Chemistry 化学

Instrumental Analysis Laboratory Report Discussion Examples | How to Write About Spectra, Peaks, and Quantitative Results

In instrumental analysis experiments, analytical instruments such as UV-Vis, IR, NMR, fluorescence analysis, atomic absorption analysis, ICP, HPLC, GC, and mass spectrometry are used to investigate the components, structure, concentration, purity, and reaction progress of samples.
The data obtained include spectra, chromatograms, peak areas, peak heights, retention times, absorbance, calibration curves, and quantitative values.

In the discussion section of an instrumental analysis experiment, it is not sufficient simply to write that “a peak appeared,” “the concentration was determined,” or “the result differed from the literature value.”
It is necessary to explain what each peak originates from, what the peak position and intensity mean, whether the calibration curve is valid, and what kinds of errors are included in the quantitative results.

This article clearly explains, as examples of discussions that can be used in instrumental analysis laboratory reports, how to write about spectra, peaks, calibration curves, and quantitative results, as well as sources of error, points for improvement, and expressions that can be used in reports.

Note:
This article is a reference intended to assist with discussions of measurement results obtained in chemistry experiments, analytical chemistry experiments, and instrumental analysis experiments at universities and similar institutions.
For the actual instrument conditions, measurement wavelengths, solvents, standard solutions, calibration-curve ranges, analysis methods, safety precautions, and waste-liquid disposal, always follow the instructions in your university’s laboratory manual and those given by your instructor or TA.

  1. What Is an Instrumental Analysis Experiment?
  2. Main Items to Include in the Results
    1. Main Items to Include in the Results
  3. Reference Experimental Values and How to Organize an Instrumental Analysis Report
    1. Reference Experimental Conditions
    2. Basic Table for Organizing Analytical Results
    3. Example of Organizing UV-Vis Results
    4. Example Calculation of UV-Vis Concentration
    5. Example of Organizing IR Spectrum Results
    6. Example of Organizing 1H NMR Results
    7. Example of Organizing GC Results
    8. Example of Organizing an HPLC Calibration Curve and Quantitative Results
    9. Example of Organizing MS Results
    10. Example of Structural Estimation by Integrating Multiple Analytical Results
    11. How to Summarize Quantitative Results
    12. Example of Checking Calibration-Curve Linearity
    13. Table for Confirming the Reliability of Peak Identification
    14. Example of Organizing Sources of Error
    15. Example of Connecting Results and Discussion
    16. Examples of Writing to Avoid in the Discussion and Improved Versions
    17. Example Calculation of Relative Error
    18. Example of a Results Table That Is Easy to Include in a Report
    19. Points for Connecting the Results to the Discussion
    20. Example Discussion
    21. Summary
  4. What Is a Spectrum?
  5. What Does a Peak Indicate?
  6. Discussion of Peak Position
  7. Discussion of Peak Intensity
  8. Discussion of Peak Area
  9. Discussion of Peak Width and Peak Shape
  10. How to Write Peak Assignments
  11. What Is a Calibration Curve?
  12. Discussion of Calibration-Curve Linearity
  13. How to Write Quantitative Results
  14. Discussion of Spectrophotometry
  15. Discussion of IR Spectra
  16. Discussion of NMR Spectra
  17. Discussion of Chromatography
  18. Discussion of Mass Spectrometry
  19. Discussion of Atomic Absorption Analysis and ICP Analysis
  20. Discussion of Blank Measurements
  21. Discussion of the Baseline
  22. Discussion of Noise
  23. Discussion When Peaks Overlap
  24. Errors Caused by Sample Preparation
  25. Errors Caused by Instrument Conditions
  26. Causes of Quantitative Values Being Too High
  27. Causes of Quantitative Values Being Too Low
  28. How to Judge the Validity of Results
  29. Discussion When the Result Differs From a Literature or Labeled Value
  30. When the Results Can Be Considered Good
  31. Example Discussion When the Experiment Did Not Go Well
  32. How to Write Points for Improvement
    1. Improvements to Sample Preparation
    2. Improvements to Standard Solutions and Calibration Curves
    3. Improvements to Measurement and Analysis
  33. Difference Between a Superficial Discussion and a Good Discussion
  34. Examples of Expressions That Can Be Used in Reports
  35. Points to Check When Discussing Instrumental Analysis Experiments
  36. Summary

What Is an Instrumental Analysis Experiment?

An instrumental analysis experiment is an experiment in which the properties of chemical substances are investigated using measuring instruments.
Information that is difficult to obtain through visual observation or manual titration alone can be quantified as spectra, peaks, electrical signals, light intensity, mass-to-charge ratios, retention times, and similar data.
Therefore, instrumental analysis is used for the quantitation of trace components, identification of unknown samples, confirmation of functional groups, separation and analysis of mixtures, and other purposes.

It is important not merely to write down the data output by the instrument, but to convert those data into chemical meaning and discuss them.
For example, an IR absorption peak represents a functional group, UV-Vis absorption represents an electronic transition or concentration, HPLC and GC peaks represent component retention times and amounts, and NMR signals represent the environments of hydrogen or carbon atoms.

Example Discussion:
In instrumental analysis, the chemical properties of a sample are observed as spectra and peaks.
By analyzing the positions, intensities, areas, and shapes of the obtained peaks, information about the components, concentrations, and structures in the sample can be obtained.
Therefore, in a report, it is necessary not only to present the measurement data but also to discuss what each peak originates from and what chemical meaning it has.

Main Items to Include in the Results

In the results of an instrumental analysis experiment, organize the analyte, sample-preparation method, instrument used, measurement conditions, spectra or chromatograms, peak positions, peak intensities, peak areas, calibration curves, quantitative results, and other information.
Because the results change when the measurement conditions change, it is important to describe the instrument conditions as clearly as possible.

Main Items to Include in the Results

  • Analyte
  • Sample name
  • Sample-preparation method
  • Dilution factor
  • Standard solutions used
  • Instrument used
  • Measurement wavelength
  • Measurement range
  • Mobile phase and stationary phase
  • Flow rate
  • Column conditions
  • Measurement temperature
  • Spectrum or chromatogram
  • Peak position
  • Peak height
  • Peak area
  • Retention time
  • Calibration curve
  • Correlation coefficient
  • Unknown-sample concentration
  • Comparison with literature values and labeled values
  • Sources of error and points for improvement

Example of How to Write the Results:
A calibration curve was prepared using standard solutions, and the concentration of the unknown sample was determined by substituting its measured value into the calibration curve.
In the obtained spectrum, a peak corresponding to the target component was confirmed.
In addition, because the calibration curve showed approximate linearity within the measured concentration range, it was considered usable for quantitation within this range.

Reference Experimental Values and How to Organize an Instrumental Analysis Report

Here, reference experimental values are used to show how spectra, peaks, calibration curves, and quantitative results obtained by instrumental analysis methods such as UV-Vis, IR, NMR, GC, HPLC, and MS can be organized in a report and connected to the discussion.

In instrumental analysis, it is important not merely to list measurement values, but to explain “what was learned” and “why that judgment can be made” based on peak positions, peak intensities, areas, retention times, calibration curves, and comparisons with standard substances.

Reference Experimental Conditions

Item Details
Analytes Unknown sample A, standard substances, standard mixture
Analytical methods used UV-Vis, IR, 1H NMR, GC, HPLC, MS
Purpose Component identification, functional-group estimation, concentration quantitation, purity evaluation
Evaluation items Absorption maximum, functional-group peaks, chemical shifts, retention times, peak areas, molecular ions, calibration curves
Points emphasized in the report Organization of results, explanation of evidence, sources of error, consistency among multiple data sets

Basic Table for Organizing Analytical Results

First, listing what was confirmed by each analytical method makes the overall report easier to understand.

Analytical Method Main Measurement Result What Can Be Determined Point to Use in the Discussion
UV-Vis λmax = 525 nm, absorbance 0.496 Absorption and concentration of a dye component Determine concentration from the calibration curve
IR Strong absorption at 1740 cm−1 and 1240 cm−1 Possibility of an ester group Judge from the combination of C=O and C-O
1H NMR 1.26 ppm triplet, 4.12 ppm quartet, 2.05 ppm singlet Ethyl group, acetyl group Estimate partial structures from integration ratios and splitting
GC Retention time 2.65 min, peak area 315000 Identification and relative amount of a volatile component Compare retention time with that of a standard substance
HPLC Retention time 4.82 min, peak area 456000 Concentration of the target component Quantify by substituting into the calibration curve
MS m/z 136, 105, 77 Molecular weight and fragments Confirm the molecular ion and characteristic fragments

Example of Organizing UV-Vis Results

In UV-Vis measurements, the absorption-maximum wavelength and absorbance are organized, and the concentration is determined from the calibration curve.

Sample Measurement Wavelength Absorbance Calibration Curve Determined Concentration Direction of Discussion
Standard solution 525 nm 0.620 A = 31000c 2.00×10−5 mol/L Used to prepare the calibration curve
Unknown sample A 525 nm 0.496 A = 31000c 1.60×10−5 mol/L Lower concentration than the standard solution
5-fold diluted sample 525 nm 0.372 A = 31000c Original sample: 6.00×10−5 mol/L Take the dilution factor into account

Example Calculation of UV-Vis Concentration

If the calibration curve is A = 31000c and the absorbance of unknown sample A is 0.496,

c = A ÷ 31000 = 0.496 ÷ 31000 = 1.60×10−5 mol/L

For a diluted sample, the concentration in the measured solution is multiplied by the dilution factor to obtain the concentration in the original sample.

Example of Organizing IR Spectrum Results

In IR analysis, characteristic absorption peaks are organized according to wavenumber, intensity, assignment, and structural significance.

Wavenumber Absorption Intensity Estimated Assignment Meaning in Structural Estimation
2980 cm−1 Medium C-H stretching Presence of an alkyl group
1740 cm−1 Strong C=O stretching Possibility of an ester carbonyl
1240 cm−1 Strong C-O stretching Possibility of ester C-O
1040 cm−1 Medium C-O stretching Ester or ether
Near 3300 cm−1 Absent No O-H absorption Alcohols and carboxylic acids are unlikely

From the IR results, the unknown sample is considered highly likely to contain an ester group.
However, because IR alone makes it difficult to distinguish the carbon skeleton or isomers, the result is judged together with NMR and MS data.

Example of Organizing 1H NMR Results

In NMR, chemical shifts, integration ratios, and splitting patterns are organized together in a table.

Chemical Shift Integration Ratio Splitting Estimated Hydrogen Structural Clue
1.26 ppm 3H triplet CH3CH2- Terminal portion of an ethyl group
2.05 ppm 3H singlet CH3CO- Acetyl group
4.12 ppm 2H quartet -OCH2- CH2 adjacent to oxygen

The triplet at 1.26 ppm and the quartet at 4.12 ppm correspond to an ethyl group with an integration ratio of 3:2.
The 3H singlet at 2.05 ppm is considered to correspond to CH3CO-, which has no neighboring hydrogens.

Example of Organizing GC Results

In GC, retention time and peak area are used for component identification and comparison of relative amounts.

Peak Retention Time Comparison With Standard Substance Peak Area Area Percentage Estimated Component
1 1.25 min Ethanol 1.25 min 185000 18.5% Ethanol
2 1.80 min Acetone 1.80 min 260000 26.0% Acetone
3 2.65 min Ethyl acetate 2.65 min 315000 31.5% Ethyl acetate
4 4.40 min Toluene 4.40 min 240000 24.0% Toluene

The area percentage is determined by dividing each peak area by the total peak area.

Area percentage (%) = Peak area of each component ÷ Total peak area × 100

However, because detector sensitivity differs among components, the area percentage cannot be regarded directly as a mass percentage.

Example of Organizing an HPLC Calibration Curve and Quantitative Results

In HPLC, a calibration curve is prepared from standard solutions, and the concentration is determined from the peak area of the unknown sample.

Standard Solution Concentration Retention Time Peak Area Use in Calibration Curve
Standard 1 5 mg/L 4.80 min 91000 Used
Standard 2 10 mg/L 4.80 min 182000 Used
Standard 3 20 mg/L 4.81 min 365000 Used
Standard 4 30 mg/L 4.81 min 548000 Used
Unknown sample 4.82 min 456000 Concentration calculation

In this example, the calibration curve is treated as follows.

Peak area = 18250 × Concentration (mg/L)

If the peak area of the unknown sample is 456000,

Concentration = 456000 ÷ 18250 = 25.0 mg/L

Example of Organizing MS Results

In MS, molecular-ion peaks and fragment peaks are organized to estimate molecular weight and partial structures.

m/z Relative Intensity Estimated Ion Meaning in Structural Estimation
136 60 Molecular ion M+ Candidate molecular weight of 136
105 100 C6H5CO+ Possibility of a benzoyl structure
77 42 C6H5+ Presence of an aromatic ring
31 9 OCH3+ Possibility of a methoxy group

From the molecular ion at m/z 136, the base peak at m/z 105, and the aromatic-derived peak at m/z 77, candidates such as aromatic esters can be considered.

Example of Structural Estimation by Integrating Multiple Analytical Results

When estimating the structure of an unknown sample, it is important not to make a judgment based on only one analytical method, but to confirm whether multiple analytical results support the same structure.

Analytical Method Observed Result Information Indicated Relationship to Candidate Structure
IR C=O at 1740 cm−1, C-O at 1240 cm−1 Ester group Supports ethyl acetate
1H NMR 1.26 ppm t, 4.12 ppm q, 2.05 ppm s Ethyl group, acetyl group Supports ethyl acetate
GC Retention time 2.65 min Matches standard ethyl acetate Supports ethyl acetate
MS Molecular ion at m/z 88, base peak at m/z 43 Molecular weight and fragments of ethyl acetate Supports ethyl acetate
Overall judgment All analytical results agree Unknown sample is highly likely to be ethyl acetate High reliability of structural estimation

How to Summarize Quantitative Results

Quantitative results become easier to understand when the measured value, calibration curve, dilution factor, and final concentration are summarized in a single table.

Analytical Method Target Component Measured Value Calibration Curve Concentration in Measured Solution Dilution Factor Concentration in Original Sample
UV-Vis Permanganate ion A = 0.496 A = 31000c 1.60×10−5 mol/L 1.60×10−5 mol/L
HPLC Caffeine Area 456000 Area = 18250C 25.0 mg/L 25.0 mg/L
GC Ethanol Area 222000 Area = 370000C 0.600% 3.00%

In a report, not only the final concentration but also which calibration curve was used and how the dilution factor was handled should be clearly stated.

Example of Checking Calibration-Curve Linearity

In quantitative analysis, confirm whether the calibration curve is linear.
If high-concentration points deviate from the straight line, the sample must be diluted before measurement.

Concentration Measured Value Theoretical Value Deviation Judgment
5 mg/L 91000 91250 −0.3% Within the linear range
10 mg/L 182000 182500 −0.3% Within the linear range
20 mg/L 365000 365000 0.0% Within the linear range
40 mg/L 730000 730000 0.0% Within the linear range
80 mg/L 1320000 1460000 −9.6% Possibly outside the linear range

If a value outside the calibration-curve range is extrapolated without adjustment, the concentration may be determined incorrectly.

Table for Confirming the Reliability of Peak Identification

In component identification, showing multiple lines of evidence such as retention time, spectrum, standard substance, and library-match score increases persuasiveness.

Confirmation Item Result Judgment Reliability
Retention time Matches standard substance within 0.02 min Match High
Spectral peaks Major peaks agree with standard data Match High
Standard addition Peak area at the same retention time increases Supports the target component High
Impurity peaks Some small peaks are present Possibility of trace impurities Caution required
Consistency with another analytical method IR, NMR, and MS support the same structure Consistent Very high

Example of Organizing Sources of Error

In an instrumental analysis report, sources of error are easier to write about when divided into “instrument,” “sample,” “operation,” and “analysis.”

Category Source of Error Effect on the Result Improvement Method
Instrument Wavelength shift, baseline drift, variation in detector sensitivity Peak positions and intensities shift Calibration, blank measurement, allow time for stabilization
Sample Concentration too high, turbidity, impurities, decomposition Absorbance or peak area becomes inaccurate Dilution, filtration, re-preparation, management of storage conditions
Operation Pipetting, variation in injection volume, dirty cell Variation in quantitative values Clean equipment, perform multiple measurements, use the internal-standard method
Analysis Peak overlap, integration-range setting, extrapolation beyond the calibration curve Errors in concentration or identification results Improve separation conditions, measure within the linear range

Example of Connecting Results and Discussion

In UV-Vis measurement, the absorption maximum of unknown sample A was observed near 525 nm.
This wavelength agreed with the absorption maximum of the standard substance, suggesting that unknown sample A may contain the same absorbing component.
The absorbance at 525 nm was 0.496, and substitution into the calibration curve A = 31000c gave a concentration of 1.60×10−5 mol/L.

In the IR spectrum, strong C=O absorption was observed at 1740 cm−1 and strong C-O absorption at 1240 cm−1.
These absorptions are characteristic of an ester group.
In addition, because no broad O-H absorption was observed near 3300 cm−1, the sample was considered not to be an alcohol or carboxylic acid.

In the 1H NMR spectrum, a 3H triplet was observed at 1.26 ppm, a 2H quartet at 4.12 ppm, and a 3H singlet at 2.05 ppm.
The combination of the peaks at 1.26 ppm and 4.12 ppm corresponds to an ethyl group, and the singlet at 2.05 ppm is considered to correspond to CH3CO-.
Together with the ester group indicated by IR, the unknown sample is highly likely to be ethyl acetate.

In GC, the retention time of the main peak of the unknown sample was 2.65 min, which agreed with the retention time of standard ethyl acetate.
Because the IR, NMR, and GC results all support ethyl acetate, the main component of the unknown sample was judged to be ethyl acetate.

Examples of Writing to Avoid in the Discussion and Improved Versions

Writing to Avoid Problem Improved Writing
I think it is ethyl acetate because a peak appeared. The basis is unclear. The C=O absorption at 1740 cm−1, C-O absorption at 1240 cm−1, and ethyl-group peaks in NMR agreed with those of ethyl acetate.
The concentration was high because the absorbance was high. No relationship to the calibration curve is shown. Substituting the absorbance of 0.496 into the calibration curve A = 31000c gave a concentration of 1.60×10−5 mol/L.
I think the error was caused by an operational mistake. Not specific. Possible sources of error include pipetting, cell contamination, insufficient blank correction, and measurement outside the calibration-curve range.
It differed from the standard value. The magnitude of the difference is unclear. The measured value was 23.8 mg/L compared with the standard value of 25.0 mg/L, giving a relative error of 4.8%.

Example Calculation of Relative Error

When comparing a measured value with a standard or labeled value, using relative error makes the magnitude of the difference easier to explain.

Relative error (%) = |Measured value − Standard value| ÷ Standard value × 100

If the standard value is 25.0 mg/L and the measured value is 23.8 mg/L,

Relative error = |23.8 − 25.0| ÷ 25.0 × 100 = 4.8%

In this case, the measured value can be described as 4.8% lower than the standard value.

Example of a Results Table That Is Easy to Include in a Report

Item Result Basis Discussion
Absorption maximum 525 nm UV-Vis spectrum Agrees with the standard substance
Quantitative result 1.60×10−5 mol/L Calibration curve A = 31000c Calculated within the linear range
Functional group Ester group C=O and C-O absorption in IR No O-H absorption
Partial structure Ethyl group, acetyl group Integration and splitting in 1H NMR Agrees with ethyl acetate
Component identification Highly likely to be ethyl acetate Consistency among IR, NMR, GC, and MS Multiple analytical results support the same conclusion

Points for Connecting the Results to the Discussion

In an instrumental analysis report, it is important not only to explain each measurement result separately, but also to combine multiple results to reach a conclusion.

  • Are the peak positions in the spectrum or chromatogram shown as specific numerical values?
  • Can the peak assignments be explained based on functional groups, partial structures, and comparison with standard substances?
  • Is the calibration-curve equation shown, and is the calculation process for the unknown-sample concentration described?
  • Are treatments such as dilution factor, blank correction, and area percentage clearly stated?
  • Has it been confirmed that the quantitative result lies within the linear range of the calibration curve?
  • Is the conclusion based on consistency among multiple analytical methods rather than on a single analytical result?
  • Have sources of error such as peak overlap, impurities, instrument conditions, and sample preparation been discussed?
  • Are the results, calculations, discussion, and conclusion connected logically?

Example Discussion

In this experiment, multiple instrumental analysis methods were used to identify and quantify the components of an unknown sample.
In UV-Vis measurement, an absorption maximum was observed at 525 nm, which agreed with the absorption maximum of the standard substance.
Substituting the absorbance of 0.496 at this wavelength into the calibration curve A = 31000c gave a concentration of 1.60×10−5 mol/L in the unknown sample.

In the IR spectrum, strong C=O absorption was observed at 1740 cm−1 and strong C-O absorption at 1240 cm−1.
These are absorptions characteristic of an ester group and indicate that the unknown sample has an ester structure.
On the other hand, because no broad O-H absorption was observed near 3300 cm−1, the possibilities of an alcohol or carboxylic acid are low.

In the 1H NMR spectrum, a triplet at 1.26 ppm and a quartet at 4.12 ppm were observed with an integration ratio of 3:2.
This combination corresponds to a CH3CH2- ethyl group.
In addition, because a 3H singlet was observed at 2.05 ppm, an acetyl group, CH3CO-, is considered to be present.
Together with the ester group indicated by IR, the unknown sample is highly likely to be ethyl acetate.

In GC, the retention time of the main peak of the unknown sample was 2.65 min and agreed with the retention time of standard ethyl acetate.
In MS, the molecular ion and major fragments were also consistent with the candidate compound.
From these results, the IR, NMR, GC, and MS analyses all support the same conclusion, and the main component of the unknown sample can be judged to be ethyl acetate.

Possible sources of error include dilution errors during sample preparation, variation in injection volume, settings for the peak integration range, and insufficient blank correction.
In quantitative analysis in particular, measurement outside the linear range of the calibration curve may prevent correct determination of concentration.
In addition, when peaks overlap on a chromatogram, their areas may be overestimated or underestimated.
Therefore, comparison with standard substances, repeated measurements, appropriate dilution, and confirmation of peak separation are necessary.

Summary

In an instrumental analysis report, the peaks in spectra and chromatograms are shown numerically, and what those peaks mean is explained together with the supporting evidence.
In quantitation, it is important to clearly organize the calibration curve, measured value, dilution factor, and final concentration.

This reference example used results from UV-Vis, IR, NMR, GC, HPLC, and MS to address peak assignments, component identification, concentration calculations using calibration curves, sources of error, and consistency among multiple analytical methods.
In a report, rather than merely listing measurement results, the discussion should connect the results to what can be concluded from them.

What Is a Spectrum?

A spectrum represents the distribution of signal intensity with respect to wavelength, wavenumber, frequency, mass-to-charge ratio, or a similar variable.
In UV-Vis, it is expressed as the relationship between absorbance and wavelength; in IR, as the relationship between transmittance or absorbance and wavenumber; and in NMR, as the relationship between chemical shift and signal intensity.
In mass spectrometry, the spectrum is obtained as the relationship between m/z and ion intensity.

Spectra contain information about the structure, functional groups, electronic state, molecular weight, chemical environment, and other properties of a substance.
In a report, attention is paid not only to the overall spectral shape but also to characteristic peaks and signals.

Example Discussion:
A spectrum is data showing how a sample responds to specific wavelengths or wavenumbers.
The characteristic peaks observed in this experiment are considered to originate from specific functional groups or components in the sample.
Therefore, the structure and components of the sample can be estimated by comparing the peak positions and intensities with those of known substances or literature values.

What Does a Peak Indicate?

A peak is a region of increased signal intensity in a spectrum or chromatogram.
Peak position provides clues to components, functional groups, electronic states, retention times, and other information.
Peak intensity and peak area may be related to the amount or concentration of a component.

However, the presence of a peak does not immediately prove that the target component is present.
Overlap with other components, noise, baseline disturbances, and peaks from solvents or impurities must also be considered.
When discussing a peak, its position, intensity, shape, width, and reproducibility are checked.

Example Discussion:
The observed peak may originate from the target component.
However, the effects of impurities or solvents with similar peak positions, peak overlap, and noise must also be considered.
To determine whether the target component is present, it is important to compare the peak position or retention time with that of a standard substance and also confirm measurement reproducibility.

Discussion of Peak Position

Peak position is one of the most important pieces of information in instrumental analysis.
In IR, it indicates the type of functional group; in UV-Vis, the wavelength of absorbed light; in NMR, the chemical environment of atoms; in HPLC and GC, the retention time of components; and in mass spectrometry, m/z.
Components and structures can be estimated by comparing peak positions with literature values or standard substances.

If a peak position differs from the literature value, possible causes include solvent, pH, temperature, concentration, instrument calibration, measurement conditions, and interactions within the sample.
In a report, not only agreement of peak positions but also the reasons for any shifts should be discussed.

Example Discussion:
Because the observed peak position approximately agreed with that of the standard substance, this peak is considered to originate from the target component.
On the other hand, the slight difference from the literature value may have been caused by differences in measurement solvent, pH, concentration, or instrument calibration.
Therefore, differences in measurement conditions must also be considered when comparing peak positions.

Discussion of Peak Intensity

Peak intensity represents the magnitude of the detected signal.
In UV-Vis it is treated as absorbance, in fluorescence analysis as fluorescence intensity, in NMR as integration value, and in chromatography as peak height or peak area.
In many cases, peak intensity is related to the amount or concentration of a component.

However, peak intensity is affected not only by concentration but also by measurement conditions, sample preparation, instrument sensitivity, cell contamination, injection volume, and peak overlap.
When it is used for quantitation, it is important to prepare a calibration curve using standard solutions and evaluate the unknown sample within that range.

Example Discussion:
A larger peak intensity may indicate that a greater amount of the target component is present in the sample.
However, because peak intensity is also affected by measurement conditions, instrument sensitivity, and sample injection volume, concentration cannot be judged directly from intensity alone.
For quantitation, a calibration curve must be prepared from the measurement results of standard solutions, and the signal of the unknown sample must be evaluated within that range.

Discussion of Peak Area

In chromatography and some spectral analyses, peak area is often used for quantitation rather than peak height.
Because peak area reflects the total amount of signal across the entire peak, it may allow the amount of a component to be evaluated even if the peak becomes slightly broader.
In HPLC and GC, quantitation is performed using a range in which peak area is proportional to the amount of component.

However, correct determination of peak area requires appropriate baseline setting.
If peaks overlap or the baseline is sloped, errors occur in the area calculation.
In a report, the effect of peak area on the quantitative value should be discussed.

Example Discussion:
In this experiment, quantitation was performed using the peak area of the target component.
Because peak area corresponds to the amount of component detected, a calibration curve can be prepared from the relationship between the peak areas and concentrations of the standard solutions.
However, if the baseline is set incorrectly or adjacent peaks overlap, errors occur in area calculation and also affect the quantitative results.

Discussion of Peak Width and Peak Shape

Peak width and peak shape are also important pieces of information.
In chromatography, a sharper peak generally indicates better separation and less diffusion or spreading within the column.
If a peak is broad, tails, or is asymmetric, possible causes include column condition, sample amount, interactions, mobile-phase conditions, and instrument problems.

In spectra as well, peak width is affected by intermolecular interactions, concentration, temperature, instrument resolution, and other factors.
For example, a broad peak in IR may indicate hydrogen bonding, while a broad signal in NMR may indicate exchange reactions or molecular motion.

Example Discussion:
Possible causes of the broad observed peak include spreading of the component during measurement and overlap of multiple nearby peaks.
In chromatography, an increase in peak width may indicate reduced resolution or diffusion within the column.
In addition, if the peak tails, the interaction between the sample and stationary phase may have been too strong.

How to Write Peak Assignments

Peak assignment means explaining which component, functional group, bond, atom, or transition an observed peak originates from.
For IR, it can be written as “absorption originating from C=O stretching vibration”; for UV-Vis, “absorption originating from a π-π* transition”; for NMR, “a signal originating from aromatic protons”; and for HPLC, “a component whose retention time agrees with that of the standard substance.”

When writing an assignment, the peak position and the basis for the assignment are stated together.
Rather than merely writing that “a peak was present,” explaining “where it appeared, what it corresponds to, and what can therefore be concluded” makes the text more suitable as a discussion.

Example of How to Write a Peak Assignment:
In the measured spectrum, a peak characteristic of the target component was observed.
Because this peak appeared at a position close to that of the standard substance or literature value, it is considered to originate from the target component.
Therefore, the target component is highly likely to be present in the sample.

What Is a Calibration Curve?

A calibration curve is a graph prepared by measuring standard solutions of known concentration and plotting the relationship between concentration and signal intensity.
The concentration of an unknown sample is determined by applying its signal to the calibration curve.
Calibration curves are commonly used in quantitative analyses such as UV-Vis, fluorescence analysis, atomic absorption analysis, HPLC, and GC.

To use a calibration curve, it is important that the signal of the unknown sample lie within the calibration-curve range.
If a concentration is determined by extrapolating a value outside the range, the error may become large.
The correlation coefficient and linearity of the calibration curve must also be confirmed.

Example of a calibration curve: Signal intensity = Slope × Concentration + Intercept

Example Discussion:
A calibration curve was prepared from the relationship between the concentrations of the standard solutions and their measured signals.
Because the calibration curve showed good linearity, signal intensity is considered proportional to concentration within this concentration range.
Because the signal of the unknown sample was within the calibration-curve range, determining the concentration using the calibration curve was considered valid.

Discussion of Calibration-Curve Linearity

The linearity of a calibration curve is closely related to the reliability of quantitative results.
When the concentrations and signals of the standard solutions have a linear relationship, the concentration of an unknown sample can be determined relatively accurately.
A correlation coefficient closer to 1 generally makes it easier to judge the linearity as high.

However, even a high correlation coefficient does not necessarily mean that the calibration curve is completely accurate.
Errors in standard-solution preparation, bias in the concentration range, blank correction, the size of the intercept, and the presence of outliers must be checked.
If high-concentration points deviate from the straight line, detector saturation or limits of the proportional relationship between absorbance and concentration may be responsible.

Example Discussion:
Because the correlation coefficient of the calibration curve was high, a good linear relationship was considered to exist between standard-solution concentration and signal intensity.
However, if points on the high-concentration side deviate from the straight line, possible causes include exceeding the detector-response range or excessively high sample concentration.
For quantitation, data within a range in which linearity has been confirmed must be used.

How to Write Quantitative Results

When writing quantitative results, include not only the determined concentration but also the calibration curve used in the calculation, dilution factor, units, number of measurements, and average value.
For example, if an unknown sample is diluted 10-fold before measurement, the concentration determined from the calibration curve is multiplied by the dilution factor to obtain the concentration in the original sample.
Forgetting to restore the dilution factor is a common mistake.

If a quantitative value differs from a literature value, labeled value, or theoretical value, the effects of sample preparation, standard-solution preparation, measurement conditions, interfering components, peak overlap, and the calibration-curve range are discussed.
It is important not to end with “the value was calculated,” but to evaluate the validity of the quantitative result.

Example of How to Write Quantitative Results:
The measured signal of the unknown sample was substituted into the calibration curve to determine the concentration of the diluted sample.
The dilution factor used during sample preparation was then taken into account to calculate the concentration of the target component in the original sample.
Because the obtained quantitative value was close to the labeled value, the analytical method used in this experiment was considered generally valid.

Discussion of Spectrophotometry

In spectrophotometry, the amount of light at a specific wavelength absorbed by a sample is measured.
Absorbance is measured at a wavelength absorbed by the target component, and the concentration is determined using a calibration curve.
Because there is a range in which absorbance is proportional to concentration, it is important to perform measurements within that range.

If absorbance is too high, the result may deviate from linearity.
Cell contamination, bubbles, insufficient blank correction, sample turbidity, and absorption by coexisting substances are also sources of error.

A = εcl

Example Discussion:
Because absorbance increased as the concentration increased, absorption of light by the target component is considered to depend on concentration.
Within the range where the calibration curve showed linearity, absorbance can be used for quantitation.
On the other hand, samples with excessively high absorbance may deviate from the linear relationship and should therefore be diluted when necessary before measurement.

Discussion of IR Spectra

In IR spectra, absorption peaks corresponding to vibrations of bonds and functional groups in molecules are observed.
Functional groups such as O-H, N-H, C=O, C-H, C=C, and C-O show absorption in characteristic wavenumber ranges.
Therefore, IR is used to confirm functional groups and structural changes before and after reactions.

In an IR discussion, state which functional group each characteristic absorption peak corresponds to.
In addition, if a peak originating from the starting material disappears after the reaction or a peak characteristic of the product appears, this change can be used to discuss reaction progress.

Example Discussion:
Because strong absorption was observed near a specific wavenumber in the IR spectrum, the corresponding functional group is considered to be present in the sample.
If a peak originating from the starting material decreases after the reaction and a peak characteristic of the product appears, this provides evidence that the intended reaction proceeded.
However, care is required in peak assignment if peaks overlap or absorption originating from moisture is present.

Discussion of NMR Spectra

In NMR spectra, differences in the chemical environments of atomic nuclei appear as chemical shifts.
In 1H NMR, information about molecular structure can be obtained from the types of hydrogen atoms, integration values, and splitting patterns.
In 13C NMR, the carbon skeleton and carbon environments around functional groups can be confirmed.

In an NMR discussion, the structure is explained based on chemical shifts, integration ratios, splitting, and agreement with known spectra.
Because solvent peaks, water peaks, and impurity peaks may also appear, care must be taken not to assign every peak to the target substance.

Example Discussion:
Because chemical shifts and integration ratios corresponding to the target compound were confirmed in the NMR spectrum, the target substance is considered to have been formed.
The splitting pattern reflects the number of neighboring hydrogens and can be used for structural estimation.
On the other hand, signals originating from solvent peaks, moisture, and impurities may also be present, so caution is required when assigning peaks.

Discussion of Chromatography

In HPLC and GC, components in mixtures are separated, and retention time and peak area are used for component identification and quantitation.
If a retention time agrees with that of a standard substance, the corresponding component may be present.
Because peak area corresponds to the amount of component, it can be quantified using a calibration curve.

In a chromatographic discussion, retention time, peak area, resolution, and peak shape are checked.
If peaks overlap, accurate quantitation becomes difficult.
Injection volume, column condition, mobile-phase composition, flow rate, and temperature also affect the results.

Example Discussion:
Because the retention time of the peak in the unknown sample agreed with that of the standard substance, this peak is considered to originate from the target component.
In addition, the concentration of the target component was determined by substituting the peak area into the calibration curve.
However, if there is overlap with a nearby peak or disturbance of the baseline, an error occurs in the peak-area calculation and affects the quantitative result.

Discussion of Mass Spectrometry

In mass spectrometry, the m/z values of ionized molecules and fragment ions are measured.
Molecular-ion peaks and fragment peaks provide information about molecular weight and structure.
In high-resolution mass spectrometry, elemental composition may also be estimated from exact mass.

In a mass-spectrometry discussion, molecular-ion peaks, major fragments, isotope peaks, and comparisons with standard substances or theoretical values are considered.
However, depending on the ionization method, the molecular ion may be weak, adduct ions may appear, or fragmentation may occur readily.

Example Discussion:
Because a peak at an m/z value corresponding to the molecular weight of the target compound was observed in the mass spectrum, the target component is highly likely to be present.
In addition, the observed fragment peaks are considered to originate from cleavage of specific bonds within the molecule.
However, because adduct ions and fragment ions may be observed depending on the ionization conditions, care is required when assigning peaks.

Discussion of Atomic Absorption Analysis and ICP Analysis

Atomic absorption analysis and ICP analysis are used to quantify the concentrations of metallic elements and inorganic elements.
A calibration curve is prepared using standard solutions, and the element concentration is determined from the signal intensity of the unknown sample.
In trace-element analysis, sample preparation, dilution, contamination, and matrix effects have a large influence on the results.

In the discussion, the linearity of the calibration curve, blank value, concentration range of the standard solutions, dilution factor, and effects of coexisting components are checked.
Trace contamination from glassware, water, and reagents may also be a source of error.

Example Discussion:
The concentration of metal ions in the unknown sample was determined from a calibration curve prepared using standard solutions.
Because the calibration curve showed linearity, the relationship between signal intensity and concentration was considered usable for quantitation within the measurement range.
However, matrix effects caused by coexisting components in the sample and trace contamination from equipment or reagents may have affected the quantitative value.

Discussion of Blank Measurements

A blank measurement is an operation in which a sample containing no target component is measured to confirm signals originating from the solvent, reagents, cell, or instrument.
If the blank value is large, factors other than the target component may be affecting the signal.
In quantitative analysis, blank correction is used to obtain a more accurate signal.

If the blank is not measured appropriately, the concentration of the target component may be overestimated.
Particularly in trace analysis, the effect of the blank value becomes relatively large.

Example Discussion:
Blank measurement is necessary to correct signals originating from reagents, solvents, and the instrument.
If the blank value is large, the measured signal includes contributions from sources other than the target component and may cause the quantitative value to be overestimated.
Therefore, appropriate blank measurement and correction are important for accurate quantitation.

Discussion of the Baseline

The baseline is the reference line in portions where no peaks are present.
The more stable the baseline is, the easier it is to accurately determine peak height and peak area.
If the baseline is sloped, wavy, or drifting, errors occur in peak reading.

Baseline disturbances may be caused by instrument stability, solvent, temperature changes, changes in the mobile phase, cell contamination, column condition, noise, and other factors.
In a report, it is useful to explain how the baseline affected the quantitative results.

Example Discussion:
If the baseline is unstable, errors occur in calculation of peak height and peak area.
Particularly in quantitation using peak area, changing the baseline position changes the area value and therefore also affects the concentration calculation.
One possible cause of the variation observed in the quantitative values in this experiment is baseline disturbance.

Discussion of Noise

Noise is unwanted fluctuation contained in the measurement signal.
If noise is large, detection of small peaks becomes difficult, and errors in reading peak height and peak area become larger.
The effect of noise is particularly important in the analysis of trace components.

Causes of noise include instrument sensitivity, electrical fluctuations, instability of the light source, temperature changes, sample turbidity, bubbles, and contamination of the cell or column.
If noise is large, stabilization of the measurement conditions and improvement of sample pretreatment are necessary.

Example Discussion:
If the measurement signal contains a large amount of noise, detection of small peaks and calculation of peak areas become difficult.
In low-concentration samples in particular, the signal from the target component is small, so the relative error caused by noise becomes large.
Therefore, to improve quantitative reliability, it is important to stabilize the instrument and remove contamination and bubbles from the cell and sample.

Discussion When Peaks Overlap

When peaks from multiple components appear at similar positions, they overlap and make accurate identification and quantitation difficult.
In chromatography, insufficient separation prevents accurate determination of peak area.
In spectral analysis, overlapping functional-group peaks or absorption bands may make assignments ambiguous.

Methods for improving peak overlap include changing measurement conditions, reviewing separation conditions, using a different wavelength or another analytical method, and comparing with standard substances.
In a report, the effect of overlap on quantitative values and assignments is discussed.

Example Discussion:
If the peak of the target component overlaps with a nearby peak, it becomes difficult to accurately determine the peak area.
As a result, the concentration of the target component may be overestimated or underestimated.
Changing the separation conditions, measuring at another wavelength, and comparing with a standard substance can improve the reliability of peak assignment and quantitation.

Errors Caused by Sample Preparation

In instrumental analysis, errors in sample preparation greatly affect the results.
If errors occur during weighing, dissolution, dilution, filtration, extraction, decomposition, pretreatment, or standard-solution preparation, the measurement signals and quantitative values shift.
Mistakes in dilution factors and standard-solution preparation in particular affect both the calibration curve and the calculated concentration of the unknown sample.

In addition, if the sample is not completely dissolved, is turbid, contains precipitate, or contains bubbles, errors occur in absorbance and peak area.
It is important to confirm sample uniformity.

Example Discussion:
Possible causes of the quantitative value differing from the expected value include errors in dilution procedures and standard-solution preparation during sample preparation.
If the concentration of the standard solution is incorrect, the entire calibration curve shifts and the concentration of the unknown sample is also calculated incorrectly.
In addition, if the sample is not completely dissolved, the actual concentration of the analyte differs from the calculated value and affects the quantitative result.

Errors Caused by Instrument Conditions

Instrument conditions also affect instrumental analysis results.
Changes in measurement wavelength, slit width, number of integrations, injection volume, flow rate, column temperature, detector sensitivity, ionization conditions, and light-source stability may change peak position, intensity, and sensitivity.

When comparing results with literature values or those of another group, it is necessary to confirm whether the instrument conditions are the same.
Even for the same sample, the results may not be completely identical if the conditions differ.

Example Discussion:
Differences in instrument conditions may explain why the measured values differed from literature values or the results of another group.
If the measurement wavelength, flow rate, column temperature, or detector sensitivity differs, peak intensity and retention time may change.
Therefore, when comparing results, not only the sample conditions but also the instrument conditions must be checked.

Causes of Quantitative Values Being Too High

Causes of quantitative values being higher than the actual value include incorrect standard-solution concentrations, insufficient blank correction, absorption by interfering components or peak overlap, concentration of the sample, calculation errors in the dilution factor, and contamination of cells or equipment.
In particular, if a coexisting component absorbs at the same wavelength as the target component or produces a peak at a similar retention time, the amount of the target component may be overestimated.

Example Discussion:
One possible reason the quantitative value was high is that components other than the target component contributed to the measurement signal.
If a coexisting substance absorbs at the same wavelength or peaks overlap in the chromatogram, the signal from the target component is overestimated.
In addition, insufficient blank correction may add signals originating from reagents or solvents and cause the calculated concentration to be too high.

Causes of Quantitative Values Being Too Low

Causes of quantitative values being lower than the actual value include sample loss, incomplete dissolution, insufficient extraction, decomposition, adsorption, excessive dilution, insufficient measurement sensitivity, and underestimation of peak area.
If the target component adsorbs to equipment during pretreatment or is partly lost during filtration, the measured value becomes lower.

In addition, if the matrix differs between the standard solutions and unknown sample, the response of the unknown sample may be lower.
In low-concentration samples, the effect of noise also becomes large.

Example Discussion:
One possible reason the quantitative value was low is that part of the target component was lost during sample pretreatment.
For example, if the target component adsorbed to equipment during filtration or transfer, the concentration in the measured solution becomes lower than the actual value.
In addition, insufficient extraction or dissolution prevents complete transfer of the target component into the measurement solution and may cause the quantitative value to be underestimated.

How to Judge the Validity of Results

The validity of instrumental-analysis results is judged by comparison with standard substances, linearity of the calibration curve, blank values, measurement reproducibility, and comparison with literature or theoretical values.
Combining multiple lines of evidence rather than relying on a single basis increases persuasiveness.

For example, if the retention time of a peak in an unknown sample agrees with that of a standard substance, the peak area lies within the calibration-curve range, and similar values are obtained upon remeasurement, the quantitative result can be considered highly reliable.
On the other hand, caution is required if peaks overlap, the value lies outside the calibration-curve range, noise is large, or reproducibility is poor.

Example Discussion:
In this experiment, the peak position of the unknown sample agreed with that of the standard substance, and the calibration curve also showed good linearity.
In addition, because the measured value of the unknown sample lay within the calibration-curve range, the quantitative result was considered generally valid.
However, because calculation of the peak area is affected by baseline setting, the quantitative value may contain a certain degree of error.

Discussion When the Result Differs From a Literature or Labeled Value

If a value obtained by instrumental analysis differs from a literature or labeled value, possible causes include sample preparation, standard-solution preparation, measurement conditions, instrument calibration, interfering components, and sample deterioration.
In addition, if the literature value is for a pure substance while the experimental sample is a mixture or commercial product, complete agreement may not occur.

When the result differs from a literature value, rather than simply writing that “there was an error,” specify which factor may have caused the measured value to become higher or lower.

Example Discussion:
Possible reasons the measured value differed from the literature value include differences in sample purity and measurement conditions.
Literature values are often obtained for pure substances under standard conditions, whereas impurities or coexisting components in the experimental sample may change peak positions or signal intensities.
Differences in instrument conditions and sample-preparation methods may also cause differences from literature values.

When the Results Can Be Considered Good

Instrumental-analysis results can be considered good when peaks corresponding to the target component are confirmed, they are consistent with standard substances or literature values, the calibration curve has good linearity, the signal of the unknown sample lies within the calibration-curve range, and the measurement is reproducible.
It is also important that noise and baseline disturbance be small and that the peaks be sufficiently separated.

In qualitative analysis, agreement of peak positions and spectral patterns is important, while in quantitative analysis, the reliability of peak area or absorbance and the validity of the calibration curve are important.
The points to emphasize should be made clear according to the purpose of the experiment.

Example Discussion:
In this experiment, a peak corresponding to the target component was clearly observed and agreed with the peak position of the standard substance.
In addition, the calibration curve prepared from the standard solutions showed good linearity, and the signal of the unknown sample was also within the calibration-curve range.
From these results, the identification and quantitation performed by instrumental analysis in this experiment were considered generally valid.

Example Discussion When the Experiment Did Not Go Well

When instrumental analysis does not go well, possible causes are considered from results such as no visible peak, a small peak, overlapping peaks, large noise, a nonlinear calibration curve, a quantitative value that deviates greatly, or poor reproducibility.
Organizing the causes separately into sample preparation, instrument conditions, standard solutions, blank, baseline, and peak processing makes the discussion easier.

Example Discussion:
In this experiment, some points on the calibration curve deviated from the straight line, reducing the reliability of the quantitative value.
Possible causes include dilution errors during preparation of the standard solutions, contamination of the cell during measurement, insufficient blank correction, and nonlinearity of the instrument response.
In addition, if the signal of the unknown sample was outside the calibration-curve range, the concentration would have been determined by extrapolation and the error may have become large.

How to Write Points for Improvement

In the discussion of an instrumental analysis experiment, including not only sources of error but also points for improvement makes the report easier to organize.
Points for improvement can be divided into sample preparation, standard-solution preparation, measurement operations, instrument conditions, and analysis methods.

Improvements to Sample Preparation

  • Dissolve the sample completely
  • Manage the dilution factor accurately
  • Perform filtration or degassing when necessary
  • Remove turbidity and precipitates
  • Prevent deterioration of the sample
  • Reduce losses during transfer

Improvements to Standard Solutions and Calibration Curves

  • Prepare standard solutions accurately
  • Set an appropriate concentration range
  • Dilute the unknown sample so that it falls within the calibration-curve range
  • Perform a blank measurement
  • Investigate the causes of outliers
  • Perform multiple measurements to confirm reproducibility

Improvements to Measurement and Analysis

  • Clean cells and equipment thoroughly
  • Remove bubbles
  • Allow the instrument to stabilize sufficiently
  • Standardize the measurement conditions
  • If peaks overlap, review the separation conditions
  • Set the baseline appropriately
  • When comparing with literature values, also check the measurement conditions

Example of How to Write Points for Improvement:
To improve quantitative accuracy, the standard solutions and unknown sample must be prepared accurately, and the unknown-sample signal must be adjusted by dilution so that it lies within the linear range of the calibration curve.
In addition, blank measurements should be performed to correct signals originating from reagents and solvents, and measurement errors can be reduced by removing contamination and bubbles from the cell.
If peaks overlap, reviewing the measurement and separation conditions can improve the accuracy of peak-area calculation.

Difference Between a Superficial Discussion and a Good Discussion

In an instrumental analysis experiment, writing only that “a peak appeared” or “the concentration was determined” results in a superficial discussion.
In a good discussion, the meaning of the peaks, basis for assignments, validity of the calibration curve, reliability of the quantitative values, and sources of error are connected.

Superficial Discussion Good Discussion
A peak appeared. Because the observed peak agreed with the peak position of the standard substance, it is considered to originate from the target component. However, it is necessary to confirm that it does not overlap with an impurity or solvent peak.
A calibration curve was prepared. Because the concentration of the standard solutions and signal intensity showed a good linear relationship, quantitation using the calibration curve was considered possible within this concentration range.
The concentration was different. Possible causes of the difference between the quantitative value and the labeled value include errors in standard-solution preparation, insufficient blank correction, interference from other components, and errors in the dilution factor.
There was noise. If noise is large, distinguishing the target signal becomes difficult in low-concentration samples, and errors in reading peak height and area may become large.

Examples of Expressions That Can Be Used in Reports

The following expressions can be used when writing the results and discussion of an instrumental analysis experiment.
Adjust the necessary parts according to your own experimental results.

  • Because the observed peak agreed with the standard substance or literature value, it is considered to originate from the target component.
  • Peak position can be used for component identification, while peak area can be used for quantitation.
  • Because the calibration curve showed linearity, signal intensity is considered proportional to concentration within this concentration range.
  • Because the signal of the unknown sample was within the calibration-curve range, the quantitative result was considered generally valid.
  • If peaks overlap, it becomes difficult to determine the peak area accurately and errors occur in the quantitative value.
  • Baseline disturbance affects the calculation of peak height and peak area.
  • If blank correction is insufficient, the concentration may be calculated including signals from sources other than the target component.
  • Errors in preparation of the standard solutions affect the entire calibration curve and also affect the quantitative value of the unknown sample.
  • The difference from the literature value may have been caused by differences in measurement conditions, sample preparation, coexisting components, or instrument calibration.
  • To improve quantitative accuracy, it is important to standardize the measurement conditions and process the standard solutions and unknown samples using the same procedure.

Points to Check When Discussing Instrumental Analysis Experiments

Checking the following points before writing the report makes the discussion easier to write.

  • Are the analytical method and analyte clearly stated?
  • Is the spectrum or chromatogram shown?
  • Are the meanings of peak position, intensity, and area explained?
  • Is there a basis for the peak assignments?
  • Has the result been compared with standard substances or literature values?
  • Are the calibration-curve equation and correlation coefficient shown?
  • Has it been confirmed that the signal of the unknown sample lies within the calibration-curve range?
  • Has the dilution factor been correctly applied?
  • Has blank correction been taken into account?
  • Have the effects of noise and baseline disturbance been considered?
  • Have the possibilities of peak overlap and interfering components been considered?
  • Do the points for improvement correspond to the sources of error?

Summary

In instrumental analysis experiments, spectra, peaks, chromatograms, calibration curves, and quantitative values are used to evaluate the components, structures, concentrations, and purity of samples.
Spectra and peaks should not simply be regarded as having “appeared” or “not appeared”; it is important to discuss which components or functional groups they originate from based on their positions, intensities, areas, and shapes.

In quantitative analysis, a calibration curve is prepared from standard solutions and the signal of the unknown sample is evaluated within that range.
The reliability of quantitative results can be judged by checking the linearity and correlation coefficient of the calibration curve, blank correction, dilution factor, and method used to calculate peak area.
If the signal of the unknown sample lies outside the calibration-curve range, dilution or remeasurement may be necessary.

In a report, rather than merely writing that “a peak was observed” or “the concentration was determined,” organize and discuss the basis for peak assignments, validity of the calibration curve, differences from literature values, noise and baseline effects, sample-preparation errors, and the influence of instrument conditions.
In instrumental analysis, it is important to convert measurement data into chemical meaning and explain what the results indicate.