Protein quantification is the process of determining the concentration or total amount of protein contained in a sample. In biochemical experiments, it is required in many situations, including comparisons of enzyme activity, adjustment of sample amounts for electrophoresis, evaluation of purification procedures, and comparison of cell extracts.
Representative methods include the Bradford, Lowry, BCA, and ultraviolet absorption methods. In educational experiments, the Bradford method is often used because it is relatively simple and can be completed in a short time.
In the Bradford method, binding of the dye to protein changes its absorption characteristics and increases absorbance at a specified wavelength. A standard curve is prepared from protein solutions of known concentration, and the concentration of an unknown sample is determined by applying its absorbance to the standard curve.
However, the range over which absorbance and protein concentration are linearly related is limited. Color development and absorbance can also be affected by protein type, detergents, salts, reducing agents, turbidity, and other components in the sample.
This article assumes a colorimetric assay using educational standard proteins and confirmed safe samples, and summarizes the introduction, precautions, result items, sample experimental values, calculations, sources of error, example results, and example discussion.
Note:
This article is a study reference for preparing biology and biochemistry laboratory reports. The conditions, values, standard curves, and results shown are illustrative examples and are not actual measurements.
Use educational standard proteins and confirmed safe samples. Do not independently handle human-derived samples, clinical samples, or unidentified biological samples.
Color reagents may contain acidic components, organic dyes, metal ions, or other substances. Avoid contact with skin and eyes and use protective eyewear, gloves, lab coats, and other protective equipment specified by the facility.
Follow product instructions or the instructor’s directions for reagents, reaction times, measurement wavelengths, and waste disposal.
Do not touch the optical surfaces of spectrophotometers or microplate readers, and check cuvettes or plates for bubbles, dirt, and scratches.
- Basics of Protein Quantification
- Main Items to Record in the Results
- Sample Experimental Conditions
- Sample Experimental Values
- Calculation Methods
- Blank Correction
- Calculating Diluted-Sample Concentration from the Standard Curve
- Original-Sample Concentration
- Dilution Factor
- Total Protein Amount
- Protein Amount per Gram of Sample
- Recovery Rate
- Mean Concentration
- Standard Deviation
- Coefficient of Variation
- Relative Error Against a Theoretical Value
- Interpreting the Coefficient of Determination
- Main Sources of Error
- Example Results
- Example Discussion
- Example Discussion When Absorbance Exceeds the Upper Limit of the Standard Curve
- Example Discussion When Absorbance Is Close to the Blank
- Example Discussion When the Standard Curve Is Not Linear
- Example Discussion When Replicate Measurements Vary Widely
- Example Discussion When the Measured Concentration Is Higher Than the Theoretical Value
- Summary
Basics of Protein Quantification
Standards
These are protein solutions of known concentration. Bovine serum albumin is commonly used as a standard protein. Multiple standard concentrations are measured to determine the relationship between concentration and absorbance.
Blank
A measurement solution containing no protein and consisting only of sample solvent and color reagent. It is used to correct for absorbance originating from the reagent and solvent themselves.
Standard Curve
A graph with protein concentration on the horizontal axis and blank-corrected absorbance on the vertical axis. When a linear regression equation is used, the protein concentration of an unknown sample can be calculated from its absorbance.
Dilution
If the absorbance of an unknown sample exceeds the range of the standard curve, dilute the sample and measure it again. Multiply the calculated concentration by the dilution factor to obtain the concentration of the original sample.
Total Protein Amount
Multiplying protein concentration by sample volume gives the total amount of protein in the sample. In purification and extraction procedures, comparison of total amount as well as concentration is important.
Main Items to Record in the Results
- Name of quantification method
- Type of standard protein
- Standard concentrations
- Volume of standard added
- Unknown sample name
- Dilution factor of unknown sample
- Volume of color reagent
- Total reaction volume
- Reaction time
- Reaction temperature
- Measurement wavelength
- Blank absorbance
- Standard absorbances
- Unknown-sample absorbances
- Blank-corrected absorbance
- Regression equation of the standard curve
- Coefficient of determination R²
- Linear range of the standard curve
- Protein concentration of diluted sample
- Protein concentration of original sample
- Sample volume
- Total protein amount
- Protein amount per gram of sample
- Recovery rate
- Replicate measurements
- Mean
- Standard deviation
- Coefficient of variation
- Presence of outliers
- Presence of samples outside the measurement range
Sample Experimental Conditions
| Item | Sample condition |
|---|---|
| Quantification method | Bradford method |
| Standard protein | Educational albumin standard |
| Standard concentration range | 0–1.0 mg/mL |
| Measurement wavelength | 595 nm |
| Reaction time | 10 min |
| Reaction temperature | Room temperature |
| Unknown sample | Confirmed safe educational protein sample |
| Unknown-sample dilution | 5-fold |
| Replicates | Three per sample |
Sample Experimental Values
Absorbance of Standards
| Standard concentration (mg/mL) | Absorbance 1 | Absorbance 2 | Absorbance 3 | Mean absorbance |
|---|---|---|---|---|
| 0.00 | 0.041 | 0.039 | 0.040 | 0.040 |
| 0.20 | 0.164 | 0.168 | 0.166 | 0.166 |
| 0.40 | 0.291 | 0.295 | 0.293 | 0.293 |
| 0.60 | 0.420 | 0.424 | 0.422 | 0.422 |
| 0.80 | 0.548 | 0.552 | 0.550 | 0.550 |
| 1.00 | 0.674 | 0.680 | 0.677 | 0.677 |
Blank-Corrected Standard Values
| Standard concentration (mg/mL) | Mean absorbance | Blank-corrected absorbance |
|---|---|---|
| 0.00 | 0.040 | 0.000 |
| 0.20 | 0.166 | 0.126 |
| 0.40 | 0.293 | 0.253 |
| 0.60 | 0.422 | 0.382 |
| 0.80 | 0.550 | 0.510 |
| 1.00 | 0.677 | 0.637 |
Sample Standard-Curve Values
| Item | Sample value |
|---|---|
| Regression equation | y = 0.638x − 0.001 |
| x | Protein concentration (mg/mL) |
| y | Blank-corrected absorbance |
| Coefficient of determination | R² = 0.9999 |
| Linear range | 0–1.0 mg/mL |
Measurements of Unknown Samples
| Sample | Dilution factor | Absorbance 1 | Absorbance 2 | Absorbance 3 | Mean absorbance |
|---|---|---|---|---|---|
| Sample A | 5-fold | 0.354 | 0.360 | 0.357 | 0.357 |
| Sample B | 10-fold | 0.482 | 0.490 | 0.486 | 0.486 |
| Sample C | 2-fold | 0.151 | 0.155 | 0.153 | 0.153 |
| Sample D | 5-fold | 0.812 | 0.826 | 0.819 | 0.819 |
Calculated Results for Unknown Samples
| Sample | Corrected absorbance | Diluted-sample concentration (mg/mL) | Original-sample concentration (mg/mL) | Assessment |
|---|---|---|---|---|
| Sample A | 0.317 | 0.498 | 2.49 | Within measurement range |
| Sample B | 0.446 | 0.701 | 7.01 | Within measurement range |
| Sample C | 0.113 | 0.179 | 0.358 | Low concentration |
| Sample D | 0.779 | 1.223 | Not suitable for calculation | Outside standard-curve range |
Sample Total Protein Amounts
| Sample | Original concentration (mg/mL) | Sample volume (mL) | Total protein (mg) |
|---|---|---|---|
| Sample A | 2.49 | 4.0 | 9.96 |
| Sample B | 7.01 | 2.5 | 17.53 |
| Sample C | 0.358 | 10.0 | 3.58 |
Sample Recovery Before and After Purification
| Stage | Concentration (mg/mL) | Volume (mL) | Total amount (mg) | Recovery (%) |
|---|---|---|---|---|
| Crude extract | 3.20 | 10.0 | 32.0 | 100 |
| After filtration | 2.85 | 9.0 | 25.65 | 80.2 |
| After purification | 4.50 | 4.0 | 18.0 | 56.3 |
Sample Replicate Measurements
| Replicate | Original-sample concentration (mg/mL) |
|---|---|
| 1 | 2.44 |
| 2 | 2.49 |
| 3 | 2.54 |
| Mean | 2.49 |
Sample Change with Reaction Time
| Reaction time | Absorbance | Observation |
|---|---|---|
| 2 min | 0.302 | Insufficient color development |
| 5 min | 0.338 | Nearly stable |
| 10 min | 0.357 | Good |
| 20 min | 0.365 | Slight increase |
| 40 min | 0.381 | Large effect of timing differences |
Calculation Methods
Blank Correction
If the mean absorbance of the unknown sample is 0.357 and the blank absorbance is 0.040:
Corrected absorbance = sample absorbance − blank absorbance
= 0.357 − 0.040
= 0.317
Calculating Diluted-Sample Concentration from the Standard Curve
If the standard curve is y = 0.638x − 0.001 and corrected absorbance y is 0.317:
0.317 = 0.638x − 0.001
x = (0.317 + 0.001) ÷ 0.638
≈ 0.498 mg/mL
Original-Sample Concentration
If the sample was diluted fivefold:
Original concentration = diluted-sample concentration × dilution factor
= 0.498 mg/mL × 5
= 2.49 mg/mL
Dilution Factor
If 4.0 mL of diluent is added to 1.0 mL of sample, the final volume is 5.0 mL.
Dilution factor = final diluted volume ÷ original sample volume
= 5.0 mL ÷ 1.0 mL
= 5-fold
Total Protein Amount
If the protein concentration is 2.49 mg/mL and the sample volume is 4.0 mL:
Total protein = protein concentration × sample volume
= 2.49 mg/mL × 4.0 mL
= 9.96 mg
Protein Amount per Gram of Sample
If 9.96 mg total protein is obtained from 5.0 g of sample:
Protein amount (mg/g) = total protein ÷ sample mass
= 9.96 mg ÷ 5.0 g
= 1.99 mg/g
Recovery Rate
If total protein is 32.0 mg in the crude extract and 18.0 mg after purification:
Recovery (%) = total amount after purification ÷ total amount before purification × 100
= 18.0 ÷ 32.0 × 100
= 56.3%
Mean Concentration
Mean concentration = (2.44 + 2.49 + 2.54) ÷ 3
= 2.49 mg/mL
Standard Deviation
s = √{Σ(xi − x̄)2/(n − 1)}
= 0.05 mg/mL
Coefficient of Variation
Coefficient of variation (%) = standard deviation ÷ mean × 100
= 0.05 ÷ 2.49 × 100
≈ 2.0%
Relative Error Against a Theoretical Value
If the theoretical concentration is 2.60 mg/mL and the measured concentration is 2.49 mg/mL:
Relative error (%) = |measured − theoretical| ÷ theoretical × 100
= |2.49 − 2.60| ÷ 2.60 × 100
≈ 4.2%
Interpreting the Coefficient of Determination
The sample standard curve had R² = 0.9999. A value closer to 1 indicates that standard concentration and absorbance fit the regression line well. However, even a high R² does not mean that unknown samples outside the measurement range can safely be extrapolated.
Main Sources of Error
| Source of error | Effect on results | Improvement |
|---|---|---|
| Error in standard preparation | Shifts the entire standard curve | Use accurate pipettes and equipment |
| Insufficient mixing of standards | Concentration is not uniform | Mix thoroughly without foaming |
| No blank measurement | Includes reagent-derived absorbance | Prepare a blank using the same solvent as the sample |
| Different blank composition | Correction becomes inaccurate | Match the solvent composition of unknown samples |
| Variation in reagent volume | Changes color intensity | Add a constant volume with the same pipette |
| Different addition order | Reaction times differ | Use a consistent order and interval |
| Insufficient reaction time | Underdeveloped color underestimates concentration | Measure after the specified time |
| Variation in reaction time | Makes sample comparisons inaccurate | Standardize with a timer |
| Different reaction temperatures | Changes color-development rate | React under the same temperature conditions |
| Absorbance outside the standard-curve range | Increases concentration-estimation error | Dilute and remeasure the sample |
| Extrapolation beyond the standard curve | Over- or underestimates concentration | Measure within the standard range |
| Dilution-factor calculation error | Produces an incorrect original concentration | Keep dilution records |
| Inappropriate pipette range | Increases relative error | Use a pipette suited to the volume |
| Residual liquid in tip | Reduces delivered volume | Dispense consistently |
| Bubbles | Light scattering increases absorbance | Remove bubbles before measurement |
| Dirty cuvette or plate | Increases absorbance | Keep optical surfaces clean |
| Different cuvette orientation | Changes effects of path length and scratches | Measure in the same orientation |
| Different liquid volumes among wells | Changes optical path length | Standardize total reaction volume |
| Sample turbidity | Light scattering overestimates concentration | Consider centrifugation, filtration, or an appropriate blank |
| Precipitate contamination | Makes measurements unstable | Use a uniform supernatant |
| Detergent contamination | May inhibit or enhance color development | Check reagent compatibility |
| Reducing-agent contamination | Interferes with color development in some assays | Select a compatible assay |
| High salt concentration | Affects color development or absorbance | Match solvent conditions between standards and samples |
| Different protein types | Color response differs between standard and unknown | Use a standard with similar properties |
| Protein denaturation or precipitation | Changes the amount of protein available for reaction | Control storage conditions and pH |
| Deteriorated standard | Actual concentration differs from the set value | Store appropriately and replace when needed |
| Incorrect wavelength setting | Absorbance is measured incorrectly | Confirm the wavelength specified by the method |
| Poor instrument zeroing | Systematically shifts all values | Calibrate with the blank |
| Too few replicates | Makes random error difficult to assess | Perform multiple measurements |
| Unconditional exclusion of outliers | Arbitrarily alters results | State the cause and exclusion criteria |
Example Results
Educational protein samples were quantified using the Bradford method. Standard protein concentrations ranged from 0 to 1.0 mg/mL, and absorbance was measured at 595 nm.
The mean blank absorbance was 0.040. Blank-corrected absorbances for 0.20, 0.40, 0.60, 0.80, and 1.00 mg/mL standards were 0.126, 0.253, 0.382, 0.510, and 0.637, respectively.
The regression equation of the standard curve was y = 0.638x − 0.001, with R² = 0.9999.
For fivefold-diluted sample A, mean absorbance was 0.357 and blank-corrected absorbance was 0.317. The standard curve gave a diluted-sample concentration of 0.498 mg/mL, corresponding to an original-sample concentration of 2.49 mg/mL.
Because sample A had a volume of 4.0 mL, its total protein amount was 9.96 mg.
Three independent measurements of sample A were 2.44, 2.49, and 2.54 mg/mL, giving a mean of 2.49 mg/mL, a standard deviation of 0.05 mg/mL, and a coefficient of variation of 2.0%.
The corrected absorbance of sample D was 0.779, exceeding the maximum standard value of 0.637. Its concentration was therefore not finalized, and further dilution and remeasurement were considered necessary.
Example Discussion
In this experiment, a standard curve was prepared from standard protein solutions using the Bradford method, and the protein concentrations of unknown samples were determined.
Within the standard protein concentration range of 0–1.0 mg/mL, absorbance at 595 nm increased with increasing concentration.
This is considered to result from increased dye binding to protein, which increases absorption at the measurement wavelength.
The blank absorbance was 0.040. Because the color reagent and solvent themselves absorb slightly, the blank value must be subtracted from sample absorbance.
Without blank correction, all sample concentrations would be overestimated.
The regression equation was y = 0.638x − 0.001 and R² was 0.9999. Because R² was very close to 1, concentration and absorbance showed a good linear relationship over the standard range used.
However, a high R² alone does not mean that every concentration can be quantified accurately. At high concentrations, the amount of dye and the instrument’s measurement range may become limiting, so absorbance may no longer increase proportionally with concentration.
For fivefold-diluted sample A, mean absorbance was 0.357 and corrected absorbance was 0.317. Substitution into the standard curve gave 0.498 mg/mL for the diluted sample, and multiplying by the dilution factor of 5 gave 2.49 mg/mL for the original sample.
Because sample A had a volume of 4.0 mL, its total protein amount was 9.96 mg. Concentration indicates how concentrated a sample is, but total protein requires multiplication by sample volume.
Sample B had a higher concentration than sample A at 7.01 mg/mL and a higher total amount of 17.53 mg, indicating that it contained more protein in the specified sample volume.
The original concentration of sample C was 0.358 mg/mL, lower than the other samples. In low-concentration samples with absorbance close to the blank, even a small measurement error can greatly affect the calculated concentration.
Therefore, accurate measurement at low concentrations may require adding standards at the low end, reducing sample dilution, or using a more sensitive assay.
Sample D had a corrected absorbance of 0.779, exceeding 0.637 for the 1.0 mg/mL standard. Although extrapolation gives 1.223 mg/mL for the diluted sample, this lies outside the range in which linearity is guaranteed and should not be adopted as a final value.
Sample D should be further diluted and remeasured so that its absorbance falls within the standard-curve range.
In the reaction-time comparison, absorbance was 0.302 after 2 minutes, 0.357 after 10 minutes, and 0.381 after 40 minutes. Color development was incomplete early in the reaction, and absorbance increased with time.
If reaction times differ among samples, identical protein concentrations may produce different absorbances. The time from reagent addition to measurement must therefore be standardized.
The three concentrations for sample A were 2.44, 2.49, and 2.54 mg/mL, with a mean of 2.49 mg/mL, standard deviation of 0.05 mg/mL, and coefficient of variation of 2.0%.
The relatively small coefficient of variation suggests reasonable reproducibility in pipetting, color development, and absorbance measurement.
However, even with a small standard deviation, a systematic error in the standard concentrations would shift all measurements in the same direction. Standard preparation and instrument calibration are therefore important in addition to replicate measurements.
Before and after purification, total protein decreased from 32.0 mg in the crude extract to 18.0 mg after purification, corresponding to a recovery of 56.3%.
Although concentration increased from 3.20 to 4.50 mg/mL after purification, volume decreased from 10.0 to 4.0 mL, so total protein decreased.
This shows that an increase in concentration alone does not indicate increased recovery. Purification should be evaluated using concentration, volume, total amount, and recovery together.
Possible causes of recovery below 100% include losses during filtration or centrifugation, adsorption to container surfaces, retention in precipitates, unrecovered liquid, and removal during purification.
However, crude extracts contain many proteins other than the target protein, so a decrease in total protein alone does not indicate purification failure. Target-protein activity or purity must be evaluated separately.
In the Bradford method, dye binding varies among proteins. If the amino-acid composition or conformation of the standard differs greatly from that of the unknown protein, equal mass concentrations may produce different color intensities.
Thus, concentrations obtained from the standard curve are values relative to the standard protein and may not exactly equal the absolute concentration of an unknown protein.
Detergents, salts, reducing agents, and buffer components may also interfere with color development. If solvent composition differs between standards and unknowns, differences unrelated to protein concentration may appear in absorbance.
To reduce this effect, solvent conditions should be matched as closely as possible and a blank using the same solvent as the sample should be prepared.
Bubbles and turbidity can also increase absorbance through light scattering rather than protein absorption. Bubbles should be removed and centrifugation or filtration considered when necessary.
Overall, by appropriately diluting unknown samples to fall within the linear range, applying blank correction, and correcting for dilution, sample A was determined to have a protein concentration of 2.49 mg/mL and total protein of 9.96 mg.
To improve reliability, standard preparation, reaction time, measurement wavelength, sample dilution, blank composition, and replicate measurements should be standardized, and values outside the standard-curve range should not be forced into an extrapolated estimate.
Example Discussion When Absorbance Exceeds the Upper Limit of the Standard Curve
If the absorbance of an unknown sample exceeds that of the highest-concentration standard, it lies outside the standard-curve range and linearity is not guaranteed. Do not adopt an extrapolated value; dilute the sample further and remeasure it within the standard range.
Example Discussion When Absorbance Is Close to the Blank
If sample absorbance is close to the blank, possible causes include low protein concentration, excessive sample dilution, insufficient color development, or an error in reagent addition. Add lower-concentration standards and review the dilution factor and reaction time.
Example Discussion When the Standard Curve Is Not Linear
Possible causes of a nonlinear standard curve include errors in standard preparation, variation in reagent volume, inconsistent reaction times, bubbles, and exceeding the measurement range. Prepare the standards again, standardize each operating condition, and repeat the measurements.
Example Discussion When Replicate Measurements Vary Widely
Large variation among replicates may result from pipetting, insufficient mixing, bubbles, differences in color-development time, or differences in liquid volume among wells. Perform each step in the same order and timing and use a pipette appropriate for the volume.
Example Discussion When the Measured Concentration Is Higher Than the Theoretical Value
Possible causes include incorrect blank correction, sample turbidity, bubbles, dilution-factor errors, or contaminants that enhance color development. Check the appearance of the measurement solution and the calculation process.
Summary
Protein quantification uses standards of known concentration to prepare a standard curve and determines unknown-sample concentration from absorbance.
Subtract blank absorbance from sample absorbance, substitute the corrected value into the regression equation, and multiply by the dilution factor to obtain the original-sample concentration.
Total protein is the product of concentration and sample volume; in purification procedures, total amount and recovery are important in addition to concentration.
Absorbance values outside the standard-curve range should not be forcibly extrapolated; dilute the sample appropriately and measure again.
Accurate quantification requires control of standard preparation, blanks, reaction time, measurement wavelength, dilution factor, solvent composition, bubbles and turbidity, and replicate measurements.