pH measurement is a basic chemistry experiment performed to investigate acid–base properties and the condition of solutions. There are also simple methods using pH test paper, but in university chemistry experiments, pH meters are often used for more accurate measurements.
A pH meter is a convenient measuring instrument, but its readings may shift depending on calibration, electrode condition, temperature, the properties of the sample solution, and how the instrument is handled during measurement. Therefore, in reports on pH measurement, it is important to discuss not only the measured values but also whether the values stabilized, whether they agreed with theoretical or expected values, and what possible causes of error can be considered.
This article explains how to interpret pH measurement results, causes of pH meter errors, reasons why measured values may not stabilize, and discussion examples that can be used in reports.
Note: This article is a reference for discussing results obtained in chemistry experiments at universities and similar institutions. For actual measurement procedures and safety precautions, always follow the instructions in your university’s laboratory manual and those given by your instructor or teaching assistant (TA).
- What Is pH?
- Results to Examine in pH Measurement
- Reference Experimental Values and Examples of Calibration and Error Analysis for pH Measurement
- Reference Experimental Conditions
- Basic pH Equation
- Example of pH Meter Calibration
- Examples of Measurements of Acidic, Neutral, and Basic Samples
- Example of Calculating the pH of a Strong Acid
- Example of Calculating the pH of a Strong Base
- Example Comparing a Weak Acid and a Strong Acid
- Example of pH Changes Caused by Dilution
- Example of pH Changes in a Buffer Solution
- Example of Changes in Measured pH with Temperature
- Example of Electrode Response Time
- Example of Reproducibility with Multiple Measurements
- Example Comparing pH Test Paper and a pH Meter
- Example of Measurement of Low-Ionic-Strength Samples
- Example of pH Decrease Due to CO2 Absorption
- Example of Systematic Error Due to Insufficient Calibration
- Main Error Factors
- Example of Writing the Results
- Points to Connect to the Discussion
- Example Discussion
- Summary
- Understanding the Basic Mechanism of a pH Meter
- Main Causes of Error in pH Meters
- Error Due to Insufficient Calibration
- Error Due to Electrode Contamination
- Error Due to Insufficient Electrode Rinsing
- Effect of Temperature on pH Measurement
- Reasons Why Measured Values Do Not Stabilize
- Error Due to Insufficient Mixing of the Solution
- Effect of Carbon Dioxide in the Air
- Points to Consider When Measuring the pH of Buffer Solutions
- Points to Consider When Measuring the pH of Strong Acids and Strong Bases
- Points to Consider When Measuring the pH of Weak Acids and Weak Bases
- Difference from pH Test Paper
- Reasons Why Theoretical and Measured Values Differ
- Discussion When Multiple Measurements Are Performed
- When the Results Can Be Considered Good
- Discussion Example When the Measurement Did Not Go Well
- How to Write Improvements
- Difference Between a Superficial Discussion and a Good Discussion
- Example Expressions That Can Be Used in Reports
- Points to Check When Discussing pH Measurement
- Summary
What Is pH?
pH is a value that indicates the degree of acidity, neutrality, or basicity of a solution. In general, a pH below 7 is considered acidic, a pH around 7 neutral, and a pH above 7 basic. However, the exact neutral point may vary depending on temperature and the type of solution.
In chemistry experiments, pH is important in many situations, including acid–base reactions, buffer solutions, hydrolysis, precipitation, and enzyme reactions. Even a small change in pH can alter the progress of a reaction or the state of the products.
Therefore, in pH measurement, it is important not only to read the numerical value but also to consider how that value relates to the purpose and expectations of the experiment.
Results to Examine in pH Measurement
For pH measurement results, organize the measured pH value, measurement conditions, temperature, stability of the value, and comparison with theoretical or expected values. Particularly when using a pH meter, it is important to determine whether the value was recorded after the measurement had stabilized.
Main Items to Include in the Results
- Measured pH
- Type of solution measured
- Temperature during measurement
- Standard solutions used to calibrate the pH meter
- Whether the measured value stabilized
- Average value when multiple measurements were performed
- Comparison with theoretical or expected values
- Changes in the solution before and after measurement
Example of Writing the Results:
When the pH of an aqueous acetic acid solution was measured using a pH meter, the pH was 3.10. The measured value became almost stable after several tens of seconds, although slight fluctuations between 3.05 and 3.15 were observed immediately after measurement began. Because an aqueous solution of acetic acid, which is a weak acid, is expected to be acidic, the obtained value generally agreed with expectations.
Reference Experimental Values and Examples of Calibration and Error Analysis for pH Measurement
This section organizes reference experimental values for pH measurements using a pH meter or pH test paper in order to discuss calibration using standard solutions, measured values of acidic, neutral, and basic samples, the effects of temperature, electrode response, buffer solutions, and changes in pH caused by dilution. Although pH measurement appears simple, the value changes depending on electrode condition, calibration, temperature, and sample properties, so it is important to discuss the results together with the measurement conditions.
pH is a value that serves as an indicator of hydrogen ion concentration and is a basic measure used to determine acidity, neutrality, and basicity. However, pH does not simply represent concentration itself; it is related to the activity of hydrogen ions in solution. Therefore, it is necessary to consider not only concentration calculations but also the effects of instrument calibration and sample conditions.
Reference Experimental Conditions
| Item | Details |
|---|---|
| Samples measured | Aqueous acid solutions, aqueous base solutions, buffer solutions, tap water, food samples, environmental water, etc. |
| Measurement methods | pH meter, glass electrode, pH test paper, pH indicators, etc. |
| Examples of calibration solutions | pH 4.01, pH 6.86 or 7.00, pH 9.18 or 10.01 |
| Measurement temperature | 25°C is often used as the reference temperature |
| Evaluation items | pH, reproducibility, difference before and after calibration, temperature effects, dilution effects, electrode response, error factors |
| Main error factors | Insufficient calibration, electrode contamination, temperature changes, insufficient stirring, low ionic strength, CO2 absorption by the sample, timing of readings |
Basic pH Equation
pH is expressed as an indicator of hydrogen ion concentration by the following equation.
pH = −log[H+]
| [H+] | pH | Solution Property | Interpretation |
|---|---|---|---|
| 1.0×10−2 mol/L | 2.00 | Acidic | High hydrogen ion concentration |
| 1.0×10−5 mol/L | 5.00 | Weakly acidic | Acidic, but not strongly acidic |
| 1.0×10−7 mol/L | 7.00 | Near neutral | Approximate value for pure water at 25°C |
| 1.0×10−10 mol/L | 10.00 | Basic | Low hydrogen ion concentration |
When pH changes by 1, the hydrogen ion concentration changes by a factor of 10. Therefore, even a small difference in pH represents a large difference in the strength of acidity or basicity of a solution.
Example of pH Meter Calibration
A pH meter is calibrated with standard buffer solutions before measurement. Calibration corrects shifts in electrode slope and zero point and improves the reliability of measured values.
| Calibration Solution | Standard pH | Reading Before Calibration | Reading After Calibration | Evaluation |
|---|---|---|---|---|
| Acidic standard solution | 4.01 | 4.13 | 4.01 | Good correction |
| Neutral standard solution | 6.86 | 6.94 | 6.86 | Good correction |
| Basic standard solution | 9.18 | 9.04 | 9.18 | Good correction |
Before calibration, the readings differed from the standard values by approximately 0.08–0.14, but after calibration they agreed with the standard values. If measurements are performed without calibration, the pH values of all samples may be systematically shifted.
Examples of Measurements of Acidic, Neutral, and Basic Samples
| Sample | Measured pH | Solution Property | Observation / Characteristic | Direction of Discussion |
|---|---|---|---|---|
| 0.010 mol/L HCl | 2.02 | Acidic | Close to theoretical value of 2.00 | Almost complete dissociation as a strong acid |
| Aqueous acetic acid solution | 3.18 | Acidic | Higher pH than a strong acid of the same concentration | Partial dissociation of a weak acid |
| Tap water | 7.35 | Near neutral | Slightly above neutral | Effect of dissolved components |
| Carbonated water | 4.05 | Acidic | Acidity derived from CO2 | Effect of carbonic acid |
| 0.010 mol/L NaOH | 12.02 | Basic | Close to theoretical value of 12.00 | Almost complete dissociation as a strong base |
| Aqueous ammonia solution | 10.80 | Basic | Lower pH than a strong base of the same concentration | Partial dissociation of a weak base |
Strong acids and strong bases give pH values close to their theoretical values, whereas weak acids and weak bases are not completely dissociated, so their pH values differ greatly even at the same concentration.
Example of Calculating the pH of a Strong Acid
0.010 mol/L HCl is a strong acid and is considered to dissociate almost completely in water.
HCl → H+ + Cl−
Therefore, [H+] can be regarded as 0.010 mol/L = 1.0×10−2 mol/L.
pH = −log(1.0×10−2) = 2.00
If the measured value is 2.02, it is close to the theoretical value of 2.00, suggesting that the calibration and measurement conditions were generally satisfactory.
Example of Calculating the pH of a Strong Base
0.010 mol/L NaOH is a strong base, and [OH−] is considered to be 1.0×10−2 mol/L.
pOH = −log[OH−] = −log(1.0×10−2) = 2.00
At 25°C, pH + pOH can be approximated as 14.00, so
pH = 14.00 − 2.00 = 12.00
If the measured value is 12.02, it is considered to agree well with the theoretical value.
Example Comparing a Weak Acid and a Strong Acid
| Sample | Concentration | Measured pH | Discussion |
|---|---|---|---|
| HCl | 0.010 mol/L | 2.02 | Almost completely dissociated because it is a strong acid |
| Acetic acid | 0.010 mol/L | 3.38 | Produces less H+ because it is a weak acid |
| Carbonated water | Apparent concentration depends on conditions | 4.05 | Changes depending on the amount of dissolved CO2 |
For a weak acid, the total acid concentration does not equal the hydrogen ion concentration, so the pH is higher than that of a strong acid.
Example of pH Changes Caused by Dilution
When a strong acid is diluted by a factor of 10 each time, the pH ideally increases by approximately 1.
| HCl Concentration | Theoretical pH | Measured pH | Difference | Direction of Discussion |
|---|---|---|---|---|
| 1.0×10−1 mol/L | 1.00 | 1.03 | +0.03 | Close to theoretical value |
| 1.0×10−2 mol/L | 2.00 | 2.02 | +0.02 | Good |
| 1.0×10−3 mol/L | 3.00 | 3.05 | +0.05 | Slight deviation |
| 1.0×10−4 mol/L | 4.00 | 4.12 | +0.12 | Error becomes noticeable in dilute solutions |
| 1.0×10−5 mol/L | 5.00 | 5.34 | +0.34 | Effect of pure water and CO2 |
As the concentration decreases, the effects of the self-ionization of pure water, dissolution of CO2 from the air, and electrode response become relatively greater.
Example of pH Changes in a Buffer Solution
A buffer solution is a solution in which the pH does not change greatly when a small amount of acid or base is added. Reference data for an acetic acid/sodium acetate buffer solution are shown below.
| Condition | Measured pH | pH Change | Direction of Discussion |
|---|---|---|---|
| Buffer solution only | 4.76 | − | Near pKa |
| Addition of a small amount of HCl | 4.68 | −0.08 | Acetate ions accept H+ |
| Addition of a small amount of NaOH | 4.84 | +0.08 | Acetic acid neutralizes OH− |
| 10-fold dilution | 4.78 | +0.02 | If the ratio remains nearly the same, the pH does not change greatly |
In a buffer solution, an acid and its conjugate base coexist, so the pH tends to remain stable against the addition of small amounts of acid or base and against dilution.
Example of Changes in Measured pH with Temperature
| Sample | 15°C | 25°C | 35°C | Direction of Discussion |
|---|---|---|---|---|
| Standard solution near pH 4 | 4.00 | 4.01 | 4.03 | Change in standard value with temperature |
| Standard solution near pH 7 | 6.90 | 6.86 | 6.84 | The neutral point changes with temperature |
| Tap water | 7.42 | 7.35 | 7.30 | Effect of dissolved components and temperature |
| Acetate buffer solution | 4.78 | 4.76 | 4.73 | Temperature dependence of the equilibrium constant |
If values measured at different temperatures are compared directly, the comparison may include effects caused by temperature differences rather than differences in the properties of the samples.
Example of Electrode Response Time
| Measurement Time | Displayed pH | Condition | Evaluation |
|---|---|---|---|
| Immediately after immersion | 7.62 | Value fluctuating | Too early to record |
| After 10 seconds | 7.45 | Still changing | Still unstable |
| After 30 seconds | 7.36 | Almost stable | Can be recorded |
| After 60 seconds | 7.35 | Stable | Representative value |
If a value is recorded before it stabilizes, the recorded pH may be higher or lower than the actual value.
Example of Reproducibility with Multiple Measurements
| Measurement | Tap Water pH | Acetate Buffer pH | 0.010 mol/L HCl pH | Evaluation |
|---|---|---|---|---|
| 1st | 7.35 | 4.76 | 2.02 | Good |
| 2nd | 7.37 | 4.75 | 2.03 | Good |
| 3rd | 7.34 | 4.77 | 2.01 | Good |
| Average | 7.35 | 4.76 | 2.02 | Representative value |
| Range | 0.03 | 0.02 | 0.02 | Small variation |
If the range of measured values is approximately 0.02–0.03, the reproducibility under the same conditions can be considered relatively good.
Example Comparing pH Test Paper and a pH Meter
| Sample | pH Meter | pH Test Paper | Difference | Direction of Discussion |
|---|---|---|---|---|
| HCl solution | 2.02 | Approx. 2 | Small | Easy to judge for a strong acid |
| Acetic acid solution | 3.18 | Approx. 3 | Somewhat coarse | Test paper relies on visual judgment |
| Tap water | 7.35 | Approx. 7 | Decimal values are difficult to read | Not suitable for precise measurement |
| NaOH solution | 12.02 | Approx. 12 | Small | Easy to judge for a strong base |
| Colored sample | 5.60 | Difficult to determine | Possibly large | Sample color interferes with visual judgment |
pH test paper is convenient, but there are individual differences in color judgment, and a pH meter is more suitable for obtaining precise values to decimal places.
Example of Measurement of Low-Ionic-Strength Samples
| Sample | Immediately After Measurement Begins | After 60 Seconds | After 180 Seconds | Direction of Discussion |
|---|---|---|---|---|
| Distilled water | 6.85 | 6.42 | 6.10 | Fluctuation due to CO2 absorption and electrode response |
| Pure water | 7.05 | 6.55 | 6.20 | Value is difficult to stabilize |
| Buffer solution | 6.86 | 6.86 | 6.86 | Easy to stabilize |
The pH of pure water is theoretically considered to be around 7, but in actual measurements it may show a value around pH 6 because of dissolution of CO2 and electrode stability.
Example of pH Decrease Due to CO2 Absorption
| Sample | Immediately After Preparation | After 10 Minutes | After 30 Minutes | Direction of Discussion |
|---|---|---|---|---|
| Distilled water | 6.85 | 6.45 | 6.20 | CO2 dissolves and shifts the solution toward weak acidity |
| 0.001 mol/L NaOH | 11.00 | 10.72 | 10.40 | OH− is consumed by CO2 |
| Buffer solution | 4.76 | 4.75 | 4.75 | Small pH change |
Particularly in dilute basic solutions, pH tends to decrease due to the effect of CO2 in the air.
Example of Systematic Error Due to Insufficient Calibration
| Sample | Measured Value Before Calibration | Measured Value After Calibration | Difference | Direction of Discussion |
|---|---|---|---|---|
| pH 4 standard solution | 4.13 | 4.01 | −0.12 | Deviation on the acidic side |
| pH 7 standard solution | 6.94 | 6.86 | −0.08 | Deviation near neutral |
| Acetic acid solution | 3.30 | 3.18 | −0.12 | Also affects sample values |
| Tap water | 7.43 | 7.35 | −0.08 | Systematic deviation |
Because deviations before calibration are also reflected in sample measurements, calibration using standard solutions is essential for pH measurement.
Main Error Factors
| Error Factor | Effect on Measured Values | Resulting Tendency | Improvement / Check |
|---|---|---|---|
| Insufficient calibration | Overall pH values shift | Systematic error | Calibrate with standard solutions before measurement |
| Dirty electrode | Slow response and unstable values | Increased variation | Clean the electrode |
| Dry electrode | Poor response | Values do not stabilize | Store in an appropriate storage solution |
| Temperature changes | Electrode response and sample pH change | Comparison becomes difficult | Measure at the same temperature |
| CO2 absorption | pH decreases | Noticeable in water and basic samples | Avoid leaving samples exposed for long periods |
| Low ionic strength | Electrode values become unstable | Values fluctuate easily in pure water | Allow a longer stabilization time |
| Insufficient stirring | Local concentration differences | Measured values vary | Stir gently under consistent conditions |
| Coloration or turbidity | Difficult to judge using test paper | Large visual error | Use a pH meter |
Example of Writing the Results
Using a pH meter, the pH of several acidic, near-neutral, and basic samples was measured. Before measurement, calibration was performed using standard buffer solutions of pH 4.01, pH 6.86, and pH 9.18. Before calibration, the displayed values differed from the standard values by approximately 0.08–0.14, but after calibration they agreed with the standard values. This confirmed that calibration has a major effect on the reliability of pH measurements.
The pH of 0.010 mol/L HCl was 2.02, and the pH of 0.010 mol/L NaOH was 12.02, both of which were close to the theoretical values. Because HCl and NaOH are a strong acid and a strong base and dissociate almost completely in water, their pH values were considered to be close to those expected from their concentrations. In contrast, the aqueous acetic acid solution had a higher pH than a strong acid of the same concentration, reflecting the fact that a weak acid dissociates only partially.
When the same samples were measured multiple times, the variation in pH was approximately 0.02–0.03, indicating relatively good reproducibility. However, immediately after the electrode was immersed in the sample, the displayed value was unstable and stabilized after approximately 30–60 seconds. Therefore, when recording pH, it is necessary to read the value after the electrode response has stabilized.
Points to Connect to the Discussion
In a discussion of pH measurement, it is important to explain not only the measured values but also calibration, comparison with theoretical values, electrode response, temperature, sample properties, and reproducibility in relation to one another.
- Can you explain the relationship pH = −log[H+]?
- Can you compare the theoretical and measured pH values of strong acids and strong bases?
- Can you explain that weak acids and weak bases do not dissociate completely, so their pH values differ even at the same concentration?
- Can you explain the meaning of calibration using standard buffer solutions?
- Can you discuss the difference between measurements before and after calibration as a systematic error?
- Can you explain pH changes caused by dilution and deviations in dilute solutions?
- Can you explain why buffer solutions tend to maintain their pH against the addition of acids or bases and dilution?
- Can you discuss how temperature affects pH and electrode response?
- Can you explain CO2 absorption, low ionic strength, electrode contamination, and reading the value before stabilization as error factors?
Example Discussion
In this experiment, the pH values of several aqueous solutions were measured using a pH meter. When the instrument was calibrated using standard buffer solutions before measurement, deviations of approximately 0.08–0.14 from the standard values were observed before calibration. After calibration, the readings agreed with the standard values, suggesting that calibration corrected shifts in the electrode zero point and sensitivity. These results indicate that if calibration is not performed in pH measurement, systematic errors may also occur in the measured values of samples.
The pH of 0.010 mol/L HCl, a strong acid, was 2.02, which was close to the theoretical value of 2.00. Because HCl dissociates almost completely in water, [H+] is approximately equal to the HCl concentration, and a pH of 2.00 is expected from pH = −log[H+]. In contrast, acetic acid is a weak acid and does not completely dissociate, so even at the same concentration its hydrogen ion concentration is lower than that of HCl, resulting in a higher pH.
The pH of 0.010 mol/L NaOH was 12.02, which was close to the theoretical value of 12.00. NaOH dissociates almost completely as a strong base, giving [OH−] = 1.0×10−2 mol/L. Therefore, pOH is 2.00, and at 25°C, pH is calculated to be 12.00 from pH + pOH = 14.00. Because the measured value was close to the theoretical value, the measurement conditions were considered to have been generally appropriate.
In the dilution experiment, when a strong acid is diluted tenfold, the pH ideally increases by approximately 1. However, when the concentration becomes extremely low, the relative effects of the self-ionization of pure water, dissolution of CO2 from the air, and unstable electrode response become greater, making deviations from theoretical values more likely. Particularly in low-ionic-strength samples such as pure water and distilled water, pH meter readings may be difficult to stabilize and may change depending on the measurement time.
Possible sources of error include insufficient calibration, electrode contamination or drying, temperature changes, CO2 absorption, insufficient stirring, and reading the value before stabilization. In addition, pH test paper involves individual differences in color judgment, and accurate determination is difficult for colored or turbid samples. Therefore, accurate pH measurement requires calibration before measurement, electrode cleaning, consistent temperature conditions, and reading the displayed value only after it has stabilized.
Summary
In pH measurement, it is necessary to consider not only whether a solution is acidic, neutral, or basic, but also comparison with theoretical values, calibration using standard solutions, electrode response, temperature, and sample properties. While theoretical pH can be calculated relatively easily for strong acids and strong bases, the effects of equilibrium and measurement conditions become more significant for weak acids, weak bases, buffer solutions, and low-ionic-strength samples.
This reference example covered the basic pH equation, pH meter calibration, measurement of acidic, neutral, and basic samples, theoretical pH values for strong acids and strong bases, pH changes caused by dilution, buffer solutions, temperature effects, electrode response time, reproducibility, comparison with pH test paper, CO2 absorption, and error factors. In a report, rather than simply presenting measured values, it is useful to discuss why a particular pH was obtained and which conditions affected the measured value.
Understanding the Basic Mechanism of a pH Meter
A pH meter is an instrument that uses a glass electrode or similar electrode to measure an electrical potential difference related to hydrogen ions in a solution and displays that value as pH. The displayed pH is affected by electrode condition, calibration, temperature, and the properties of the sample solution.
Because a pH meter displays a numerical value, it may appear extremely accurate at first glance. However, the value displayed by the instrument may also contain error. Therefore, in a report, rather than treating the displayed value as an absolutely correct value, it is necessary to discuss it while considering the measurement conditions.
Main Causes of Error in pH Meters
There are several causes of error in pH measurement. Typical examples include improper calibration, electrode contamination, temperature differences, insufficient mixing of the sample, insufficient electrode rinsing, and reading the value before it has stabilized.
| Cause of Error | What Happens | Effect on Results |
|---|---|---|
| Insufficient calibration | Measurement is performed with a shifted reference | All pH values shift higher or lower |
| Dirty electrode | Response becomes slower and the value is more difficult to stabilize | Measured values vary |
| Insufficient electrode rinsing | The previous sample contaminates the next sample | Affects the next measured value |
| Temperature difference | Electrode response and solution equilibrium change | pH deviates from the expected value |
| Insufficient mixing | The pH of the entire solution is not uniform | The value changes depending on the measurement location |
| Reading before stabilization | The value is recorded while the display is still changing | Reproducibility decreases |
Error Due to Insufficient Calibration
A pH meter must be calibrated using standard solutions before measurement. If calibration is insufficient, the reference of the pH meter shifts, and all measured values may appear higher or lower than they should.
Particularly when accurately measuring acidic or basic solutions, it is important that the instrument has been calibrated using a standard solution with a pH close to the measurement range. Errors may also occur if the standard solution used for calibration is old, contaminated, or at a temperature substantially different from that of the sample.
Discussion Example:
One possible reason why the measured pH differed from the expected value is insufficient calibration of the pH meter. Because a pH meter displays measured values using standard solutions as references, if calibration is not performed correctly, all measured values may shift higher or lower. Therefore, the deviation in the obtained pH may be attributable not only to the properties of the sample solution but also to the calibration condition of the measuring instrument.
Error Due to Electrode Contamination
If the pH meter electrode is dirty, its response to the solution may become slower and the value may be more difficult to stabilize. Particularly after measuring viscous solutions, solutions containing precipitates, or solutions containing oils or organic substances, contamination may remain on the electrode surface.
Electrode contamination leads to variation in measured values and delayed response. If it took a long time for the value to stabilize or if the value fluctuated even when the same solution was measured, the condition of the electrode can be included in the discussion.
Discussion Example:
One possible reason why it took time for the measured value to stabilize is that contamination may have adhered to the surface of the pH electrode. If the electrode surface is contaminated, its response to the solution becomes slower, making the displayed value more difficult to stabilize. As a result, variation may have occurred in the measured values, affecting accurate pH readings.
Error Due to Insufficient Electrode Rinsing
In pH measurement, the electrode is rinsed before measuring a different sample. If the previous sample remains on the electrode, it may contaminate the next solution being measured and cause the pH to deviate from its original value.
Particularly after measuring a strong acid or strong base solution, measuring another solution without thoroughly rinsing the electrode may have a significant effect on the next measured value.
Discussion Example:
One possible reason why the measured value shifted toward a more acidic or basic value than expected is insufficient rinsing of the electrode. If the previously measured solution remained on the electrode surface, it may have entered the next sample solution and changed its pH. Therefore, a value different from the original pH of the sample may have been obtained.
Effect of Temperature on pH Measurement
pH is affected by temperature. When temperature changes, the equilibrium state of the solution and the response of the electrode change, and the pH value may also change. In addition, if there is a large temperature difference between the standard solution and the sample solution, the calibration and measurement conditions will not match, which may cause error.
Therefore, for accurate pH measurement, it is useful to record the temperature at the time of measurement. In a report, when comparing measured values with theoretical or literature values, you can consider whether the temperature conditions were the same.
Discussion Example:
One possible reason why the measured pH did not agree with the theoretical value is a difference in temperature during measurement. pH changes with temperature, and the response of the pH electrode is also affected by temperature. If the temperature during calibration with the standard solution differed from the temperature during sample measurement, an error may have occurred in the measured value.
Reasons Why Measured Values Do Not Stabilize
With a pH meter, the value may fluctuate immediately after the electrode is placed in the solution. If the value is recorded before it stabilizes, the result may have poor reproducibility.
Possible reasons why the value does not stabilize include delayed electrode response, insufficient mixing of the solution, temperature changes, electrode contamination, and ongoing reactions within the sample.
| Condition | Possible Cause |
|---|---|
| The value continues to change slowly | Delayed electrode response, temperature changes, progression of a reaction |
| The value fluctuates up and down | Electrode contamination, poor contact, nonuniform solution |
| The value differs with each measurement | Insufficient rinsing, poor calibration, insufficient sample mixing |
| The value is far from the expected value | Calibration error, sample preparation error, electrode deterioration |
Discussion Example:
Possible reasons why the displayed pH did not stabilize during measurement include insufficient mixing of the solution and delayed electrode response. If the pH is not uniform throughout the solution, the electrode measures the local pH near the electrode, causing the value to fluctuate. Therefore, it is necessary to mix the solution thoroughly and record the value only after the display has stabilized.
Error Due to Insufficient Mixing of the Solution
Immediately after adding an acid or base, or immediately after preparing a buffer or diluted solution, the pH may not be uniform throughout the solution. If the solution is measured before sufficient mixing, the local pH near the electrode may be measured, and the value may not represent the entire solution.
Discussion Example:
One possible cause of the variation observed in the measured pH values is insufficient mixing of the solution. If the solution is not thoroughly stirred after adding an acid or base, regions with different pH values may remain within the solution. Therefore, the pH near the electrode may differ from the average pH of the entire solution, possibly causing an error in the measured value.
Effect of Carbon Dioxide in the Air
Water and basic solutions may be affected by carbon dioxide in the air. When carbon dioxide dissolves in water, carbonic acid is formed, which may cause the pH to decrease slightly. Particularly in pure water and dilute basic solutions, this effect may not be negligible.
Discussion Example:
One possible reason why the measured pH was lower than expected is that carbon dioxide from the air dissolved in the solution. When carbon dioxide dissolves in water, it forms carbonic acid and shifts the solution slightly toward the acidic side. Therefore, particularly in pure water and dilute basic solutions, the time the solution was left standing and its contact with air may have affected the measured pH.
Points to Consider When Measuring the pH of Buffer Solutions
A buffer solution is a solution whose pH does not change greatly when a small amount of acid or base is added. In buffer solution experiments, discuss whether the measured pH was close to the theoretical or expected value and how much the pH changed after adding an acid or base.
If the pH of a buffer solution differs from the expected value, possible causes include concentration errors during preparation, insufficient mixing, temperature, and the calibration condition of the pH meter.
Discussion Example:
The change in pH was small even after a small amount of acid was added to the buffer solution. This is considered to be because the conjugate base in the buffer solution accepted some of the added hydrogen ions, suppressing a rapid decrease in pH. On the other hand, possible reasons why the measured value did not completely agree with the theoretical value include concentration errors during preparation of the buffer solution and calibration errors of the pH meter.
Points to Consider When Measuring the pH of Strong Acids and Strong Bases
For strong acids and strong bases, pH can sometimes be predicted from the concentration. However, if the concentration is extremely low or the measuring instrument is insufficiently calibrated, the expected and measured values may differ.
In addition, strongly basic solutions may absorb carbon dioxide from the air, causing the pH to decrease. If the electrode is not sufficiently rinsed after measuring a strong acid or strong base, the next measurement may also be affected.
Discussion Example:
One possible reason why the pH of the strongly basic solution was lower than expected is absorption of carbon dioxide from the air. When carbon dioxide dissolves in the solution, part of the base may be neutralized, causing the pH to decrease. In addition, the calibration of the pH meter and the condition of electrode rinsing may also have affected the deviation in the measured value.
Points to Consider When Measuring the pH of Weak Acids and Weak Bases
Weak acids and weak bases do not completely dissociate in water. Therefore, compared with strong acids and strong bases of the same concentration, their pH values are closer to neutral. In pH measurements of weak acids and weak bases, ionization equilibrium and the effect of concentration can be included in the discussion.
Discussion Example:
The aqueous acetic acid solution showed an acidic pH, but its pH is considered to be higher than that of a strong acid at the same concentration. This is because acetic acid is a weak acid and does not completely dissociate in water. Possible reasons why the measured value differed from the theoretical value include errors in concentration preparation, temperature, and the calibration condition of the pH meter.
Difference from pH Test Paper
pH test paper is a simple method for determining pH, but individual differences in color judgment exist, and values as precise as those obtained with a pH meter cannot be obtained. In contrast, a pH meter can measure pH numerically, but it requires calibration and electrode maintenance.
| Method | Characteristics | Causes of Error |
|---|---|---|
| pH test paper | Easy to use for measurement | Color appearance, individual differences in judgment, sample color |
| pH meter | Provides numerical measurements | Calibration, electrode condition, temperature, insufficient rinsing |
Discussion Example:
One possible reason for the difference between the values measured with pH test paper and a pH meter is the difference in measurement method. With pH test paper, there are individual differences in color judgment, and the color of the sample solution may also affect the determination. In contrast, a pH meter measures pH numerically, but errors may occur depending on calibration and electrode condition.
Reasons Why Theoretical and Measured Values Differ
If the theoretical and measured pH values do not agree, possible causes include sample preparation errors, concentration deviations, temperature, electrode condition, and improper calibration. In addition, theoretical calculations often assume ideal conditions, so they may differ from actual solutions.
| Cause | Possible Effect |
|---|---|
| Concentration preparation error | Theoretical and measured pH values differ |
| Approximation of ionization equilibrium | Differences from theoretical values occur for weak acids and weak bases |
| Temperature difference | Equilibrium and electrode response change |
| Improper pH meter calibration | All measured values shift |
| Carbon dioxide in the air | The pH of water and basic solutions decreases |
| Insufficient mixing of the solution | pH differs depending on the measurement location |
Discussion Example:
Possible reasons why the measured pH did not agree with the theoretical value include concentration errors during solution preparation and calibration errors of the pH meter. In addition, because theoretical values are calculated assuming ideal conditions, actual solutions may differ due to the effects of temperature and activity. Therefore, the difference between the measured and theoretical values is considered to be attributable to both sample preparation and measurement conditions.
Discussion When Multiple Measurements Are Performed
When pH is measured multiple times, the reproducibility of the measurement can be evaluated by examining the average and variation of the measured values. If the values agree closely, reproducibility is high; if the variation is large, there may be problems with the measurement procedure or sample condition.
Example Results:
1st measurement: pH 6.82
2nd measurement: pH 6.85
3rd measurement: pH 6.83
Average: pH 6.83
Discussion Example:
The three measured values were within the range of 6.82–6.85, and no large variation was observed. This suggests that the reproducibility of the measurement was relatively high. On the other hand, possible reasons why a difference from the theoretical value remained include the calibration condition of the pH meter and concentration errors during solution preparation.
When the Results Can Be Considered Good
pH measurement results can be considered good when the measured values are close to the theoretical or expected values, there is no large variation among repeated measurements, and the displayed values are stable. It is also important that the determination of acidity, neutrality, or basicity is consistent with the purpose of the experiment.
Discussion Example:
The measured pH was within the expected range, and no large variation was observed among repeated measurements. This suggests that the calibration of the pH meter and the condition of the electrode were generally appropriate and that the reproducibility of the measurements was relatively high. In addition, the obtained pH is considered to have reflected the acid–base properties of the sample solution well.
Discussion Example When the Measurement Did Not Go Well
When pH measurement does not go well, consider the cause based on circumstances such as the value not stabilizing, a large deviation from the theoretical value, or different values being obtained for each measurement. Rather than writing only “measurement error,” it is important to identify specific possible causes.
Discussion Example:
The measured value differed greatly from the theoretical value, and the displayed value was also difficult to stabilize. Possible causes include insufficient calibration of the pH meter, contamination remaining on the electrode surface, or insufficient mixing of the solution. In particular, because the measured value continued to change over time, it is also possible that the value was read before the electrode response had stabilized.
How to Write Improvements
In a discussion of pH measurement, writing not only the causes of error but also improvements makes the report easier to organize. It is important to describe improvements specifically in relation to deviations and variations in the measured values.
Improvements for Increasing Measurement Accuracy
- Properly calibrate the pH meter with standard solutions before measurement.
- Thoroughly rinse the electrode after each measurement.
- Pay attention to contamination and deterioration of the electrode surface.
- Thoroughly mix the solution before measurement.
- Record the value only after the display has stabilized.
- Record the temperature during measurement.
- Perform multiple measurements to confirm reproducibility.
- For solutions that are easily affected by carbon dioxide in the air, minimize the time they are left standing.
Example of Writing Improvements:
To measure pH more accurately, the pH meter should be calibrated with appropriate standard solutions before measurement, and the electrode should be thoroughly rinsed after each measurement. In addition, thoroughly mixing the sample solution and reading the displayed value only after it has stabilized can reduce variation in the measured values.
Difference Between a Superficial Discussion and a Good Discussion
In a discussion of pH measurement, writing only “there was an error in the pH meter” makes the content superficial. Explaining what caused the error and how the measured value changed makes the discussion more convincing.
| Superficial Discussion | Good Discussion |
|---|---|
| There was an error in the pH meter. | If the calibration of the pH meter was insufficient, all measured values may shift higher or lower. Therefore, the difference from the theoretical value may be attributable not only to the properties of the sample solution but also to the calibration condition of the measuring instrument. |
| The value did not stabilize. | Possible reasons why the displayed value did not stabilize include delayed electrode response and insufficient mixing of the solution. If measurement is performed while the pH is not uniform throughout the solution, the local pH near the electrode is read, making the value more likely to fluctuate. |
| It differed from the theoretical value. | Possible reasons why the measured value differed from the theoretical value include concentration errors during solution preparation, temperature during measurement, and the calibration condition of the pH meter. In addition, because theoretical calculations assume ideal conditions, actual solutions may differ due to the effects of activity and temperature. |
Example Expressions That Can Be Used in Reports
The following expressions can be used when writing pH measurement results and discussions. Adjust the necessary parts according to your own experimental results.
- The measured pH generally agreed with the expected acidic or basic range.
- One possible reason why the measured value differed from the theoretical value is a calibration error of the pH meter.
- If the electrode was insufficiently rinsed, the previous sample may have contaminated the next sample and affected the measured value.
- If the value was read before the display stabilized, variation may have occurred in the pH measurement.
- If the measurement temperature differs, the equilibrium of the solution and electrode response change, affecting the measured pH.
- If the solution was insufficiently mixed, the local pH near the electrode may have been measured.
- Dissolution of carbon dioxide from the air may have shifted the pH of the solution toward the acidic side.
- Because repeated measurements gave similar values, the reproducibility of the measurement is considered to have been relatively high.
- For more accurate measurement, calibration, electrode rinsing, temperature control, and confirmation that the measured value has stabilized are important.
Points to Check When Discussing pH Measurement
Before writing the report, checking the following points can make it easier to write the discussion.
- Is the measured pH acidic, neutral, or basic?
- How does it compare with the theoretical or expected value?
- Was the measured value stable?
- If multiple measurements were performed, was the variation large?
- Was the pH meter calibrated with standard solutions?
- Was the electrode thoroughly rinsed?
- Was there any contamination or deterioration of the electrode?
- Was the solution thoroughly mixed?
- Was the measurement temperature recorded?
- Is the solution susceptible to the effects of carbon dioxide in the air?
- Could there have been concentration errors during sample preparation?
- Were specific improvements described?
Summary
pH measurement is a basic experiment used to evaluate whether a solution is acidic, neutral, or basic. A pH meter allows pH to be measured numerically, but errors may occur due to calibration, electrode condition, temperature, and the mixing condition of the solution.
In a discussion of pH measurement, check whether the measured value agreed with the theoretical or expected value, whether the value was stable, and whether repeated measurements showed reproducibility. If the measured value deviated, it is important to specifically consider factors such as improper calibration, electrode contamination, insufficient rinsing, temperature differences, insufficient mixing, and the effects of carbon dioxide.
In a report, rather than simply writing that “the pH was measured,” explain what the value means in relation to the purpose of the experiment. Understanding the causes of pH meter errors and how to interpret the results makes it easier to write discussions for acid–base experiments and buffer solution experiments.
