GC-MS is an analytical method in which components in a mixture are separated by gas chromatography and the mass spectrum of each component is measured with a mass spectrometer.
By combining the retention time obtained by GC with the molecular-ion peaks, fragment peaks, and isotope peaks obtained by MS, it can be used to identify unknown components and confirm reaction products.
In a discussion of GC-MS, it is not sufficient simply to write that “the library search matched,” “a molecular-ion peak was observed,” or “fragments were observed.”
It is necessary to explain which peak corresponds to the molecular ion, which bonds were cleaved to produce the fragments, what the isotope peaks and base peak mean, and to what extent identification can be made from both retention time and the mass spectrum.
This article clearly explains, as examples of discussions that can be used in GC-MS laboratory reports, molecular-ion peaks and fragment analysis, retention time, library searches, isotope peaks, the concept of qualitative analysis, sources of error, and points for improvement.
Note:
This article is a reference intended to assist with discussions of GC-MS results obtained in instrumental-analysis experiments, organic-chemistry experiments, and analytical-chemistry experiments at universities and similar institutions.
For the actual ionization method, measurement mode, column conditions, library searches, fragment assignments, and safety precautions, always follow the instructions in your university’s laboratory manual and those given by your instructor or TA.
- What Is GC-MS?
- Main Items to Include in the Results
- Reference Experimental Values and Analysis Examples for GC-MS
- Reference Experimental Conditions
- Information Obtained From GC-MS
- GC-MS Measurement Results for a Standard Mixture
- Example Calculation of Relative Composition From Peak Area
- MS Fragment Analysis of Acetone
- MS Fragment Analysis of Toluene
- MS Fragment Analysis of Methyl Benzoate
- Example of Distinguishing Components With the Same Molecular Weight
- Example GC-MS Measurement of Unknown Sample A
- Area Percentages of Unknown Sample A
- Example of Interpreting Library-Search Results
- Spectral Disturbance Caused by Coelution
- Changes in Retention Time and Separation Caused by the Temperature Program
- Errors When Judging From Retention Time Alone
- Reference Example of Quantitation by the Internal-Standard Method
- Example of How to Write the Results
- Points for Connecting the Results to the Discussion
- Example Discussion
- Summary
- How to Read a Total Ion Chromatogram
- Discussion of Retention Time
- What Is a Mass Spectrum?
- What Is a Molecular-Ion Peak?
- When the Molecular-Ion Peak Is Difficult to Observe
- What Is a Base Peak?
- What Is a Fragment Peak?
- Concept of Fragment Analysis
- How to Interpret Representative Fragments
- Discussion of Isotope Peaks
- How to Interpret Compounds Containing Chlorine and Bromine
- Discussion of Library Searches
- Discussion of Mass-Spectrum Match Score
- Discussion When Peaks Overlap
- Discussion of Extracted-Ion Chromatograms
- Discussion of the EI Method
- Difference From the CI Method
- Concept of Qualitative Analysis
- Concept of Quantitative Analysis
- Relationship Between the Molecular-Ion Peak and Molecular Weight
- Fragment Analysis and Structural Estimation
- Discussion of Background Peaks
- Discussion of Solvent Peaks
- When GC-MS Results Can Be Considered Good
- Example Discussion When the Experiment Did Not Go Well
- How to Write Points for Improvement
- Difference Between a Superficial Discussion and a Good Discussion
- Examples of Expressions That Can Be Used in Reports
- Points to Check When Discussing GC-MS
- Summary
What Is GC-MS?
GC-MS is an analytical method that combines GC and MS.
First, components in the sample are separated by GC according to differences in time, and each component is introduced into the MS.
In MS, the introduced components are ionized, and the resulting molecular ions and fragment ions are detected according to their mass-to-charge ratio, or m/z.
With GC alone, components are estimated from retention time, but it may be difficult to distinguish another component with a similar retention time.
GC-MS provides a mass spectrum in addition to retention time, allowing more reliable identification based on information about molecular weight and partial structure.
Example Discussion:
In GC-MS, components can be separated by GC and the mass spectrum of each component can be obtained by MS.
Because the peaks observed in this experiment can be analyzed from both retention time and mass spectrum, the reliability of component identification is higher than with GC alone.
In particular, molecular-ion peaks and characteristic fragment peaks provide important clues for estimating the molecular weight and structure of compounds.
Main Items to Include in the Results
In GC-MS results, organize the retention time of the chromatogram, peak area, mass spectrum of each peak, molecular-ion peak, base peak, major fragment peaks, isotope peaks, library-search results, and other information.
Because GC and MS conditions affect the identification results, the measurement conditions should also be clearly stated.
Main Items to Include in the Results
- Sample name
- Analyte
- Column used
- Carrier gas
- Column temperature conditions
- Inlet temperature
- Injection volume
- Ionization method
- Measurement mode
- Measured m/z range
- Retention time
- Total ion chromatogram
- Peak area
- Molecular-ion peak
- Base peak
- Major fragment peaks
- Isotope peaks
- Library-search results
- Candidate compounds
- Basis for identification
- Sources of error and points for improvement
Example of How to Write the Results:
Multiple peaks were observed in the total ion chromatogram obtained by GC-MS measurement.
The retention time and mass spectrum of each peak were examined, and candidate compounds were estimated from the molecular-ion peak and major fragment peaks.
In addition, the library-search results were compared with the retention times of standard substances to identify the target components.
Reference Experimental Values and Analysis Examples for GC-MS
Here, the process of identifying components in a mixed sample using retention time, molecular-ion peaks, fragment peaks, and library-search results obtained by GC-MS is organized using reference experimental values.
In GC-MS, a mixture is separated into individual components by gas chromatography (GC), and the mass spectrum of each peak is measured by mass spectrometry (MS).
GC retention time reflects the volatility of each component and its interaction with the column, while molecular-ion peaks and fragment peaks in MS provide clues to molecular weight and structural estimation.
Reference Experimental Conditions
| Item | Details |
|---|---|
| Measurement target | Standard mixture, fragrance-component mixture, unknown sample |
| Measurement method | GC-MS |
| Column | Nonpolar capillary column |
| Carrier gas | Helium |
| Ionization method | EI method, 70 eV |
| Measurement range | m/z 30–300 |
| Evaluation items | Retention time, peak area, molecular ion, base peak, fragments, library-match score |
Information Obtained From GC-MS
| Information | Main Meaning | Use in Analysis |
|---|---|---|
| Retention time | Time at which a component elutes from the GC column | Comparison with standard substances or library information |
| Peak area | Approximate amount of the detected component | Used to compare relative amounts of components |
| Molecular-ion peak | Clue to molecular weight | Confirmation of the molecular weight of candidate compounds |
| Fragment peak | Ion formed when a molecule fragments | Estimation of partial structures |
| Library-match score | Similarity to a standard spectrum | Assistance in component identification |
GC-MS Measurement Results for a Standard Mixture
A reference example is shown for measurement of a standard mixture containing acetone, toluene, ethyl acetate, methyl benzoate, and limonene.
| Peak | Retention Time | Estimated Component | Molecular Formula | Molecular Weight | Molecular Ion | Base Peak | Library Match |
|---|---|---|---|---|---|---|---|
| 1 | 2.15 min | Acetone | C3H6O | 58 | m/z 58 | m/z 43 | 96% |
| 2 | 3.42 min | Ethyl acetate | C4H8O2 | 88 | m/z 88 | m/z 43 | 94% |
| 3 | 5.80 min | Toluene | C7H8 | 92 | m/z 92 | m/z 91 | 98% |
| 4 | 9.65 min | Limonene | C10H16 | 136 | m/z 136 | m/z 68 | 93% |
| 5 | 12.30 min | Methyl benzoate | C8H8O2 | 136 | m/z 136 | m/z 105 | 97% |
Even limonene and methyl benzoate, which have the same molecular weight of 136, can be distinguished because their retention times and fragment peaks differ.
In GC-MS, it is important to combine retention time and mass spectrum.
Example Calculation of Relative Composition From Peak Area
The relative amount of each component in a mixture can be determined from the peak areas in the GC chromatogram.
Here, the calculation is performed as a simple area percentage without correcting for differences in sensitivity.
| Peak | Estimated Component | Peak Area | Area Percentage | How to Interpret the Result |
|---|---|---|---|---|
| 1 | Acetone | 125000 | 8.3% | Minor component |
| 2 | Ethyl acetate | 210000 | 14.0% | Moderate |
| 3 | Toluene | 365000 | 24.3% | One of the major components |
| 4 | Limonene | 480000 | 32.0% | Largest amount |
| 5 | Methyl benzoate | 320000 | 21.3% | One of the major components |
| Total | – | 1500000 | 100.0% | – |
The area percentage is calculated using the following equation.
Area percentage (%) = Area of each peak ÷ Total peak area × 100
If the peak area of limonene is 480000 and the total peak area is 1500000,
Area percentage = 480000 ÷ 1500000 × 100 = 32.0%
However, because ionization efficiency and detection sensitivity differ among components, the area percentage cannot be regarded directly as a mass percentage.
MS Fragment Analysis of Acetone
| m/z | Relative Intensity | Interpretation | Structural Clue |
|---|---|---|---|
| 58 | 35 | Molecular ion M+ | Molecular weight 58 |
| 43 | 100 | CH3CO+ | Characteristic of carbonyl compounds |
| 15 | 22 | CH3+ | Originating from a methyl group |
| 29 | 10 | CHO+, etc. | Oxygen-containing fragment |
The molecular-ion peak at m/z 58 indicates a molecular weight of 58, while the strong peak at m/z 43 suggests a compound containing a carbonyl group.
MS Fragment Analysis of Toluene
| m/z | Relative Intensity | Interpretation | Structural Clue |
|---|---|---|---|
| 92 | 70 | Molecular ion M+ | Molecular weight 92 |
| 91 | 100 | C7H7+ | Benzyl / tropylium ion |
| 65 | 20 | C5H5+ | Originating from an aromatic ring |
| 39 | 12 | C3H3+ | Small hydrocarbon ion |
Because m/z 91 appears strongly as the base peak, it provides a clue for considering the presence of a benzyl structure or alkylbenzene.
MS Fragment Analysis of Methyl Benzoate
| m/z | Relative Intensity | Interpretation | Structural Clue |
|---|---|---|---|
| 136 | 65 | Molecular ion M+ | Molecular weight 136 |
| 105 | 100 | C6H5CO+ | Benzoyl cation |
| 77 | 45 | C6H5+ | Phenyl cation |
| 31 | 10 | OCH3+ | Originating from a methoxy group |
The difference from m/z 136 to m/z 105 is 31, which can be considered to correspond to loss of an OCH3 group.
The strong peak at m/z 105 provides a clue indicating a compound with an aromatic carbonyl structure.
Example of Distinguishing Components With the Same Molecular Weight
In GC-MS, even components having the same molecular weight can be distinguished from differences in retention time and fragmentation pattern.
Limonene and methyl benzoate both have a molecular weight of 136, but their spectra differ greatly.
| Item | Limonene | Methyl Benzoate | Point of Distinction |
|---|---|---|---|
| Molecular weight | 136 | 136 | Cannot be distinguished by molecular weight alone |
| Retention time | 9.65 min | 12.30 min | Behavior on the column differs |
| Base peak | m/z 68 | m/z 105 | Fragments differ |
| Characteristic peaks | m/z 93, 68, 67 | m/z 105, 77, 31 | Reflect differences in structure |
| Structural characteristics | Terpene hydrocarbon | Aromatic ester | Functional groups differ |
Even when the molecular-ion peaks are the same, the components can be judged to be different because the retention times and base peaks differ.
Example GC-MS Measurement of Unknown Sample A
A reference example is shown in which three major peaks were observed when unknown sample A was measured by GC-MS.
| Peak | Retention Time | Peak Area | Major m/z | Molecular Ion | Library Candidate | Match |
|---|---|---|---|---|---|---|
| 1 | 5.81 min | 180000 | 91, 92, 65 | m/z 92 | Toluene | 97% |
| 2 | 9.66 min | 420000 | 68, 93, 67, 136 | m/z 136 | Limonene | 94% |
| 3 | 12.28 min | 250000 | 105, 136, 77, 31 | m/z 136 | Methyl benzoate | 96% |
Unknown sample A may contain toluene, limonene, and methyl benzoate.
In particular, peaks 2 and 3 both have a molecular ion at m/z 136, but because their retention times and fragments differ, they can be judged to be different components.
Area Percentages of Unknown Sample A
| Peak | Estimated Component | Peak Area | Area Percentage | How to Interpret the Result |
|---|---|---|---|---|
| 1 | Toluene | 180000 | 21.2% | Moderate |
| 2 | Limonene | 420000 | 49.4% | Main component |
| 3 | Methyl benzoate | 250000 | 29.4% | Secondary component |
| Total | – | 850000 | 100.0% | – |
From the area percentages, limonene was detected as the largest peak in unknown sample A.
However, because GC-MS peak area is affected by differences in sensitivity among components, a calibration curve or internal-standard method is necessary to determine accurate concentrations.
Example of Interpreting Library-Search Results
In GC-MS, the measured spectrum can be compared with a database to display candidate compounds.
However, even when a candidate with a high match score is obtained, it is important to confirm the result together with retention time and comparison with a standard substance.
| Unknown Peak | Retention Time | Candidate 1 | Match | Candidate 2 | Match | Judgment |
|---|---|---|---|---|---|---|
| Peak 1 | 5.81 min | Toluene | 97% | Ethylbenzene | 82% | Toluene is more likely |
| Peak 2 | 9.66 min | Limonene | 94% | Pinene | 76% | Limonene is more likely |
| Peak 3 | 12.28 min | Methyl benzoate | 96% | Ethyl benzoate | 70% | Methyl benzoate is more likely |
Library-match score is a powerful clue for component identification, but misidentification may occur because of coelution, noise, or isomers with similar spectra.
Spectral Disturbance Caused by Coelution
Coelution refers to two components that are not sufficiently separated and are detected overlapping near the same retention time.
When coelution occurs, the mass spectrum becomes a mixture of multiple components and the library-match score may decrease.
| Condition | Retention Time | Major Observed m/z | Library Match | How to Interpret the Result |
|---|---|---|---|---|
| Sufficiently separated | 9.65 min | 68, 93, 67, 136 | 94% | Easy to identify as limonene |
| Mild coelution | 9.62–9.80 min | 68, 93, 105, 136 | 78% | Peaks from another component are mixed in |
| Strong coelution | 9.60–9.90 min | 68, 91, 105, 136, 77 | 55% | Difficult to analyze as a single component |
If coelution is suspected, the separation must be improved by changing the column conditions or temperature-program conditions.
Changes in Retention Time and Separation Caused by the Temperature Program
In GC, retention time and peak separation change depending on the oven temperature program.
If the temperature is increased too rapidly, the analysis time becomes shorter, but peaks may overlap more easily.
| Temperature Program | Limonene Retention Time | Methyl Benzoate Retention Time | Approximate Resolution | How to Interpret the Result |
|---|---|---|---|---|
| 5°C/min | 12.80 min | 16.40 min | Good | Good separation but long analysis time |
| 10°C/min | 9.65 min | 12.30 min | Good | Standard conditions |
| 20°C/min | 6.40 min | 7.55 min | Slightly reduced | Short analysis time |
| 40°C/min | 4.20 min | 4.65 min | Insufficient | Coelution is likely |
In GC-MS component identification, it is necessary to check not only the mass spectrum but also whether the components are sufficiently separated in the chromatogram.
Errors When Judging From Retention Time Alone
Retention time is a clue for component identification, but it is risky to identify a component from retention time alone.
Different compounds may show similar retention times.
| Candidate Component | Retention Time | Molecular Ion | Base Peak | Point for Judgment |
|---|---|---|---|---|
| Limonene | 9.65 min | m/z 136 | m/z 68 | Terpene-derived fragments |
| p-Cymene | 9.70 min | m/z 134 | m/z 119 | Originating from an aromatic hydrocarbon |
| Benzaldehyde | 9.58 min | m/z 106 | m/z 77 | Originating from an aromatic aldehyde |
Even if retention times are close, compounds can be judged to be different if their molecular ions or fragments differ.
Therefore, it is important to examine both retention time and the MS spectrum.
Reference Example of Quantitation by the Internal-Standard Method
For more accurate quantitation, an internal standard is added and the peak-area ratio is used.
Here, a reference example using ethylbenzene as the internal standard is shown.
| Sample | Limonene Area | Internal-Standard Area | Area Ratio | Concentration Determined From Calibration Curve |
|---|---|---|---|---|
| Standard 1 | 120000 | 200000 | 0.60 | 10 mg/L |
| Standard 2 | 245000 | 205000 | 1.20 | 20 mg/L |
| Standard 3 | 480000 | 200000 | 2.40 | 40 mg/L |
| Unknown sample | 360000 | 200000 | 1.80 | 30 mg/L |
The internal-standard method can compensate to some extent for variations in injection volume and instrument sensitivity.
It is effective when greater quantitative accuracy than simple area percentage is required.
Example of How to Write the Results
When the standard mixture was measured by GC-MS, peaks were observed at retention times of 2.15, 3.42, 5.80, 9.65, and 12.30 min.
Comparison of the mass spectrum of each peak with the library suggested acetone, ethyl acetate, toluene, limonene, and methyl benzoate.
For toluene, a molecular-ion peak was observed at m/z 92 and a base peak at m/z 91.
m/z 91 corresponds to a benzyl cation or tropylium ion and is a fragment characteristic of alkylbenzenes.
Because these spectral characteristics and the retention time agreed with those of the standard substance, peak 3 was judged to be toluene.
In unknown sample A, a molecular-ion peak at m/z 136 and a base peak at m/z 68 were observed for the peak at a retention time of 9.66 min.
On the other hand, a molecular-ion peak at m/z 136 was also observed for the peak at a retention time of 12.28 min, but the base peak was m/z 105.
Therefore, although the two components have the same molecular weight, they have different fragmentation patterns and are likely to be limonene and methyl benzoate, respectively.
Points for Connecting the Results to the Discussion
In a GC-MS discussion, it is important to interpret retention time, peak area, molecular ions, fragments, and library-match scores together.
- Can the separation of components be confirmed from retention times?
- Can the molecular weight be estimated from the molecular-ion peak?
- Can the base peak and fragment peaks be explained in relation to the structure?
- Can compounds with the same molecular weight be distinguished from differences in retention time and fragments?
- Can the result be judged together with standard substances and retention time rather than being determined solely from the library-match score?
- Can it be explained that peak-area percentage is not necessarily the same as exact mass percentage?
- Can it be discussed that coelution produces mixed spectra and lowers the library-match score?
- Can it be explained that temperature-program and column conditions affect retention time and separation?
- Can it be explained that the internal-standard method can compensate for variation in injection volume and sensitivity?
Example Discussion
In this experiment, GC-MS was used to separate and identify components in a mixed sample.
Multiple peaks were observed in the GC chromatogram, and the mass spectrum at each retention time was analyzed.
Peak 1 showed a retention time of 2.15 min, a molecular ion at m/z 58, and a base peak at m/z 43, which agreed with the spectrum of acetone.
m/z 43 corresponds to CH3CO+ and is a fragment characteristic of carbonyl compounds.
For peak 3, a molecular-ion peak at m/z 92 and a base peak at m/z 91 were observed.
m/z 91 is stabilized as a benzyl cation or tropylium ion and therefore appears strongly in alkylbenzenes such as toluene.
Because the retention time and library-match score also agreed well with toluene, this peak was judged to be toluene.
In unknown sample A, two peaks having a molecular ion at m/z 136 were observed.
One had a base peak at m/z 68 and showed fragments characteristic of limonene.
The other had a base peak at m/z 105, corresponding to a benzoyl cation, and was therefore considered to be methyl benzoate.
These results show that even compounds having the same molecular weight can be distinguished by combining retention time and fragmentation pattern.
When the area percentages were calculated, unknown sample A contained 49.4% limonene, 29.4% methyl benzoate, and 21.2% toluene.
However, GC-MS peak area is affected not only by component concentration but also by ionization efficiency and detection sensitivity.
Therefore, area percentage is only a relative guide, and accurate quantitation requires a calibration curve or internal-standard method.
Possible sources of error include coelution, variation in injection volume, deterioration of the column, inappropriate temperature-program conditions, and misidentification in library searches.
When coelution occurs, the spectrum contains fragments originating from multiple components and the library-match score decreases.
In addition, if the temperature ramp is too fast, the retention time becomes shorter but separation between peaks may become poorer.
Therefore, in GC-MS analysis, it is necessary to examine not only the mass spectrum but also the separation state in the chromatogram.
Summary
In GC-MS, mixtures can be separated by GC and the molecular weight and fragments of each component can be examined by MS.
Combining retention time, molecular-ion peaks, base peaks, fragments, and library-match scores improves the reliability of component identification.
This reference example dealt with component identification, area percentage, distinction between components with the same molecular weight, coelution, and the internal-standard method using a standard mixture and unknown sample A.
In a report, it is useful to discuss retention time, molecular ions, fragment analysis, library searches, peak separation, and precautions for quantitation in relation to one another.
How to Read a Total Ion Chromatogram
In GC-MS, components separated by GC are introduced into the MS over time, and the total intensity of ions detected at each time is recorded.
This graph is called the total ion chromatogram, or TIC.
In a TIC, retention time is plotted on the horizontal axis and total ion intensity on the vertical axis.
Each peak in the TIC represents a component introduced into the MS at a particular time.
Selecting each peak makes it possible to examine the mass spectrum corresponding to that retention time.
Therefore, in GC-MS, components are estimated by combining the peak positions in the TIC with the mass spectrum of each peak.
Example Discussion:
Because multiple peaks were observed in the total ion chromatogram, the sample was considered to contain multiple volatile components.
The retention time of each peak provides separation information from GC, while the mass spectrum at that retention time provides information about the molecular weight and structure of the component.
Therefore, in GC-MS, it is important to identify components by combining retention time and mass spectrum.
Discussion of Retention Time
Retention time is important in GC-MS just as it is in GC analysis.
Retention time is the time required for a component to move through the column and reach the detector.
It changes depending on the volatility of the component, interaction with the stationary phase, column temperature, carrier-gas flow rate, and type of column.
If the retention time agrees with that of a standard substance, it becomes one piece of evidence for identification.
However, retention time alone may not be sufficient to distinguish another component.
In GC-MS, the reliability of identification is improved by confirming agreement of the mass spectrum in addition to retention time.
Example Discussion:
Because the retention time of the unknown peak was close to that of the standard substance, the peak may have originated from the target component.
However, because another component may have a similar retention time, identification cannot be confirmed from retention time alone.
In this experiment, the validity of the identification was evaluated by checking the molecular-ion peak and fragment peaks in the mass spectrum in addition to the retention time.
What Is a Mass Spectrum?
A mass spectrum is a graph showing the intensity of ionized molecules and fragment ions at each m/z.
m/z is plotted on the horizontal axis and relative intensity on the vertical axis.
Normally, the strongest peak is set to 100% and the intensities of the other peaks are shown relative to it.
A mass spectrum contains molecular-ion peaks, base peaks, fragment peaks, and isotope peaks.
From this information, molecular weight, the presence of elements, partial structures, and the ease of bond cleavage can be estimated.
Particularly in the EI method, molecules undergo relatively extensive fragmentation, producing many peaks useful for structural analysis.
Example Discussion:
In a mass spectrum, molecules or fragments are ionized and detected as peaks at each m/z.
Among the observed peaks, the molecular-ion peak can be used to estimate molecular weight, while fragment peaks can be used to estimate partial structures.
Therefore, in GC-MS, information about the structure of an unknown component can be obtained by analyzing the peak pattern of the mass spectrum.
What Is a Molecular-Ion Peak?
A molecular-ion peak is a peak corresponding to a molecule that has lost an electron and become ionized.
In general, the m/z of the molecular-ion peak corresponds to the molecular weight.
For example, if a compound has a molecular weight of 100 and a charge of +1, the molecular-ion peak appears near m/z 100.
However, the molecular-ion peak does not always appear clearly for every compound.
In the EI method, molecules may undergo extensive fragmentation, so the molecular-ion peak may be small or difficult to observe.
If the molecular-ion peak cannot be seen, library searches, fragment peaks, and results obtained by soft-ionization methods are used as references.
Example Discussion:
If a peak is observed in the mass spectrum at an m/z considered to correspond to the molecular weight, this peak may be the molecular-ion peak.
The molecular-ion peak is an important clue for estimating the molecular weight of a compound.
However, because molecules readily fragment in the EI method, the molecular-ion peak may become weak and does not necessarily appear as the strongest peak.
When the Molecular-Ion Peak Is Difficult to Observe
Causes of a molecular-ion peak being difficult to observe include the compound readily decomposing during ionization, instability of the molecular ion, extensive fragmentation, the measured m/z range, insufficient concentration, and effects of background.
In compounds such as alcohols, ethers, esters, and compounds with branched structures, the molecular-ion peak may become weak.
If the molecular-ion peak is not clear, it is important not to assume that the peak at the highest m/z is the molecular ion.
The result should be judged by combining library searches, isotope peaks, consistency of fragments, and comparison with standard substances.
Example Discussion:
In this experiment, it was difficult to confirm a clear molecular-ion peak corresponding to the molecular weight.
This was considered to be because the molecular ion became unstable during EI ionization and immediately fragmented.
In such cases, the component must be identified comprehensively from the major fragment peaks, library-search results, and comparison with standard substances rather than determining the molecular weight from the molecular-ion peak alone.
What Is a Base Peak?
The base peak is the peak with the greatest intensity in a mass spectrum.
Normally, this peak is assigned a relative intensity of 100%, and the intensities of the other peaks are expressed relative to it.
The base peak often corresponds to the stable ion produced in the greatest amount and reflects the fragmentation characteristics of the compound.
The base peak is not necessarily the molecular-ion peak.
If a fragment ion more stable than the molecular ion is generated in large amounts, that fragment becomes the base peak.
Therefore, examining the m/z of the base peak makes it possible to consider which partial structure readily remains as a stable ion.
Example Discussion:
The strongest peak in the mass spectrum is the base peak and is considered to correspond to the stable ion produced in the greatest amount.
If the base peak is a fragment peak rather than the molecular-ion peak, this indicates that the fragment ion is particularly stable.
Therefore, the m/z of the base peak provides an important clue for considering partial structures and bonds that are readily cleaved in the compound.
What Is a Fragment Peak?
A fragment peak is a peak corresponding to a smaller ion formed when the molecular ion undergoes bond cleavage.
In the EI method, molecules are ionized by high-energy electrons, so the molecular ion may undergo further decomposition and produce many fragment peaks.
Fragment peaks indicate which bonds in the molecule are readily cleaved and which partial structures readily form stable ions.
Therefore, fragment analysis is important for structural estimation of unknown compounds and for confirming the validity of library-search results.
Example Discussion:
The observed fragment peaks were considered to originate from ions formed when the molecular ion cleaved at specific bonds.
The m/z of fragment peaks provides clues to the presence of partial structures and functional groups in the molecule.
Therefore, by analyzing not only the molecular-ion peak but also the combination of major fragment peaks, it is possible to confirm whether the structure of the candidate compound is valid.
Concept of Fragment Analysis
In fragment analysis, the portion lost from the molecular-ion peak is considered.
For example, a peak 15 mass units lower than the molecular ion may correspond to loss of a methyl group, a peak 18 lower may correspond to loss of water, and a peak 28 lower may correspond to loss of CO or ethylene.
However, because actual fragments depend on the structure of the compound and the stability of the ions, it is important not to make a definitive assignment from the mass difference alone.
Fragment ions tend to appear strongly when they have structures stable as carbocations, resonance-stabilized structures, or structures containing aromatic rings.
Therefore, the m/z of the peak is compared with the structural formula of the compound to estimate which bond was cleaved.
Example Discussion:
Because a fragment peak having a specific mass difference from the molecular-ion peak was observed, part of the molecule may have been lost.
For example, a peak 15 mass units lower than the molecular ion may correspond to loss of a methyl group.
However, because fragmentation depends on ion stability and molecular structure, it is necessary to consider not only the mass difference but also whether the resulting ion can exist stably.
How to Interpret Representative Fragments
In mass spectra, characteristic fragments may be observed depending on the type of compound.
In alkyl chains, fragment peaks corresponding to the number of carbon atoms are likely to appear, while in aromatic compounds, peaks corresponding to benzyl cations or tropylium ions may be important.
Alcohols may undergo dehydration, while esters and ketones may show cleavage around the carbonyl group.
However, representative fragments are only clues for discussion.
In actual peak assignment, the molecular formula, molecular-ion peak, isotope peaks, and agreement with library spectra are checked.
| Observed Characteristic | Possible Meaning | Point for Discussion |
|---|---|---|
| Peak 15 lower than the molecular ion | Possibility of methyl-group loss | Check whether the structure contains a methyl group |
| Peak 18 lower than the molecular ion | Possibility of water loss | Check whether the structure readily undergoes dehydration, such as an alcohol |
| Stable aromatic-derived peak | Possibility of a fragment containing an aromatic ring | Compare the presence of an aromatic ring with IR and NMR |
| Strong low-m/z peak | Stable small fragment | May reflect structural characteristics as the base peak |
Example Discussion:
Because multiple peaks smaller than the molecular ion were observed in the mass spectrum, the molecular ion was considered to have fragmented through several pathways.
These peaks may correspond to loss of alkyl groups, cleavage around functional groups, or formation of stable ions containing aromatic rings.
In fragment analysis, it is important to confirm whether the observed m/z values can be reasonably explained from the structural formula.
Discussion of Isotope Peaks
Isotope peaks are peaks originating from molecules or fragments containing naturally occurring isotopes.
For example, because carbon contains a certain proportion of 13C, an M+1 peak may appear one mass unit above the molecular-ion peak.
Compounds containing chlorine or bromine show characteristic M and M+2 peak patterns.
Isotope peaks are useful for estimating elemental composition.
Particularly in compounds containing chlorine or bromine, the intensity ratio of isotope peaks is characteristic and provides an important clue for discussing the presence of halogen elements.
Example Discussion:
Because M+1 and M+2 peaks were observed near the molecular-ion peak, ions containing naturally occurring isotopes were considered to have been detected.
The M+1 peak may mainly arise from the presence of 13C.
In addition, a characteristic intensity ratio between M and M+2 provides evidence for the possible presence of halogen elements such as chlorine or bromine.
How to Interpret Compounds Containing Chlorine and Bromine
Isotope peaks are extremely important for compounds containing chlorine or bromine.
Chlorine has 35Cl and 37Cl, and the molecular-ion peak and M+2 peak appear in a characteristic ratio.
Bromine has 79Br and 81Br, and M and M+2 may appear with nearly equal intensities.
Such isotope patterns are useful for confirming the validity of library-search results.
If a candidate compound contains chlorine or bromine, the measured spectrum should be checked for the corresponding isotope peaks.
Example Discussion:
If M and M+2 peaks are observed near the molecular ion in a characteristic ratio, the compound may contain chlorine or bromine.
In particular, compounds containing bromine may show M and M+2 peaks at nearly equal intensities.
Confirming whether this isotope pattern agrees with the elemental composition of the candidate compound improves the reliability of identification.
Discussion of Library Searches
In GC-MS, the obtained mass spectrum can be compared with spectra registered in a database to search for candidate compounds.
Library searches are useful for identifying unknown components, but it is risky to treat the search result itself as definitive identification.
This is because another compound having a similar fragmentation pattern may appear as a candidate.
When discussing library-search results, similarity, agreement of the molecular-ion peak, agreement of major fragment peaks, isotope pattern, retention time, and comparison with standard substances are checked.
If possible, a standard substance is measured to confirm whether both retention time and mass spectrum agree.
Example Discussion:
A candidate compound was suggested by the library search, but identification cannot be confirmed from the search result alone.
It is necessary to confirm whether the molecular-ion peak, major fragment peaks, and isotope peaks of the candidate compound agree with the measured spectrum.
In addition, if the retention time and mass spectrum agree with those of a standard substance, the reliability of identification becomes even higher.
Discussion of Mass-Spectrum Match Score
In a library search, the degree of agreement between the measured spectrum and the database spectrum may be shown numerically.
The higher the match score, the stronger the candidate compound, but the score changes depending on measurement conditions, background, peak overlap, concentration, and ionization conditions.
Even when the match score is high, caution is required if the molecular-ion peak or major fragments cannot be explained.
Conversely, even if the match score is somewhat low, agreement of retention time and major peaks with a standard substance can provide evidence for identification.
It is important to examine the content of the peaks rather than relying only on the numerical value.
Example Discussion:
Because the library-search match score was high, the candidate compound is likely to be the component of the unknown peak.
However, the match score is an indicator of the overall similarity of the spectra and does not confirm identification.
It is necessary to confirm that the major fragment peaks, molecular-ion peak, isotope peaks, and retention time do not contradict the candidate compound.
Discussion When Peaks Overlap
If separation by GC is insufficient, multiple components may elute near the same retention time and appear as a single peak in the TIC.
In this case, the mass spectrum at that retention time contains ions originating from multiple components, causing library-search results to become unclear or leading to misidentification.
If peak overlap is suspected, it can be checked by examining an extracted-ion chromatogram, changing the column-temperature conditions, using another column, or adding a standard substance.
Tracking only a specific m/z may make it possible to distinguish overlapping components.
Example Discussion:
Even if a peak appears as a single peak in the TIC, if multiple components overlap, the obtained mass spectrum contains fragments originating from several components.
As a result, the library-search match score may decrease or an incorrect candidate compound may be suggested.
If peak overlap is suspected, the separation state must be checked using an extracted-ion chromatogram or by changing the measurement conditions.
Discussion of Extracted-Ion Chromatograms
An extracted-ion chromatogram, or EIC, is a chromatogram that extracts only a specific m/z and displays its change over time.
In a TIC, signals from multiple components are summed, whereas in an EIC, attention can be focused on a specific ion, making it useful for checking peak overlap and trace components.
By selecting an m/z characteristic of the target component, it is possible to confirm at what retention time that component appears.
However, because a fragment with the same m/z may also be generated from another component, it is important not to confirm identification using only one m/z.
Example Discussion:
By using an extracted-ion chromatogram, the time dependence of an m/z characteristic of the target component can be examined.
Even a trace component hidden by signals from other components in the TIC may become easier to detect by extracting a characteristic m/z.
However, because fragments with the same m/z may also be produced from other components, multiple characteristic ions must be used in combination for confirmation.
Discussion of the EI Method
The EI method commonly used in GC-MS is electron ionization.
Because molecules are ionized by collision with high-energy electrons, many fragment ions are generated in addition to molecular ions.
Therefore, EI spectra contain abundant structural information and are well suited to library searches.
On the other hand, molecular-ion peaks may become weak in the EI method.
When molecular weight must be confirmed, it is important to check whether the molecular-ion peak is visible.
If the molecular ion is difficult to confirm, a soft-ionization method such as CI may be effective.
Example Discussion:
In the EI method, molecules are ionized by high-energy electrons, so many fragment peaks are observed.
These fragment peaks are useful for structural analysis and library searches.
On the other hand, in compounds with unstable molecular ions, the molecular-ion peak may become weak, so care is required when confirming molecular weight.
Difference From the CI Method
The CI method is chemical ionization.
It is often treated as a gentler ionization method than EI and may make it easier to obtain information about molecular weight.
In CI, the molecule may be observed not as the molecular ion itself but as a protonated ion or an adduct ion originating from the reagent gas.
The EI method provides abundant fragment information and is suitable for library searches.
The CI method may be advantageous for molecular-weight estimation.
Which method is appropriate depends on the purpose of the analysis.
Example Discussion:
In the EI method, many fragment peaks are obtained, making it useful for structural estimation and library searches.
On the other hand, in the CI method, ions with masses close to that of the molecule are more readily observed and may help confirm molecular weight.
If the molecular-ion peak is difficult to confirm by EI, confirmation using a soft-ionization method such as CI can be effective.
Concept of Qualitative Analysis
In qualitative analysis by GC-MS, components are estimated by comprehensively considering retention time, molecular-ion peaks, fragment peaks, isotope peaks, and library-search results.
Rather than making a definitive judgment from a single piece of evidence, it is important to confirm that multiple pieces of information agree with the candidate compound.
In particular, if a standard substance can be measured, it is checked whether the retention time and mass spectrum agree between the standard substance and the unknown sample.
If no standard substance is available, it is appropriate to describe the result as “tentative identification” based on library-search results and fragment analysis.
Example Discussion:
In this experiment, the unknown component was estimated by comprehensively considering the retention time, molecular-ion peak, major fragment peaks, and library-search results.
If the retention time and mass spectrum agree with those of a standard substance, the reliability of identification is high.
On the other hand, if no confirmation with a standard substance is performed, it is appropriate to treat the result as tentative identification based on the library search.
Concept of Quantitative Analysis
GC-MS may also be used for quantitation using peak area or the area of a specific ion.
In some cases, the TIC peak area is used, while in others an m/z characteristic of the target component is selected and the area of the extracted-ion peak is used.
For quantitation, a calibration curve is prepared from standard solutions and the peak area of the unknown sample is substituted into the calibration curve.
Because the MS response differs depending on the component and ionization efficiency, a calibration curve using a standard substance is important.
If there are matrix effects or variation in injection volume, the internal-standard method is effective.
Particularly in quantitative analysis, not only library searches for identification but also reproducibility of peak areas and linearity of the calibration curve must be checked.
Example Discussion:
In GC-MS quantitation, the peak area corresponding to the target component or the extracted-ion area of a characteristic m/z can be used.
If a calibration curve prepared from standard solutions shows linearity, the unknown sample can be quantified within that range.
However, because ionization efficiency and matrix effects differ among components, correction using standard substances or an internal standard is important for accurate quantitation.
Relationship Between the Molecular-Ion Peak and Molecular Weight
If a molecular-ion peak is clearly observed, the molecular weight can be estimated from its m/z.
In ordinary GC-MS, ions are often detected as singly charged positive ions, so m/z is close to the molecular weight.
Isotope peaks such as M+1 and M+2 may also appear near the molecular-ion peak.
When considering a molecular-weight candidate, it is necessary to confirm whether the molecular-ion peak is really the molecular ion, whether there is a small peak at a higher m/z, and whether the relationship with the isotope peaks is natural.
If a fragment peak is mistaken for the molecular ion, the molecular weight will be underestimated.
Example Discussion:
Because a peak was observed at an m/z corresponding to the molecular weight of the candidate compound, this peak may be the molecular-ion peak.
The molecular-ion peak is an important basis for molecular-weight estimation, but care must be taken not to mistake a fragment peak for the molecular ion.
Confirming the consistency of isotope peaks and major fragments improves the reliability of molecular-weight estimation.
Fragment Analysis and Structural Estimation
In fragment analysis, the structural formula of the candidate compound is used to consider which bonds would have to break to produce the observed m/z ions.
Characteristic cleavage may occur at positions where stable cations can readily form, near heteroatoms, at benzylic positions, and around carbonyl groups.
If fragment peaks can be naturally explained from the structural formula, the validity of the candidate compound increases.
Conversely, if the major peaks cannot be explained from the candidate structure, possibilities such as an incorrect library candidate, overlapping peaks, or contamination by background must be considered.
Example Discussion:
The observed major fragment peaks can be explained by cleavage of specific bonds in the structure of the candidate compound.
In particular, cleavage at positions adjacent to an aromatic ring or heteroatom tends to produce relatively stable ions.
The fact that the major fragments do not contradict the candidate structure supports the validity of the library-search result.
Discussion of Background Peaks
In GC-MS, peaks originating from background rather than sample components may also be observed.
Possible causes include column bleed, solvent, air, moisture, silicone contamination, sample containers, septa, and insufficient cleaning.
It is important not to mistake these peaks for target components.
To check the effects of background, a blank measurement is performed and compared with the sample measurement.
If the same peak also appears in the blank, it may not originate from the sample.
In trace analysis, the effect of background becomes particularly large.
Example Discussion:
Background-derived contamination may be responsible for peaks that cannot be explained from the sample structure.
Components originating from solvents, column bleed, septa, or sample containers may be detected by GC-MS.
Therefore, it is important to perform a blank measurement and check whether peaks having the same retention time or m/z as those in the sample are also present in the blank.
Discussion of Solvent Peaks
In GC-MS, the sample solvent may be observed as a large peak.
If the solvent peak is too large, it may obscure components eluting nearby or place a load on the MS detector.
Solvent-derived fragments may also become mixed into the spectrum.
The solvent is selected by considering sample solubility, volatility, compatibility with the column, and overlap of retention time with the target component.
If the solvent peak overlaps with the target component, it is necessary to change the solvent or adjust the temperature conditions.
Example Discussion:
If a large solvent peak is observed in the chromatogram, solvent-derived signals may interfere with detection of nearby target components.
In addition, if solvent-derived fragments become mixed into the mass spectrum, they may affect the library-search results.
Therefore, it is important to select conditions under which the target component and solvent peak do not overlap.
When GC-MS Results Can Be Considered Good
GC-MS results can be considered good when the peak of the target component is clearly separated in the TIC, its retention time agrees with that of a standard substance or the expected value, molecular-ion peaks and characteristic fragment peaks are observed in the mass spectrum, and the results are also consistent with the library search.
In addition, if the isotope-peak pattern agrees with the elemental composition of the candidate compound and the major fragments can be explained from the structural formula, the reliability of identification becomes higher.
Measuring a standard substance and confirming agreement of both retention time and spectrum provides even stronger evidence.
Example Discussion:
In this experiment, the target peak was clearly observed in the TIC and was sufficiently separated from adjacent peaks.
In addition, the mass spectrum showed a molecular-ion peak corresponding to the molecular weight of the candidate compound and fragment peaks that could be explained from the structure.
Because the library-search results, retention time, and fragmentation pattern did not contradict one another, this peak was considered to originate from the target component.
Example Discussion When the Experiment Did Not Go Well
When GC-MS analysis does not go well, possible causes are considered from results such as an invisible molecular-ion peak, low library-match score, overlapping peaks, large background, a large solvent peak, shifted retention time, or fragments that do not agree with the candidate structure.
Organizing the possible causes into GC separation, MS conditions, sample preparation, concentration, contamination, and limitations of library searches makes the discussion easier.
Example Discussion:
In this experiment, the library-search match score was not sufficiently high.
One possible cause is that another component overlapped with the target peak and fragments originating from multiple components were mixed together.
In addition, because the molecular-ion peak was weak and the search was based mainly on the major fragments, multiple candidate compounds may have been suggested.
For more reliable identification, the GC conditions must be changed to improve peak separation, and retention time and mass spectrum should be compared with those of a standard substance.
How to Write Points for Improvement
In a GC-MS discussion, including not only sources of error but also points for improvement makes the report easier to organize.
Improvements can be divided into sample preparation, GC separation, MS measurement, and data analysis.
Improvements to Sample Preparation
- Dilute the sample to an appropriate concentration
- Prevent loss of volatile components
- Store the sample in a sealed container
- Remove nonvolatile components and solids
- Perform a blank measurement
- Use solvents and containers with little contamination
Improvements to GC Separation
- Review the column-temperature conditions
- Adjust the temperature ramp rate
- Optimize the carrier-gas flow rate
- Use an appropriate injection volume
- Search for conditions that avoid peak overlap
- Consider a column with another stationary phase
Improvements to MS Measurement
- Set an appropriate measured m/z range
- Check the background
- Check contamination of the ion source
- Consider a soft-ionization method when necessary
- Measure and compare a standard substance
Improvements to Analysis
- Confirm the molecular-ion peak
- Assign the major fragment peaks
- Check isotope peaks
- Do not accept library-search results uncritically
- Judge the result using both retention time and mass spectrum
- Use extracted-ion chromatograms
Example of How to Write Points for Improvement:
To improve identification accuracy by GC-MS, the GC conditions must first be adjusted so that the target component and other components are sufficiently separated.
In addition, it is important to perform a blank measurement and exclude peaks originating from solvents, column bleed, or contamination.
In the analysis, the validity of candidate compounds must be judged comprehensively from not only library-search results but also molecular-ion peaks, fragment peaks, isotope peaks, and retention time.
Difference Between a Superficial Discussion and a Good Discussion
In a GC-MS discussion, simply writing that “it appeared in the library search” or that “there was a molecular ion” results in a superficial discussion.
A good discussion relates retention time, molecular-ion peaks, fragments, isotope peaks, and library-search results.
| Superficial Discussion | Good Discussion |
|---|---|
| It matched in the library search. | A candidate compound was suggested by the library search, and because the molecular-ion peak, major fragment peaks, and retention time did not contradict the candidate compound, it was considered highly likely to be the target component. |
| There was a molecular-ion peak. | Because a peak was observed at an m/z corresponding to the molecular weight of the candidate compound, this peak may be the molecular-ion peak. The molecular-ion peak provides evidence for molecular-weight estimation. |
| Fragments appeared. | The observed fragment peaks can be explained as ions produced by cleavage of specific bonds in the candidate compound, supporting the validity of the candidate structure. |
| The component could not be identified. | Because the molecular-ion peak was weak and overlap of multiple component peaks was also suspected, the reliability of the library-search result was considered to have decreased. More reliable identification requires improved separation conditions and confirmation using a standard substance. |
Examples of Expressions That Can Be Used in Reports
The following expressions can be used when writing the results and discussion of GC-MS.
Adjust the necessary parts according to your own experimental results.
- In GC-MS, combining retention time and mass spectrum improves the reliability of component identification.
- The molecular-ion peak provides important evidence for estimating the molecular weight of a compound.
- In the EI method, molecules readily fragment, so the molecular-ion peak may become weak.
- The base peak is the peak with the greatest intensity in a mass spectrum and often corresponds to a stable fragment ion.
- Fragment peaks are considered to originate from ions produced by cleavage of specific bonds in the molecule.
- The isotope-peak pattern provides clues to the presence of particular elements such as chlorine or bromine.
- Library-search results are useful, but identification cannot be confirmed from the search result alone.
- If peaks overlap, ions originating from multiple components may be mixed in the mass spectrum.
- An extracted-ion chromatogram can be used to track an m/z characteristic of the target component.
- If both retention time and mass spectrum agree with those of a standard substance, the reliability of identification becomes high.
Points to Check When Discussing GC-MS
Checking the following points before writing the report makes the discussion easier to write.
- Have the retention times of peaks in the TIC been checked?
- Has the mass spectrum of each peak been checked?
- Has the molecular-ion peak been searched for?
- Has the possibility of a weak molecular-ion peak been considered?
- Have the base peak and major fragment peaks been organized?
- Has it been checked whether the fragments can be explained from the candidate structure?
- Has the isotope-peak pattern been checked?
- Are library-search results evaluated with supporting evidence?
- Are both retention time and mass spectrum used for judgment?
- Have the effects of peak overlap and background been considered?
- Have comparisons with standard substances and blanks been performed?
- Do the points for improvement correspond to the sources of error?
Summary
GC-MS is an analytical method that combines component separation by GC with mass-spectrum measurement by MS.
GC provides retention time, while MS provides molecular-ion peaks, base peaks, fragment peaks, and isotope peaks.
Combining this information makes it possible to identify components more reliably than with GC alone.
Molecular-ion peaks are important for estimating molecular weight, but in the EI method the molecular ion may become weak.
Fragment peaks provide clues to bond cleavage and stable partial structures in the molecule.
Isotope peaks are useful for discussing the number of carbon atoms and the presence of elements such as chlorine and bromine.
In a report, rather than simply writing that “it matched in the library search,” organize and discuss retention time, molecular-ion peaks, major fragments, isotope patterns, comparison with standard substances, and the effects of peak overlap and background.
In GC-MS, confirming that multiple pieces of evidence do not contradict one another leads to persuasive identification.
