading absorbance.
If an outlier is excluded, the reason must be based on the experimental operation rather than simply because the value is inconvenient.
Example Discussion:
A dilution error during preparation of the standard solution may explain why some measurement points on the calibration curve deviated from the approximation line.
If the standard-solution concentration differs from the set value, its absorbance will deviate from the trend of the concentration series.
In addition, differences in dispensing volume and absorbance errors caused by bubbles may also contribute to variation in the calibration curve.
Discussion of the Calibration-Curve Intercept
The intercept of the calibration curve corresponds to the absorbance when the protein concentration is zero.
If blank correction is appropriate, the intercept is expected to be close to zero.
However, if there is reagent-derived absorbance, contamination of the cuvette, or insufficient blank correction, the intercept may deviate from zero.
Example Discussion:
Insufficient blank correction may explain why the intercept of the calibration curve deviated from zero.
If reagents without protein still show absorbance, failure to subtract that absorbance appropriately leaves absorbance even at a concentration of zero.
Therefore, the deviation of the intercept may reflect reagent-derived absorbance or background absorption of the measurement system.
Effects of Protein Denaturation and Degradation
Proteins may denature or degrade because of temperature, pH, enzymatic degradation, freeze-thaw cycles, organic solvents, and other factors.
When denaturation or degradation occurs, reactivity with the color reagent may change or proteins may precipitate and become difficult to measure.
As a result, the quantified value may differ from the actual amount of protein.
Example Discussion:
If the proteins denatured or degraded during sample storage, errors may have occurred in the quantified value.
If denaturation causes proteins to precipitate, the amount of soluble protein in the measurement solution decreases and the concentration may be underestimated.
In addition, if degradation changes reactivity with the color reagent, absorbance may no longer accurately reflect the original amount of protein.
Buffer Conditions of the Standard Curve and Unknown Sample
If the buffer conditions of the standard protein and unknown sample differ, differences may occur in the color reaction or absorbance.
For example, if only the unknown sample contains high concentrations of salts, surfactants, reducing agents, or chelating agents, it may not be possible to compare it under the same conditions as the standard curve.
For more accurate quantification, the standard protein may be dissolved in the same buffer as the unknown sample when preparing the calibration curve.
This makes it easier to standardize the effects of buffer components between the standard solutions and unknown sample.
Example Discussion:
If the buffer conditions differed between the standard protein and unknown sample, the effects on the color reaction may also have differed.
If surfactants or reducing agents in the unknown sample affected absorbance, the concentration determined from the standard curve would not be accurate.
Therefore, it is desirable to make the solution conditions of the standard solutions and unknown sample as similar as possible.
When the Results Can Be Considered Good
Protein quantification results can be considered good when the calibration curve shows good linearity, the absorbance of the unknown sample falls within the calibration-curve range, and variation among repeated measurements is small.
It is also important that blank correction be appropriate and that the sample dilution factor and buffer conditions be correctly taken into account.
Example Discussion:
The calibration curve showed good linearity, and the absorbance of the unknown sample also fell within the calibration-curve range.
In addition, variation among repeated measurements of the same sample was small and the reproducibility of the measurement was relatively good.
From these results, the protein concentration determined in this experiment is considered generally reasonable.
Example Discussion When the Experiment Did Not Go Well
If protein quantification does not go well, the causes are considered from results such as a non-linear calibration curve, an unknown sample outside the range, variation in repeated measurements, excessively high or low absorbance, or a large blank.
Organizing the causes into the calibration curve, sample, reagents, operation, and measuring instrument makes the discussion easier.
Example Discussion:
In this experiment, some measurement points on the calibration curve deviated from the straight line, creating uncertainty in calculation of the unknown-sample concentration.
Possible causes include dilution errors in the standard solutions, variation in pipetting, insufficient mixing of the color reagent, and differences in reaction time.
In addition, if buffer components in the unknown sample affected the color reaction, absorbance may not have accurately reflected the amount of protein.
How to Write Points for Improvement
In a discussion of protein quantification, including not only sources of error but also points for improvement makes the report easier to organize.
Improvements are easier to organize when divided into standard-solution preparation, pipetting, reaction conditions, measurement conditions, and sample pretreatment.
Improvements to Standard Solutions and the Calibration Curve
- Prepare standard protein concentrations accurately
- Prepare the dilution series carefully
- Mix each standard solution thoroughly
- Confirm the linear range of the calibration curve
- If an outlier appears, check the operation record
Improvements to Sample Measurement
- Dilute unknown samples so that they fall within the calibration-curve range
- Correctly reflect the dilution factor in the calculation
- Check the sample for turbidity or precipitates
- Perform centrifugation or filtration when necessary
- Match the buffer conditions of the standard solutions and unknown samples
Improvements to Operations and Measurements
- Use the micropipette accurately
- Avoid bubbles in the tip
- Keep the color-development reaction time consistent
- Perform blank correction appropriately
- Avoid contamination of cuvettes and plates
- Perform multiple measurements and use the mean value
Example of How to Write Points for Improvement:
To improve the accuracy of protein quantification, the dilution series of standard proteins must be prepared accurately and unknown samples must be measured within the linear range of the calibration curve.
In addition, it is important to standardize the color-development reaction time among samples and perform blank correction appropriately.
Furthermore, to reduce the effects of interfering substances in the unknown sample, it is desirable to make the buffer conditions of the standard solutions and unknown sample as similar as possible.
Difference Between a Superficial Discussion and a Good Discussion
In a discussion of protein quantification, simply writing that “the concentration was determined from the calibration curve” or “there was an error” results in a superficial discussion.
Relating calibration-curve linearity, the absorbance range of the unknown sample, blank correction, dilution factor, interfering substances, and operational errors produces a more persuasive discussion.
| Superficial Discussion | Good Discussion |
|---|---|
| The concentration was determined from the calibration curve. | Because a linear relationship was obtained between standard protein concentration and absorbance, the concentration was determined by substituting the absorbance of the unknown sample into the calibration curve. Because the absorbance of the unknown sample fell within the calibration-curve range, relatively reliable quantification without extrapolation is considered to have been possible. |
| The absorbance was high. | Possible reasons the absorbance of the unknown sample was high include not only a high protein concentration but also sample turbidity, insufficient blank correction, and color development caused by interfering substances. |
| There was an error. | Possible sources of error in the quantified value include errors in dilution of the standard solutions, variation in pipetting, differences in color-development reaction time, contamination of the cuvette, and the effects of surfactants or reducing agents in the unknown sample. |
Examples of Expressions That Can Be Used in Reports
The following expressions can be used when writing the results and discussion of protein quantification.
Adjust the necessary parts according to your own experimental results.
- A calibration curve was prepared from the standard protein concentrations and absorbance values.
- Because the calibration curve showed good linearity, protein concentration could be determined from absorbance within the measurement range.
- Because the absorbance of the unknown sample fell within the calibration-curve range, the concentration could be calculated without extrapolation.
- Because the unknown sample was diluted before measurement, the dilution factor was taken into account when determining the original sample concentration.
- Blank correction was used to subtract absorbance originating from the reagents and solvent.
- On the high-concentration side, the color reaction may become saturated and the proportional relationship between absorbance and concentration may break down.
- Surfactants or reducing agents in the sample may have affected the color reaction and caused errors in the quantified value.
- Contamination or bubbles in the cuvette can cause absorbance to appear high.
- If reaction time differs among samples, the extent of color development changes and variation in absorbance occurs.
- If the properties of the standard protein and unknown sample differ, color intensity may differ even at the same concentration.
Points to Check When Discussing Protein Quantification
Checking the following points before writing the report makes the discussion easier to write.
- Have you stated the quantification method used?
- Have you written the type of standard protein?
- Have you shown the calibration-curve equation and R2?
- Have you confirmed the linear range of the calibration curve?
- Have you confirmed that the absorbance of the unknown sample lies within the calibration-curve range?
- Have you performed blank correction?
- Have you correctly reflected the dilution factor in the calculation?
- Have you discussed variation among repeated measurements?
- Have you considered errors caused by pipetting and reaction time?
- Have you considered the effects of interfering substances?
- Have you considered differences between the standard protein and the unknown sample?
- Do the points for improvement correspond to the sources of error?
Summary
In protein quantification, a calibration curve is prepared using a standard protein, and the protein concentration is determined from the absorbance of an unknown sample.
Good linearity of the calibration curve and an unknown-sample absorbance that falls within the calibration-curve range are important for reliable quantification.
If the unknown sample is diluted before measurement, the dilution factor must always be taken into account when determining the original sample concentration.
Major sources of error in protein quantification include errors in preparing standard solutions, pipetting, insufficient blank correction, differences in color-development reaction time, contamination of cuvettes or plates, interfering substances in the sample, and protein denaturation or precipitation.
In addition, if the properties of the standard protein and unknown sample differ, the intensity of color development may differ even at the same concentration.
In a report, rather than simply writing that “the concentration was determined from the calibration curve,” discuss calibration-curve linearity, measurement range, blank correction, dilution factor, interfering substances, and operational errors in relation to one another.
Explaining which factors may cause the concentration to be estimated too high or too low produces a more persuasive discussion of protein quantification.
